Skip to main content
. 1998 Sep;36(9):2686–2689. doi: 10.1128/jcm.36.9.2686-2689.1998

TABLE 3.

Comparison of recovery rates of probable contaminants in FAN and 80A blood culture bottles

Taxon No. of isolates recovered by:
P value
FAN and 80A FAN only 80A only
Non-S. aureus staphylococci 75 132 91 0.01
Micrococcus spp. 0 5 1 0.22
Staphylococcus aureus 1 3 0 0.25
Viridans group streptococci 3 8 6 0.79
Enterococci 2 2 9 0.07
Corynebacterium spp. 2 6 12 0.24
Bacillus spp. 0 4 3 1.00
Enterobacteriaceaea 0 6 2 0.29
Other gram-negative bacillib 4 3 4 1.00
Moraxella-Neisseria 1 4 0 0.13
Anaerobesc 1 3 3 0.68
Fungid 1 1 3 0.62
 Totals 90 177 134 0.02
a

Five Escherichia coli, one Enterobacter cloacae, one Klebsiella pneumoniae, and one K. oxytoca isolate. 

b

Two Acinetobacter, one Capnocytophaga, one Comamonas, one Eikenella corrodens, one Flavimonas, one Pseudomonas putida, and one Stenotrophomonas maltophilia isolate and three miscellaneous gram-negative bacilli. 

c

Four Propionibacterium acnes, one Bacteroides fragilis group, one Clostridium perfringens, and one Peptostreptococcus isolate. 

d

Two Candida albicans, one Alternaria, one Rhizopus, and one Penicillium isolate.