Mean differences between rCT and sCT for metrics calculated from acoustic simulations presented as mean ± standard deviation. TR: time reversal.

Simulation Peak pressure (%) Focal position (mm) Focal volume (%)
No correction 12.4±8.1 1.3±1.2 25.5±20.5
Kranion 7.3±6.0 0.6±0.7 20.4±23.5