B1 Field inhomogeneity correction for qDESS T2 mapping: application to rapid bilateral knee imaging

Marco Barbieri 1, Lauren E Watkins 1,2, Valentina Mazzoli 1, Arjun D Desai 1,3, Elka Rubin 1, Andrew Schmidt 1, Garry Evan Gold 1,2, Brian Andrew Hargreaves 1,2,3, Akshay Sanjay Chaudhari 1,4, Feliks Kogan 1
PMCID: PMC10524110  NIHMSID: NIHMS1899479  PMID: 37142852

Abstract

Purpose

T2 mapping is a powerful tool for studying osteoarthritis (OA) changes and bilateral imaging may be useful in investigating the role of between-knee asymmetry in OA onset and progression. The quantitative double-echo in steady-state (qDESS) can provide fast simultaneous bilateral knee T2 and high-resolution morphometry for cartilage and meniscus. The qDESS uses an analytical signal model to compute T2 relaxometry maps, which require knowledge of the flip angle (FA). In the presence of B1 inhomogeneities, inconsistencies between the nominal and actual FA can affect the accuracy of T2 measurements. We propose a pixel-wise B1 correction method for qDESS T2 mapping exploiting an auxiliary B1 map to compute the actual FA used in the model.

Methods

The technique was validated in a phantom and in vivo with simultaneous bilateral knee imaging. T2 measurements of femoral cartilage (FC) of both knees of six healthy participants were repeated longitudinally to investigate the association between T2 variation and B1.

Results

The results showed that applying the B1 correction mitigated T2 variations that were driven by B1 inhomogeneities. Specifically, T2 left–right symmetry increased following the B1 correction ρc=0.74>ρc=0.69. Without the B1 correction, T2 values showed a linear dependence with B1. The linear coefficient decreased using the B1 correction (from 24.3 ± 1.6 ms to 4.1 ± 1.8) and the correlation was not statistically significant after the application of the Bonferroni correction (p value > 0.01).

Conclusion

The study showed that B1 correction could mitigate variations driven by the sensitivity of the qDESS T2 mapping method to B1, therefore, increasing the sensitivity to detect real biological changes. The proposed method may improve the robustness of bilateral qDESS T2 mapping, allowing for an accurate and more efficient evaluation of OA pathways and pathophysiology through longitudinal and cross-sectional studies.

Keywords: QMRI, QDESS, T2 mapping, B1 correction, Bilateral knee imaging

Introduction

Knee osteoarthritis (OA) remains a leading cause of disability in developed countries. Despite its prevalence and large socioeconomic burden, OA remains a poorly understood disease, with multiple pathways and risk factors leading to the onset of this debilitating disease. Qualitative Magnetic Resonance Imaging (MRI) can effectively assess OA-related morphological changes in the knee joint, such as cartilage loss, osteophytes, and bone marrow lesions. However, these structural changes are observed at a late stage of the disease, when treatments are unlikely to be effective. There is thus a goal to detect cartilage microstructure changes at an earlier time point when therapies are more likely to be effective. Quantitative MRI methodologies have shown promise to provide biomarkers sensitive to early OA changes that can help to study early OA onset and progression as well as to monitor response to treatments [1, 2].

The transverse relaxation time T2 has been demonstrated to be sensitive to collagen microstructure and hydration changes in knee articular cartilage and has shown promise for evaluating early changes that are predictive of OA progression [1, 3]. Furthermore, for studying knee OA, bilateral examinations are commonly performed in order to evaluate OA-related changes [4], as well as to investigate the role of between-knee asymmetry in OA [5, 6]. For example, people sustaining a complete anterior crucial ligament (ACL) tear are more likely to develop post-traumatic OA in the injured knee. The contralateral uninjured knee is often used as an internal reference to evaluate longitudinal changes and study post-traumatic OA pathophysiology [7]. The 3D double-echo in steady-state (DESS) pulse sequence [8–10] has been widely applied in musculoskeletal imaging due to its image contrast and high SNR efficiency. DESS enables high-resolution imaging of cartilage, including assessment of cartilage thickness and defects [11–13]. Separation of the two echoes in the DESS sequence in a method known as quantitative DESS [14, 15] (qDESS) can further provide accurate and reproducible T2 measurements [16, 17] by fitting their signals to an extended phase graph signal model [18–20]. In addition, the SNR efficiency and multiple contrasts provided by this approach have shown promise for bilateral clinical knee assessment in 5 min or less [5, 21–23]. However, as for other T2 mapping methods, T2 estimation accuracy is affected by field inhomogeneities [20, 24]. Specifically, qDESS T2 mapping uses an analytical signal model to compute T2 relaxometry maps [16], which requires knowledge of the flip angle (FA). The presence of B1 inhomogeneities leads to inconsistencies between the prescribed and actual FA provided to the spin ensemble at each voxel, which affects the accuracy of T2 measurements. B1 can be a confounding factor both for cross-sectional and longitudinal studies since B1 inhomogeneities introduce bias and variance in T2 mapping. A B1 correction for qDESS T2 mapping may be particularly relevant for bilateral knee acquisition, where B1 inhomogeneities may be larger than single knee acquisitions because of excitation over a larger field of view (FOV) and knee positioning that is further removed from isocenter.

In this work, we aim to investigate to what extent B1 inhomogeneities affect qDESS T2 mapping accuracy and to what extent applying a B1 correction is able to mitigate those effects. For B1 correction, we propose a pixel-wise method for qDESS T2 mapping that exploits an auxiliary B1 map to compute the actual FA to be used in the analytical model for computing the T2 estimate. We validated the B1 correction technique with phantom and in vivo measurements. Subsequently, we investigated our primary aim with cross-sectional and longitudinal T2 measurements of femoral cartilage (FC) in simultaneous bilateral knee imaging.

Theory

Quantitative DESS, represented in Fig. 1A, is a gradient-spoiled steady-state sequence that samples two echoes per repetition TR separated by a spoiler gradient. The two signals have different contrast mechanisms and are functions of the extrinsic scan parameters (TR, TE, flip angle α, gradient amplitude G, and gradient duration τ) as well as the intrinsic tissue parameters (T1, T2, and diffusivity).

Fig. 1.

Fig. 1

A qDESS pulse sequence diagram with phase graph representation. B T2 error as a function of B1 ratio, defined as the ratio between the nominal and actual FA, obtained running Bloch simulations of the qDESS sequence for a voxel composed of cartilage (T1=1400  ms and T2=45  ms, D=1.25×10−9m2/s). Estimation of T2 was performed using Eq. 1 considering a nominal FA = 20°

Sveinsson et al. [16] proposed a simple analytic model that can be used for T2 estimation from the ratio between the second and first echoes of qDESS. The model is expressed in Eq. 1:

S2S1=e−2TR−TET2−TR−τ3Δk2Dsin2α21+e−TRT1−TRΔk2D1+cos αe−TRT1−TRΔk2D. (1)

Here, TR and TE represent the repetition and the first echo time, respectively, α is the flip angle, D is the diffusivity, and the dephasing per unit length induced by the unbalanced gradient is denoted by Δk=γGτ where G and τ are the spoiler amplitude and duration, respectively, and γ is the gyromagnetic ratio. When applying Eq. 1 to femoral cartilage, T1 and D are usually fixed to 1200 ms and 1.25×10−9m2/s, respectively. This simple model, which allows estimating T2 from qDESS, also allows one to appreciate the mechanism for which B1 inhomogeneities can introduce errors in qDESS T2 mapping. The FA must be known to correctly estimate T2 from Eq. 1. B1 inhomogeneities across the imaged volume directly affect the actual FA experienced by the spins within voxels introducing inaccuracies in T2 estimation if not taken into account. If one defines the B1 ratio as the ratio between the nominal and actual FAs, then an auxiliary B1 map can be used to input, voxel-wise, the actual FA in Eq. 1 and correct for B1 inhomogeneities.

Figure 1B reports Bloch simulations run to show the dependence of the T2 estimation error as a function of the ratio between the nominal and actual FA, the B1 ratio, experienced by the spin ensemble. Bloch simulations were run in MATLAB (version 20a, Mathworks, Natick, MA) using an in-house library with the following parameters:T1=1400 ms, T2=45 ms, D=1.25×10−9m2/s, nominal FA = 20°, TE=5 ms and TR=15 ms.

Methods

MRI acquisitions

All MRI acquisitions were performed on a GE 3T SIGNA Premier scanner (GE Healthcare, Milwaukee, WI, USA) using 16-channel flexible phased-array receive-only coils (NeoCoil, Pewaukee, WI, USA). Informed consent was obtained from all subjects in accordance with the Institutional Review Board protocol at our institution. Simultaneous bilateral scans used two identical coils wrapped around each knee.

The MRI protocol consisted of a 3D qDESS sequence for T2 mapping with a 20-degree FA. Data for generating a B1 map, performed using a double-angle method [25] (DAM), was acquired with magnetization prepared (MP-DAM) sequence [26–28] with two flip angles (30° and 60° flip angle excitation pulses followed by a crusher gradient) followed by elliptic centric 3D SPGR Readout. A diagram of the pulse sequences is reported in Fig. S1 of Supplementary Information. The 3D qDESS used a spectrally selective excitation with nominal flip angle α= 20° and a spoiler gradient, which had a spoiler moment of 3.13 mT/m ms with duration τ=1.6  ms. Other sequence parameters varied depending on the type of acquisition (phantom vs in vivo, unilateral vs bilateral) and are summarized in Table 1.

Table 1.

Summary of the sequence parameters

Sequence Phantom
In vivo unilateral
In vivo bilateral
parameter qDESS MP-DAM qDESS MP-DAM qDESS MP-DAM
Scan plane Sagittal Sagittal Sagittal Sagittal Sagittal Sagittal
Echo times (ms) 5/25 1.3 5/25 2 5/25 2
TR (ms) 15 3.67 15 4.5 15 4.5
FOV (cm) 16 × 16 × 12 16 × 16 × 12 16 × 16 × 12 16 × 16 × 12 16 × 16 × 36 16 × 16 × 36
Acquisition matrix 128 × 128 128 × 128 320 × 320 256 × 256 320 × 320 256 × 256
Slice thickness (mm) 2 2 1.4 4 1.4 4
Number of slices 64 64 120 42 240 84
Acquisition time (m:s) 0:50 0:40 1:03 0:54 2:40 1:27

The parameters for MP-DAM are for the readout of the longitudinally stored magnetization prepared signal with 30°/60° flip angles

Phantom measurements

A T2 phantom made of three pairs of vials containing 2%, 3%, and 4% agarose, with T2 ranging from 30 to 70 ms, was scanned. T2 maps computed from qDESS with and without using the B1 corrections were compared to validate the B1 correction method.

The MRI phantom protocol was repeated with modifications to the transmit gain (TG) of the pulse transmitter. A variation in the TG modifies the B1 experienced by the spins in each voxel proportionally to the extent of the variation. The range of TG values was selected so that the smallest and largest TG values corresponded to a 20% reduction and increase with respect to the default TG value, respectively. qDESS T2 maps were computed with and without applying the B1 correction, and within each vial of the phantom the linear correlation between T2 and B1 was assessed.

In vivo knee imaging

The knee of a healthy volunteer was repeatedly scanned unilaterally, varying the TG. The range of TG values was selected so that the smallest and largest TG values corresponded to a 20% reduction and increase with respect to the default TG value, respectively. The linear dependence of T2 on B1 values was assessed using a linear fit, which was performed with and without applying the B1 correction.

Six healthy volunteers (4 females and 2 males) were scanned using a simultaneous bilateral knee acquisition [21], once a week for 8 weeks, to evaluate the effect of the B1 correction on T2 longitudinal stability and between-knee asymmetry. During the second week, each subject was scanned once a day for 5 days.

Data processing and analysis

For in vivo longitudinal bilateral knee acquisitions, with reference to Fig. 2, for each subject within each timepoint, the following data processing protocol was applied.

Fig. 2.

Fig. 2

Pipeline used to process the longitudinal in vivo bilateral acquisitions. For each subject, within each time point B1 map was first obtained from the MP-DAM images and smoothed with a 2D median filter. The FC was automatically segmented from the qDESS images and subdivided into 6 sub-regions: anterior/central/posterior for the medial/lateral sides. The B1 map was then co-registered to qDESS and T2 maps were then obtained with and without B1 correction and average quantitative values computed within each FC sub-region. Finally, for visualization, the 3D segmented volume was projected onto a 2D space

  1. The B1 map was computed from the MP-DAM acquisition using an in-house script developed in MATLAB (version 20a, Mathworks, Natick, MA), which can be found in the public repository that contains the data of this work (see data availability statement). A customized 2D median filter with intensity thresholding was applied to smooth the B1 map. Thus, within the kernel, only pixels for which their intensity was above a set threshold value were considered in the computation of the median. This aimed to avoid bias in the median due to pixels of very low SNR, such as those in the bone/cartilage interface (see arrow in Fig. 2 panel 1). Since the DAM is magnitude based, in such region the estimation of B1 is inaccurate. Specifically, at each step of the convolution, the median was evaluated on the B1 map, but the decision of which pixels to include in the computation was performed on the raw MP-DAM images belonging to the first flip angle acquisition. The threshold was automatically adjusted for each acquisition, as it was set to the 75-th percentile of the image intensity values. The selection of the percentile threshold was done empirically and aimed to assure the exclusion of the pixels belonging to the cartilage-bone region.

  2. The femoral cartilage (FC) was automatically segmented from the qDESS acquisition using the open-source Deep Open-Source Medical Image Analysis (DOSMA) framework [29, 30](https://github.com/ad12/DOSMA), and further subdivided into 6 sub-regions: anterior/central/posterior for the medial/lateral sides.

  3. The B1 map was then registered to the first echo of qDESS through a rigid transformation using Elastix [31, 32], so that B1 and T2 maps were co-registered for each acquisition.

  4. T2 maps were computed on a pixel-by-pixel basis by analytically inverting the qDESS signal model [16] expressed in Eq. 1, both using the nominal FA and the actual FA computed by multiplying the nominal FA with the B1 ratio estimated from the auxiliary double-angle acquisition. In performing the computation, T1 and D were fixed to 1200 ms and 1.25×10−9m2/s, respectively.

  5. The average T2 values, with and without B1 correction, and B1 values within each FC sub-region were then computed and used to perform statistical analysis.

  6. For visualization purpose only, the 3D maps were projected onto a 2D space (Fig. 2 panel 6). To visualize the between-knee asymmetry. Pixel-wise subtraction was evaluated by mirroring and registering the right 2D projected FC to the left 2D FC projection for direct comparison.

For all the phantom and unilateral in vivo knee measurements, the procedure used to obtain T2 and B1 maps was the same as described above with the only exception that the customized 2D smoothing filter was not applied for the phantom experiments. For the phantom, the T1 for T2 computation was fixed to 1600 ms, which had been previously estimated using a variable flip angle T1 mapping method.

Statistical analysis

For the phantoms and in vivo knee experiments, the dependence of T2 values, obtained with and without applying the B1 correction, on B1 was assessed by fitting a linear model to the experimental data. The linear coefficient of the model, a, also referred to as slope, was used to assess the strength of the linear dependence between T2 and B1, and the fit p-value was used to test the statistical significance of the linear correlation.

For the in vivo longitudinal bilateral acquisitions, a repeated measurement Analysis of Variance (RMANOVA) was used to determine the within-subject standard deviation (SD), also called measurement error SD, of three parameters across time points and in each FC sub-region. The parameters were: T2 without B1 correction, T2 with B1 correction and B1. The coefficient of repeatability (RC) and the coefficient of variation (CV) were also computed to characterize longitudinal stability. The rationale for considering each FC sub-region separately was to eliminate variance related to differences in T2 values across FC sub-regions. To assess the dependence of T2 on B1, the following multivariate linear regression model was used:

T2=β0+β1×Knee+β2×Sub-region+β3      ×Subject+β4×Time+β5×B1. (2)

The statistical significance of each coefficient was tested with a significance level of α=0.05, and the Bonferroni correction was used to set the significance threshold for the p value. Since 5 predictors were considered, the significant threshold was set to 0.05/5, which is equal to 0.01. The model was fitted to T2 values obtained with and without B1 correction.

Finally, Bland–Altman (BA) analysis was used to assess T2 left–right asymmetry with and without using the B1 correction. Lin’s concordance correlation coefficient ρc and the coefficient of variation of the root mean squared error (CVRMSE) were also evaluated as global similarity metrics.

All the statistical analyses were run using in-house scripts developed in Python.

Results

Phantom validation

The Phantom T2 maps, obtained with and without applying the B1 correction, are shown in Fig. 3, along with the B1 map and the comparison among T2 distributions within each pair of vials (Fig. 3B). B1 values are dimensionless as they are expressed as the ratio between the actual and nominal FA. The B1 map showed a variation in the B1 across the phantom within the range [0.8, 1.1], which corresponds roughly to ± 20% FA variations with respect to the nominal value. The spatial variation of B1 was such that for each pair of vials, one of the two vials experienced a lower B1 than the other. The T2 map computed without applying the B1 correction showed that in each pair of identical vials, an asymmetry in the T2 values between the two vials was observed. This can be better appreciated by looking at the T2 distributions (first row of Fig. 3B). When the T2 map is computed by applying the proposed B1 correction, the T2 differences between paired vials in the phantom were greatly reduced. This can be seen in the appearance of the 2D T2 maps and by observing the overlapping histograms of the T2 distributions (second row of Fig. 3B).

Fig. 3.

Fig. 3

A Sketch of the T2 phantom (left) along with its T1 map, and T2 map (bottom right) computed from qDESS both with and without correcting for B1 (top right) that consisted of roughly ± 20% FA variations. B1 values are dimensionless, being the ratio between the actual and nominal flip angles. The T2 phantom was designed to have 3 pairs of vials with different T2 relaxation times. B Comparisons between the T2 distributions of the paired samples without (top row) and with (bottom row) B1 corrections demonstrate that using B1 correction reduces errors in the paired T2 calculations

The correlation plots between T2 and B1 values for each vial of the T2 phantom, obtained with various TGs, are shown in Fig. 4. The linear coefficient and the associated fit p-value are reported for each correlation plot. When T2 was computed without correcting for B1, the linear coefficients were greater than or equal to 21 ms (21–50 ms), with p values (p) less than or equal to 0.001. When the B1 correction was applied, the slopes of the linear fit were reduced, and in almost all vials, the correlation was not statistically significant (p > 0.05). For vials 2 and 6, the correlations were still statistically significant after applying the B1 correction (p < 0.05), but the slopes of the linear fit were substantially reduced compared to the case in which the B1 correction was not applied: the slopes decreased from 41 ms to 8.5 ms and from 50 ms to 8.4 ms for vials 2 and 6, respectively.

Fig. 4.

Fig. 4

Correlation plots between T2 and B1 for each phantom vial acquired varying the TG. The first column reports the correlation plots obtained using T2 values computed without applying the B1 correction, whereas in the second column T2 values are computed using the B1 correction. Points represent the average parameter values within the vial, while error bars represent the standard deviation of parameter values within the vial

Validation in vivo

The correlation plots between T2 and B1 values in FC sub-regions obtained with varying TG weights are shown in Fig. 5. For each FC sub-region, the correlation plot obtained using T2 values computed without applying the B1 correction is drawn in blue, whereas red is used for T2 values computed using the B1 correction. The points represent the average parameter values within the sub-region, while the error bars represent the standard deviation. The linear coefficient and the fit p-value are reported for each correlation plot.

Fig. 5.

Fig. 5

Correlation plots between T2 and B1 in each FC sub-regions acquired varying the TG. For each FC sub-region, the correlation plot obtained using T2 values computed without applying the B1 correction is drawn in blue, whereas red is used for T2 values computed using the B1 correction. Points represent the average parameter values within the sub-region, while error bars represent the standard deviation of parameter values within the sub-region

Measured B1 values varied from 0.75 to 1.3. Without applying the B1 correction, for all sub-regions, a statistically significant linear correlation between T2 and B1 values was observed (p ≤ 0.002) and a slope greater than or equal to 18 ms. When the B1 correction was applied, the linear coefficients were greatly reduced for all the sub-regions. Furthermore, the correlations were not statistically significant in all the FC sub-regions (p > 0.05), with the only exception of the lateral posterior region (p = 0.01).

In vivo bilateral knee acquisitions

Longitudinal variation

Subjects #5 and 6 did not complete the study and only data from the first 6 weeks were available for these subjects. Furthermore, for subject #3, the MP-DAM data did not correctly reconstruct at the scanner for the first time point, which was then discarded from the analysis. Hence, the RMANOVA test was run over 8 total time points (from week #2 to week #6) for which data from all subjects were available.

For each parameter, the RMANOVA test was unable to reject the null hypothesis with a significance value of 0.05. The summary statistics of the repeated measurement analysis are reported for the left and right knee in Table 2. The value of each statistic is reported as the averaged value across different sub-regions, and the standard deviation across sub-regions is also reported. Overall, B1 was relatively stable during time points, with a CV of 4.1% both in the left and right knee. Without the B1 correction, the average within-subject SD was 1.8 ms and 1.3 ms for the left and right knees, respectively, and the corresponding RCs were 5.1 ms and 3.6 ms. The CV was 6% in both knees. When the B1 correction was applied, the average within-subject SD was 1.9 ms and 1.5 ms for the left and right knees, respectively. The RC was 5.3 ms and 4.1 ms in the left and right knee, respectively. The CV was 6% in both knees.

Table 2.

Summary statics for the longitudinal in vivo simultaneous bilateral knee acquisitions on six healthy volunteers

Statistic (avg ± sd) Left knee
Right knee
T2
B1 T2
B1
W/O Corr W/Corr W/O Corr W/Corr
Within-subject SD (1.8 ± 0.4) ms (1.9 ± 0.5) ms 0.04 ± 0.01 (1.3 ± 0.3) ms (1.5 ± 0.1) ms 0.03 ± 0.01
RC (5.1 ± 2.5) ms (5.3 ± 3.1) ms 0.1 ± 0.1 (3.6 ± 0.8) ms (4.1 ± 0.4) ms 0.1 ± 0.1
CV (%) 6.0 ± 1.1 6.0 ± 1.2 4.1 ± 0.6 4.4 ± 0.6 4.6 ± 0.4 4.1 ± 0.7

The value of each statistic is reported as the mean value averaged across the six FC sub-regions, and the standard deviation across FC sub-regions is also reported

Dependence of T2 on B1

The linear coefficients for the multivariate linear regression model fitted to the T2 values obtained with and without B1 correction are reported in Table 3. The p value, R-squared R2, and its relative importance on the total model R2 are reported for each linear coefficient. Without applying the B1 correction, the linear regression model explained 26% of the variance (total R2=0.26). Out of the 5 explanatory variables, B1 had the most relative importance, explaining 64.4% of the total R2. The associated linear coefficient was equal to (24.3 ± 1.6) ms with a p value of 7×10−47, showing a positive linear correlation between T2 and B1. Of the other explanatory variables, the FC Sub-region and Knee had the highest relative importance: 28.8% and 5.3%, respectively. The remaining variables were progressively negligible, and the linear coefficient associated with the variable Time was not statically significant (p value > 0.01).

Table 3.

Summary of the multivariate linear regression model expressed in Eq. 2 run on T2 data with and without applying the B1 correction

Variable β±SE p value R2 R2 rel. imp. (%)
Without B1 correction
Intercept 9.7 ± 1.5 1.3×10−10 – –
Subject 0.20 ± 0.07 4×10−3 0.003 1.2
Knee 1.1 ± 0.3 6×10−5 0.014 5.3
Sub-region −0.78 ± 0.07 5×10−28 0.075 28.8
Time −0.02 ± 0.03 0.5 0.0003 0.1
B1 24.3 ± 1.6 7×10−47 0.17 64.6
Total: 0.26 100
With B1 correction
Intercept 30.2 ± 1.7 4×10−62 – –
Subject 0.26 ± 0.08 6×10−4 0.01 6.6
Knee 1.03 ± 0.29 4×10−4 0.008 5.2
Sub-region −0.90 ± 0.08 3×10−29 0.13 84.3
Time −0.04 ± 0.04 0.3 0.001 0.9
B1 4.1 ± 1.8 0.02 0.005 2.9
Total 0.15 100

For each variable of the model, the linear coefficient β is reported along with its p value, R2, and the relative importance of such R2 on the total R2 of the model. Following the Bonferroni correction, the threshold for statistical significance was set to 0.01

When the B1 correction was applied, the total explanatory power of the regression model was reduced to 15% (total R2=0.15). Specifically, the linear coefficient associated with B1 was reduced to (4.1 ± 1.8) ms, explaining only the 2.9% of the total R2 and it was not statically significant (p value > 0.01). The most important explanatory variable was the FC Sub-region, explaining 84.3% of the total R2, with a linear coefficient equal to −0.9 ± 0.08. The remaining variables were progressively negligible, and the linear coefficient associated with the variable Time was not statically significant (p value > 0.01).

T2 left–right knee asymmetry and correlation between T2 and B1

An example of T2 left–right asymmetry for one time point of subject #5 is displayed in Fig. 6. Each row displays: the parameter of interest (B1 and T2 in the top and bottom panels, respectively); the 2D projections of FC for the left and the mirrored right knee; the pixel-wise difference; and the BA plots between the left and right parameter average values in the six FC sub-regions. In Fig. 6, B1 differed by up to 0.3 between regions in the left and right knees, as highlighted by the 2D projections. Left–right symmetry (bottom panel of Fig. 6) increased following a B1 correction, evident through visual inspection of the unrolled T2 maps and the BA analysis. In particular, when the B1 correction was not applied, the mean bias, CV, and ρc were equal to −4.3 ms, 12.3 %, and 0.31, respectively. When the B1 correction was applied, the mean bias, CV, and ρc were equal to 0.5 ms, 7.0 %, and 0.82, respectively.

Fig. 6.

Fig. 6

Example of comparison between the left and right T2 maps of FC where B1 differs substantially between the left and right knee. Each row displays the quantitative 2D projected and unrolled parameter (B1 and T2 in the top and bottom panels, respectively) for the left and the mirrored right knee. The pixel-wise difference and the BA plots between left and right parameter values in the six FC sub-regions are also reported

The BA plots regarding left–right asymmetry for B1 and T2, computed with and without B1 correction, considering all the subjects and time points are reported in Fig. 7. On average B1 ratio differed by about 0.1 between the left and right knees, with LOA equal to ± 0.17 and an RMSECV of 7%. Poor agreement between left and right B1 values was observed as summarized by Lin’s concordance coefficient being equal to 0.19. Regarding T2, the left–right symmetry improved when the B1 was applied compared to the case in which the B1 correction was not applied. Specifically, the RMSECV was reduced from 5 to 4.4% and Lin’s concordance coefficient was increased from 0.69 to 0.74. The limits of agreement were ± 5.2 ms when B1 correction was not applied, and ± 5.4 ms when B1 correction was applied.

Fig. 7.

Fig. 7

BA plots between left and right values for B1 (right), T2 without B1 correction (center), and T2 with B1 correction. All subjects and time points are reported in the plot, and each color represents one of the FC sub-regions. For each plot, the limits of agreement (LOA), the RMSECV and Lin’s concordance coefficient ρc are reported

Discussion

qDESS is becoming a widely utilized tool for studying knee OA due to its capability of providing accurate T2 relaxation time measurements and SNR-efficient morphological bilateral imaging in 5 min or less. As for many other T2 mapping methods, qDESS T2 mapping is sensitive to B1 inhomogeneities, which becomes increasingly important to account for when performing simultaneous bilateral T2 mapping. Therefore, (i) we proposed a pixel-wise B1 correction method for qDESS T2 mapping that exploits an auxiliary B1 map, and (ii) investigated to what extent the correction impacts qDESS T2 mapping in simultaneous bilateral knee imaging.

Our experiments with the T2 phantom demonstrated the errors in T2 calculations that occur with B1 variations. Without B1 correction, qDESS T2 relaxation times for sample pairs varied over the field of view, with a mean variation in the order of 10 ms. This variation was removed when the B1 correction was applied. Similarly, phantom and in vivo experiments run with varying TG showed that the proposed B1 correction was able to considerably mitigate the dependence of qDESS T2 on B1. Furthermore, these experiments demonstrated that the proposed method was able to correct T2 measurements for a wide range of B1 values (0.68–1.34).

The results regarding the in vivo longitudinal bilateral knee acquisitions gave several insights on the impact of B1 on qDESS T2 mapping in vivo. First, Regarding T2 longitudinal repeatability, there was a negligible difference between the case in which the B1 correction was applied compared to the case in which B1 correction was not applied both for the right and left knee. In both cases, the average CVs were considered acceptable considering that they were within the repeatability CVs reported in the literature for qDESS T2 mapping [33]. The longitudinal study was carried out using the same scanner, and each subject was acquired under similar conditions. As a consequence, within each subject, B1 was relatively stable across time as highlighted by the high repeatability metrics reported in Table 2. Although not surprising, these results suggest that B1 correction might be less concerning when dealing with longitudinal studies where the measurement conditions do not change across time points. Nonetheless, the multivariate linear regression analysis showed that the proposed correction method greatly mitigated the dependence of T2 on B1. The interpretation of this result is that even if B1 did not vary significantly across timepoints, it appeared to vary across subjects. Thus, B1 correction may be more important for cross-sectional studies to reduce a systematic source of bias and variability in T2 measurements. Furthermore, the results suggested that asymmetries in T2 relaxation times between the left and right knees of asymptomatic volunteers were partially related to B1 inhomogeneities. It is important to note that there is no clear understanding of whether or not T2 values in left and right knees should be symmetrical in healthy volunteers. There is a lack of studies in the literature addressing this scientific question. The main finding of our study was that left–right asymmetries in B1 appeared to contribute to left–right asymmetries in T2. For example, when considering the results reported in Fig. 6, B1 values were higher in the left knee compared with the right knee. As a consequence, when the B1 correction was not applied T2 values in the left knee were also higher than T2 values in the right knee. Applying the B1 correction removed this bias due to B1, and left–right T2 asymmetry was reduced. The key point is that our work clearly suggests that B1 correction mitigated variations driven by the sensitivity of the T2 mapping method to B1 and did not reflect actual biological differences between knees. This finding is particularly relevant for bilateral scanning and applications at ultra-high fields, but can also help ensure accuracy for unilateral longitudinal scans. Our study showed that T2 errors due to B1 inhomogeneities are in the order of 5 ms or more with a 20% variation in B1, as also demonstrated by Bloch simulations in Fig. 1B. These errors are relevant as studies investigating the longitudinal variation of T2 in cartilage in knee OA over a period of two years have found variations in the order of 1.5–2 ms [34, 35]. Similarly, longitudinal studies of T2 in knee cartilage after ACL reconstruction over a period of 24 months have observed average differences of about 5 ms between T2 at 6 months after surgery and at 24 months [36]. Thus, taking into account B1 inhomogeneities in qDESS T2 mapping might be beneficial for obtaining meaningful results from cross-sectional and longitudinal studies, as well as studies investigating the use of the contralateral knee as a reference.

Although only healthy subjects were considered in this study, the obtained results on the effect of B1, and B1 correction, in qDESS T2 mapping are independent of the underlying knee condition. In fact, preliminary investigations on the application of the same protocol to two subjects that had undergone ACL surgery in one of their knees showed that multivariate linear regression analysis gave results that were consistent with those obtained in healthy volunteers (see Table S1 of supplementary material). Overall, our experiments suggest that implementing a B1 correction should be considered when designing studies investigating T2 with qDESS. Future studies can further investigate the impact of B1 on qDESS T2 mapping when performing multi-center and multi-vendor studies since B1 inhomogeneities might show a higher variance than that observed in our study. Furthermore, this work may provide a protocol to effectively investigate the scientific value of using the contralateral (healthy) knee as an intrinsic reference for studying knee pathologies using bilateral qDESS T2 mapping.

There are some limitations to our methodology. The proposed B1 correction requires an additional acquisition to obtain the auxiliary B1 map. However, B1 mapping can be performed relatively quickly, requiring just about a minute and a half in our bilateral studies. When used in tandem with qDESS, the SNR efficiency of qDESS still allows for reduced scan times compared to non-model based T2 mapping methods, such as MESE which normally require more than 5 min per knee and would still require a B1 correction. The qDESS time efficiency is particularly relevant with simultaneous bilateral acquisitions. Further, additional post-processing is needed to extract B1 corrected T2 relaxation times. However, the additional accuracy in T2 estimation may be critical to obtain meaningful information from cross-sectional and longitudinal studies. It is worth noting that in estimating T2 through Eq. 1, the values of T1 and D were assumed to be constant across the cartilage tissue and subjects. This in principle could introduce biases whenever this condition is not true. In practice, T1 of cartilage is usually considered to be constant across cartilage also in other techniques such as delayed Gadolinium-Enhanced MRI [37, 38]. Furthermore, simulations have shown that for the typical T1 value assumed for cartilage (1200 ms), if the actual T1 is 20% lower or higher than the assumed value, the error in T2 estimation is only 4% [16]. Thus, the method has low sensitivity to T1 variations. The sensitivity of T2 to diffusion is a function of the diffusion coefficient and the spoiling gradient area. In our case, the gradient area was set to the minimum value to achieve full spoiling. Thus, the sequence has minimal sensitivity to diffusion [16] and it is not thought to introduce significant errors in the estimation of T2. However, further studies could study whether in presence of pathological changes the diffusion coefficient of cartilage increases to the point of introducing biases in the estimation of T2 with qDESS. Lastly, it is worth mentioning that in our study, we used receive-only coils, and the transmission was performed with the body coil. Thus, when using transmit–receive coils, the size effect of variations in B1 might be different. Nonetheless, the outlined methodology will work independently from the setup utilized, provided that an accurate B1 map can be acquired.

Conclusions

A B1 correction for qDESS T2 mapping was proposed to improve the accuracy of the method. The correction exploited an auxiliary B1 map to compute the actual FA to be input into the model used to estimate T2. Validation of the methodology was performed in phantom and in vivo. Further, in vivo longitudinal bilateral knee acquisitions demonstrated that the B1 correction mitigated variations that were driven by B1 inhomogeneities rather than reflecting biological differences. The proposed method may improve the robustness of bilateral qDESS T2 mapping, allowing more accurate and efficient evaluation of OA pathophysiology through longitudinal and cross-sectional studies.

Supplementary Material

Supplementary Material

Acknowledgements

This work was supported by GE Healthcare and NIH Grants R01-AR077604, R01-EB002524, R01-AR074492, R21EB030180, K24-AR062068, and R00-EB022634.

Footnotes

Supplementary Information The online version contains supplementary material available at https://doi.org/10.1007/s10334-023-01094-y.

Declarations

Conflict of interest The authors declare no conflicts of interest.

Ethical approval This study received Institutional Review Board approval.

Informed consent Written informed consent was obtained from the volunteer.

Data availability

The MRI data that support the findings of this study are openly available in Zeonodo at https://doi.org/10.5281/zenodo.7478225.

References

  • 1.Liu F, Chaudhary R, Hurley SA, Munoz Del Rio A, Alexander AL, Samsonov A et al. (2014) Rapid multicomponent T2 analysis of the articular cartilage of the human knee joint at 3.0T. J Magn Reson Imaging 39(5):1191–1197. 10.1002/jmri.24290 [DOI] [PubMed] [Google Scholar]
  • 2.Ling W, Regatte RR, Navon G, Jerschow A (2008) Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST). Proc Natl Acad Sci 105(7):2266–2270. 10.1073/PNAS.0707666105 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Monu UD, Jordan CD, Samuelson BL, Hargreaves BA, Gold GE, McWalter EJ (2017) Cluster analysis of quantitative MRI T2 and T1ρ relaxation times of cartilage identifies differences between healthy and ACL-injured individuals at 3T. Osteoarthr Cartil 25(4):513–520. 10.1016/j.joca.2016.09.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Jungmann PM, Brucker PU, Baum T, Link TM, Foerschner F, Minzlaff P et al. (2015) Bilateral cartilage T2 mapping 9 years after Mega-OATS implantation at the knee: a quantitative 3T MRI study. Osteoarthr Cartil 23(12):2119–2128. 10.1016/j.joca.2015.06.013 [DOI] [PubMed] [Google Scholar]
  • 5.Barbieri M, Desai AD, Kogan F, Mazzoli V, Rubin E, Castellani G et al. (2019) Rapid Quantitative Bilateral knee imaging with fully automated femoral cartilage analysis In: Proc. of ISMRM 27th Annual Meeting and Exhibition. Montreal, Canada: Proc. of ISMRM 27th Annual Meeting and Exhibition; 11–16 [Google Scholar]
  • 6.Black M, Younk K, Chaudhari A, Kogan F, Sveinsson B, McWalter E et al. (2020) T2 cluster asymmetry can detect differences in cartilage of ACL-injured subjects 3 months post-surgery In: Proc. of 2020 International Workshop on Osteoarthritis Imaging. Salzburg, Austria: Proc. of 2020 International Workshop on Osteoarthritis Imaging [Google Scholar]
  • 7.Casula V, Tajik BE, Kvist J, Frobell R, Haapea M, Nieminen MT et al. (2022) Quantitative evaluation of the tibiofemoral joint cartilage by T2 mapping in patients with acute anterior cruciate ligament injury vs contralateral knees: results from the subacute phase using data from the NACOX study cohort. Osteoarthr Cartil 30(7):987–997. 10.1016/j.joca.2022.02.623 [DOI] [PubMed] [Google Scholar]
  • 8.Bruder H, Fischer H, Graumann R, Deimling M (1988) A new steady-state imaging sequence for simultaneous acquisition of two MR images with clearly different contrasts. Magn Reson Med 7(1):35–42. 10.1002/MRM.1910070105 [DOI] [PubMed] [Google Scholar]
  • 9.Redpath TW, Jones RA (1988) FADE-A new fast imaging sequence. Magn Reson Med 6(2):224–234. 10.1002/mrm.1910060211 [DOI] [PubMed] [Google Scholar]
  • 10.Lee SY, Cho ZH (1988) Fast SSFP gradient echo sequence for simultaneous acquisitions of FID and echo signals. Magn Reson Med 8(2):142–150. 10.1002/mrm.1910080204 [DOI] [PubMed] [Google Scholar]
  • 11.Eckstein F, Westhoff J, Sittek H, Maag KP, Haubner M, Faber S et al. (2013) In vivo reproducibility of three-dimensional cartilage volume and thickness measurements with MR imaging. Am J Roentgenol 170(3):593–597. 10.2214/AJR.170.3.9490936 [DOI] [PubMed] [Google Scholar]
  • 12.Moriya S, Miki Y, Yokobayashi T, Ishikawa M (2009) Three-dimensional double-echo steady-state (3D-DESS) magnetic resonance imaging of the knee: contrast optimization by adjusting flip angle. Acta Radiologica 50(5):507–511. 10.1080/02841850902849444 [DOI] [PubMed] [Google Scholar]
  • 13.Hardy PA, Recht MP, Piraino D, Thomasson D (1996) Optimization of a dual echo in the steady state (DESS) free-precession sequence for imaging cartilage. J Magn Reson Imaging 6(2):329–335. 10.1002/JMRI.1880060212 [DOI] [PubMed] [Google Scholar]
  • 14.Welsch GH, Scheffler K, Mamisch TC, Hughes T, Millington S, Deimling M et al. (2009) Rapid estimation of cartilage T2 based on double echo at steady state (DESS) with 3 Tesla. Magn Reson Med 62(2):544–549. 10.1002/MRM.22036 [DOI] [PubMed] [Google Scholar]
  • 15.Staroswiecki E, Granlund KL, Alley MT, Gold GE, Hargreaves BA (2012) Simultaneous estimation of T2 and apparent diffusion coefficient in human articular cartilage in vivo with a modified three-dimensional double echo steady state (DESS) sequence at 3 T. Magn Reson Med 67(4):1086–1096. 10.1002/MRM.23090 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Sveinsson B, Chaudhari AS, Gold GE, Hargreaves BA (2017) A simple analytic method for estimating T2 in the knee from DESS. Magn Reson Imaging 38:63–70. 10.1016/j.mri.2016.12.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Chaudhari AS, Black MS, Eijgenraam S, Wirth W, Maschek S, Sveinsson B et al. (2018) Five-minute knee MRI for simultaneous morphometry and T2 relaxometry of cartilage and meniscus and for semiquantitative radiological assessment using double-echo in steady-state at 3T. J Magn Reson Imaging 47(5):1328–1341. 10.1002/JMRI.25883 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Hennig J (1991) Echoes-how to generate, recognize, use or avoid them in MR-imaging sequences. Part I: fundamental and not so fundamental properties of spin echoes. Concepts Magn Reson 3(3):125–143. 10.1002/cmr.1820030302 [DOI] [Google Scholar]
  • 19.Hennig J (1991) Echoes-how to generate, recognize, use or avoid them in MR-imaging sequences. Part II: Echoes in imaging sequences. Concepts Magn Reson 3(4):179–192. 10.1002/cmr.1820030402 [DOI] [Google Scholar]
  • 20.Weigel M (2015) Extended phase graphs: dephasing, RF pulses, and echoes—pure and simple. J Magn Reson Imaging 41(2):266–295. 10.1002/jmri.24619 [DOI] [PubMed] [Google Scholar]
  • 21.Kogan F, Levine E, Chaudhari AS, Monu UD, Epperson K, Oei EHG et al. (2018) Simultaneous bilateral-knee MR imaging. Magn Reson Med 80(2):529–537. 10.1002/MRM.27045 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Chaudhari AS, Stevens KJ, Sveinsson B, Wood JP, Beaulieu CF, Oei EHG et al. (2019) Combined 5-minute double-echo in steady-state with separated echoes and 2-minute proton-density-weighted 2D FSE sequence for comprehensive whole-joint knee MRI assessment. J Magn Reson Imaging 49(7):e183–e194. 10.1002/jmri.26582 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Chaudhari AS, Grissom MJ, Fang Z, Sveinsson B, Lee JH, Gold GE et al. (2021) Diagnostic accuracy of quantitative multicontrast 5-minute knee MRI using prospective artificial intelligence image quality enhancement. Am J Roentgenol 216(6):1614–1625. 10.2214/AJR.20.24172. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Prasloski T, Mädler B, Xiang QS, MacKay A, Jones C (2012) Applications of stimulated echo correction to multicomponent T2 analysis. Magn Reson Med 67(6):1803–1814. 10.1002/MRM.23157 [DOI] [PubMed] [Google Scholar]
  • 25.Stollberger R, Wach P (1996) Imaging of the active B1 field in vivo. Magn Reson Med 35(2):246–251. 10.1002/MRM.1910350217 [DOI] [PubMed] [Google Scholar]
  • 26.Cunningham CH, Pauly JM, Nayak KS (2006) Saturated double-angle method for rapid B1+ mapping. Magn Reson Med 55(6):1326–1333. 10.1002/MRM.20896 [DOI] [PubMed] [Google Scholar]
  • 27.Chung S, Kim D, Breton E, Axel L (2010) Rapid B1+ mapping using a pre-conditioning RF pulse with TurboFLASH readout. Magn Reson Med 64(2):439. 10.1002/MRM.22423 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Watkins LE, Rubin EB, Mazzoli V, Uhlrich SD, Desai AD, Black M et al. (2020) Rapid volumetric gagCEST imaging of knee articular cartilage at 3 T: evaluation of improved dynamic range and an osteoarthritic population. NMR Biomed 33(8):e4310. 10.1002/nbm.4310 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Desai AD, Barbieri M, Mazzoli V, Rubin E, Black M, Watkins L et al. (2019) DOSMA: a deep-learning, open-source framework for musculoskeletal MRI analysis In: Proc. of ISMRM 27th Annual Meeting and Exhibition. Montreal, Canada: Proc. of ISMRM 27th Annual Meeting and Exhibition; 11–16 [Google Scholar]
  • 30.Desai AD, Gold GE, Hargreaves BA, Chaudhari AS (2019. feb) Technical considerations for semantic segmentation in MRI using Convolutional Neural Networks. arXiv:1902.01977v1 [eessIV]. arXiv:1902.01977 [Google Scholar]
  • 31.Klein S, Staring M, Murphy K, Viergever MA, Pluim JPW (2010) Elastix: a toolbox for intensity-based medical image registration. IEEE Trans Med Imaging 29(1):196–205. 10.1109/TMI.2009.2035616 [DOI] [PubMed] [Google Scholar]
  • 32.Shamonin DP, Bron EE, Lelieveldt BPF, Smits M, Klein S, Staring M (2014) Fast parallel image registration on CPU and GPU for diagnostic classification of Alzheimer’s disease. Front Neuro-inform 0(JAN):50. 10.3389/FNINF.2013.00050 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Jordan CD, McWalter EJ, Monu UD, Watkins RD, Chen W, Bangerter NK et al. (2014) Variability of CubeQuant T1ρ, quantitative DESS T2, and cones sodium MRI in knee cartilage. Osteoarthr Cartil 22(10):1559–1567. 10.1016/j.joca.2014.06.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Pan J, Pialat JB, Joseph T, Kuo D, Joseph GB, Nevitt MC et al. (2011) Knee cartilage T2 characteristics and evolution in relation to morphologic abnormalities detected at 3-T MR imaging: a longitudinal study of the normal control cohort from the osteoarthritis initiative. Radiology 261(2):507–515. 10.1148/radiol.11102234 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Kretzschmar M, Heilmeier U, Yu A, Joseph GB, Liu F, Solka M et al. (2016) Longitudinal analysis of cartilage T2 relaxation times and joint degeneration in African American and Caucasian American women over an observation period of 6 years- Data from the Osteoarthritis Initiative. Osteoarthritis Cartilage 24(8):1384. 10.1016/J.JOCA.2016.03.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Lansdown DA, Xiao W, Zhang AL, Allen CR, Feeley BT, Li X et al. (2020) Quantitative imaging of anterior cruciate ligament (ACL) graft demonstrates longitudinal compositional changes and relationships with clinical outcomes at 2 years after ACL reconstruction. J Orthop Res 38(6):1289–1295. 10.1002/JOR.24572 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Williams A, Mikulis B, Krishnan N, Gray M, McKenzie C, Burstein D (2007) Suitability of T1Gd as the “dGEMRIC index” at 1.5T and 3.0T. Magn Reson Med 58(4):830–834. 10.1002/MRM.21376 [DOI] [PubMed] [Google Scholar]
  • 38.Williams A, Gillis A, McKenzie C, Po B, Sharma L, Micheli L et al. (2012) Glycosaminoglycan distribution in cartilage as determined by delayed gadolinium-enhanced MRI of cartilage (dGEMRIC). Potential Clin Appl 182(1):167–172. 10.2214/AJR.182.1.1820167 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Material

Data Availability Statement

The MRI data that support the findings of this study are openly available in Zeonodo at https://doi.org/10.5281/zenodo.7478225.