Table 1.

Summary of studies evaluating ECV quantification with CT, detailing scan, reconstruction, and contrast media protocols

Scanner type Non-enhanced scan Timing of LE scan [min] Slice thickness/ Tube voltage ECG-gating Contrast media [mgI/mL] Dose of contrast media Effective radiation dosea
Single-energy protocols
 Nacif, 2012 [14] Aquilion One (Toshiba) Yes 10 3 mm 120 kVp Prospective Iopamidol (370) Bolus of 125 mL ± 24 mL 1.98 ± 0.16 mSv for non-enhanced and LE scan
 Bandula, 2013 [12] Somatom Sensation 64 (Siemens) Yes 25 5 mm 120 kVp Retrospective Iohexol (300) Bolus of 1 mL/kg body weight + infusion of 1.88 mL/kg/h 10.7 ± 3.4 mSv (total cardiac CT protocol)
 Hamdy, 2018 [20] Somatom Force (Siemens) Yes 3, 5, 7 1 mm 80 kVp Prospective Iopamiron (370) Bolus of 1.6 mL/kg body weight 11.8 ± 2.9 mSv (total cardiac CT protocol)
 Scully, 2020 [21] Somatom Force (Siemens) Yes 3, 5 2 mm 80 kVp Prospective Iohexol (300) Bolus of 90 mL 5.1 ± 0.3 mSv for non-enhanced and LE scan
 Tamarappoo, 2020 [10] Somatom Definition Flash (Siemens) Yes 5 3 mm 120 kVp Prospective Iohexol (350) Bolus of 100 mL 25.7 ± 6.9 mSv (total cardiac CT protocol)
Dual-energy protocols
 Lee, 2016 [18] Somatom Definition Flash (Siemens) No 12 0.75 mm 100/140 kVp Retrospective Iopamidol (370) Bolus of 1.8 mL/kg body weight 5.59 ± 0.85 mSv (total cardiac CT protocol)
 Abadia, 2019 [15] Somatom Force (Siemens Healthcare) No 7 1.5 mm 90/150 kVp Prospective Iopamidol (370) Bolus of 65 mL 3.61 mSv [2.78–4.21] (total cardiac CT protocol)
 Oda, 2019 [22] iQon Spectral CT (Philips) No 7 0.67 mm 120 kVp Prospective Iopamidol (370) Bolus of 1.5 mL/kg body weight 4.8 ± 1.6 mSv for LE scan
 Ohta, 2020 [23] Discover CT 750 HD (GE) No 7–8 0.625 mm 80/140 kVp Prospective Iopamidol (370) Bolus of 0.9 mL/kg body weight + infusion until 1.4 mL/kg/scan 3.42 ± 0.87 mSv for LE scan
 Dubourg, 2021 [17] Discover CT 750HD (G) No 7 0.625 mm 80/140 kVp Prospective Iohexol (350) Bolus of 65 mL 1.89 ± 0.38 mSv for LE scan
 Qi, 2022 [24] Somatom Force (Siemens) No 7 0.6 mm 90/150 kVp Prospective Iopromide (370) Two boli of 50 mL + 50 mL 5.9 mSv [4.5–8.4] (total cardiac CT protocol)
Photon-counting detector CT scanner
 Mergen, 2022 [25] NAEOTOM Alpha (Siemens) Yes 5 1.5 mm 120 kVp Prospective Iopromide (370) Bolus of 100 mL 1.2 mSv [0.97–1.75] for LE scan

LE Late enhancement

aEffective radiation dose was calculated by multiplying the DLP by a factor of 0.014, if not available in the publication