Abstract
Purpose
mapping is a powerful tool for studying osteoarthritis (OA) changes and bilateral imaging may be useful in investigating the role of between-knee asymmetry in OA onset and progression. The quantitative double-echo in steady-state (qDESS) can provide fast simultaneous bilateral knee and high-resolution morphometry for cartilage and meniscus. The qDESS uses an analytical signal model to compute relaxometry maps, which require knowledge of the flip angle (FA). In the presence of inhomogeneities, inconsistencies between the nominal and actual FA can affect the accuracy of measurements. We propose a pixel-wise correction method for qDESS mapping exploiting an auxiliary map to compute the actual FA used in the model.
Methods
The technique was validated in a phantom and in vivo with simultaneous bilateral knee imaging. measurements of femoral cartilage (FC) of both knees of six healthy participants were repeated longitudinally to investigate the association between variation and .
Results
The results showed that applying the correction mitigated variations that were driven by inhomogeneities. Specifically, left–right symmetry increased following the correction . Without the correction, values showed a linear dependence with . The linear coefficient decreased using the correction (from 24.3 ± 1.6 ms to 4.1 ± 1.8) and the correlation was not statistically significant after the application of the Bonferroni correction (p value > 0.01).
Conclusion
The study showed that correction could mitigate variations driven by the sensitivity of the qDESS mapping method to , therefore, increasing the sensitivity to detect real biological changes. The proposed method may improve the robustness of bilateral qDESS mapping, allowing for an accurate and more efficient evaluation of OA pathways and pathophysiology through longitudinal and cross-sectional studies.
Keywords: QMRI, QDESS, T2 mapping, B1 correction, Bilateral knee imaging
Introduction
Knee osteoarthritis (OA) remains a leading cause of disability in developed countries. Despite its prevalence and large socioeconomic burden, OA remains a poorly understood disease, with multiple pathways and risk factors leading to the onset of this debilitating disease. Qualitative Magnetic Resonance Imaging (MRI) can effectively assess OA-related morphological changes in the knee joint, such as cartilage loss, osteophytes, and bone marrow lesions. However, these structural changes are observed at a late stage of the disease, when treatments are unlikely to be effective. There is thus a goal to detect cartilage microstructure changes at an earlier time point when therapies are more likely to be effective. Quantitative MRI methodologies have shown promise to provide biomarkers sensitive to early OA changes that can help to study early OA onset and progression as well as to monitor response to treatments [1, 2].
The transverse relaxation time has been demonstrated to be sensitive to collagen microstructure and hydration changes in knee articular cartilage and has shown promise for evaluating early changes that are predictive of OA progression [1, 3]. Furthermore, for studying knee OA, bilateral examinations are commonly performed in order to evaluate OA-related changes [4], as well as to investigate the role of between-knee asymmetry in OA [5, 6]. For example, people sustaining a complete anterior crucial ligament (ACL) tear are more likely to develop post-traumatic OA in the injured knee. The contralateral uninjured knee is often used as an internal reference to evaluate longitudinal changes and study post-traumatic OA pathophysiology [7]. The 3D double-echo in steady-state (DESS) pulse sequence [8–10] has been widely applied in musculoskeletal imaging due to its image contrast and high SNR efficiency. DESS enables high-resolution imaging of cartilage, including assessment of cartilage thickness and defects [11–13]. Separation of the two echoes in the DESS sequence in a method known as quantitative DESS [14, 15] (qDESS) can further provide accurate and reproducible measurements [16, 17] by fitting their signals to an extended phase graph signal model [18–20]. In addition, the SNR efficiency and multiple contrasts provided by this approach have shown promise for bilateral clinical knee assessment in 5 min or less [5, 21–23]. However, as for other mapping methods, estimation accuracy is affected by field inhomogeneities [20, 24]. Specifically, qDESS mapping uses an analytical signal model to compute relaxometry maps [16], which requires knowledge of the flip angle (FA). The presence of inhomogeneities leads to inconsistencies between the prescribed and actual FA provided to the spin ensemble at each voxel, which affects the accuracy of measurements. can be a confounding factor both for cross-sectional and longitudinal studies since B1 inhomogeneities introduce bias and variance in mapping. A correction for qDESS mapping may be particularly relevant for bilateral knee acquisition, where inhomogeneities may be larger than single knee acquisitions because of excitation over a larger field of view (FOV) and knee positioning that is further removed from isocenter.
In this work, we aim to investigate to what extent inhomogeneities affect qDESS mapping accuracy and to what extent applying a correction is able to mitigate those effects. For correction, we propose a pixel-wise method for qDESS mapping that exploits an auxiliary map to compute the actual FA to be used in the analytical model for computing the estimate. We validated the correction technique with phantom and in vivo measurements. Subsequently, we investigated our primary aim with cross-sectional and longitudinal measurements of femoral cartilage (FC) in simultaneous bilateral knee imaging.
Theory
Quantitative DESS, represented in Fig. 1A, is a gradient-spoiled steady-state sequence that samples two echoes per repetition separated by a spoiler gradient. The two signals have different contrast mechanisms and are functions of the extrinsic scan parameters (, , flip angle , gradient amplitude G, and gradient duration ) as well as the intrinsic tissue parameters (, , and diffusivity).
Fig. 1.

A qDESS pulse sequence diagram with phase graph representation. B error as a function of ratio, defined as the ratio between the nominal and actual FA, obtained running Bloch simulations of the qDESS sequence for a voxel composed of cartilage ( and , ). Estimation of was performed using Eq. 1 considering a nominal FA = 20°
Sveinsson et al. [16] proposed a simple analytic model that can be used for estimation from the ratio between the second and first echoes of qDESS. The model is expressed in Eq. 1:
| (1) |
Here, and represent the repetition and the first echo time, respectively, is the flip angle, is the diffusivity, and the dephasing per unit length induced by the unbalanced gradient is denoted by where G and are the spoiler amplitude and duration, respectively, and is the gyromagnetic ratio. When applying Eq. 1 to femoral cartilage, and are usually fixed to 1200 ms and , respectively. This simple model, which allows estimating from qDESS, also allows one to appreciate the mechanism for which inhomogeneities can introduce errors in qDESS mapping. The FA must be known to correctly estimate from Eq. 1. inhomogeneities across the imaged volume directly affect the actual FA experienced by the spins within voxels introducing inaccuracies in estimation if not taken into account. If one defines the ratio as the ratio between the nominal and actual FAs, then an auxiliary map can be used to input, voxel-wise, the actual FA in Eq. 1 and correct for inhomogeneities.
Figure 1B reports Bloch simulations run to show the dependence of the estimation error as a function of the ratio between the nominal and actual FA, the ratio, experienced by the spin ensemble. Bloch simulations were run in MATLAB (version 20a, Mathworks, Natick, MA) using an in-house library with the following parameters:, , , nominal FA = 20°, and .
Methods
MRI acquisitions
All MRI acquisitions were performed on a GE 3T SIGNA Premier scanner (GE Healthcare, Milwaukee, WI, USA) using 16-channel flexible phased-array receive-only coils (NeoCoil, Pewaukee, WI, USA). Informed consent was obtained from all subjects in accordance with the Institutional Review Board protocol at our institution. Simultaneous bilateral scans used two identical coils wrapped around each knee.
The MRI protocol consisted of a 3D qDESS sequence for mapping with a 20-degree FA. Data for generating a map, performed using a double-angle method [25] (DAM), was acquired with magnetization prepared (MP-DAM) sequence [26–28] with two flip angles (30° and 60° flip angle excitation pulses followed by a crusher gradient) followed by elliptic centric 3D SPGR Readout. A diagram of the pulse sequences is reported in Fig. S1 of Supplementary Information. The 3D qDESS used a spectrally selective excitation with nominal flip angle and a spoiler gradient, which had a spoiler moment of 3.13 mT/m ms with duration . Other sequence parameters varied depending on the type of acquisition (phantom vs in vivo, unilateral vs bilateral) and are summarized in Table 1.
Table 1.
Summary of the sequence parameters
| Sequence | Phantom |
In vivo unilateral |
In vivo bilateral |
|||
|---|---|---|---|---|---|---|
| parameter | qDESS | MP-DAM | qDESS | MP-DAM | qDESS | MP-DAM |
| Scan plane | Sagittal | Sagittal | Sagittal | Sagittal | Sagittal | Sagittal |
| Echo times (ms) | 5/25 | 1.3 | 5/25 | 2 | 5/25 | 2 |
| TR (ms) | 15 | 3.67 | 15 | 4.5 | 15 | 4.5 |
| FOV (cm) | 16 × 16 × 12 | 16 × 16 × 12 | 16 × 16 × 12 | 16 × 16 × 12 | 16 × 16 × 36 | 16 × 16 × 36 |
| Acquisition matrix | 128 × 128 | 128 × 128 | 320 × 320 | 256 × 256 | 320 × 320 | 256 × 256 |
| Slice thickness (mm) | 2 | 2 | 1.4 | 4 | 1.4 | 4 |
| Number of slices | 64 | 64 | 120 | 42 | 240 | 84 |
| Acquisition time (m:s) | 0:50 | 0:40 | 1:03 | 0:54 | 2:40 | 1:27 |
The parameters for MP-DAM are for the readout of the longitudinally stored magnetization prepared signal with 30°/60° flip angles
Phantom measurements
A phantom made of three pairs of vials containing 2%, 3%, and 4% agarose, with ranging from 30 to 70 ms, was scanned. maps computed from qDESS with and without using the corrections were compared to validate the correction method.
The MRI phantom protocol was repeated with modifications to the transmit gain (TG) of the pulse transmitter. A variation in the TG modifies the experienced by the spins in each voxel proportionally to the extent of the variation. The range of TG values was selected so that the smallest and largest TG values corresponded to a 20% reduction and increase with respect to the default TG value, respectively. qDESS maps were computed with and without applying the correction, and within each vial of the phantom the linear correlation between and was assessed.
In vivo knee imaging
The knee of a healthy volunteer was repeatedly scanned unilaterally, varying the TG. The range of TG values was selected so that the smallest and largest TG values corresponded to a 20% reduction and increase with respect to the default TG value, respectively. The linear dependence of on values was assessed using a linear fit, which was performed with and without applying the correction.
Six healthy volunteers (4 females and 2 males) were scanned using a simultaneous bilateral knee acquisition [21], once a week for 8 weeks, to evaluate the effect of the correction on longitudinal stability and between-knee asymmetry. During the second week, each subject was scanned once a day for 5 days.
Data processing and analysis
For in vivo longitudinal bilateral knee acquisitions, with reference to Fig. 2, for each subject within each timepoint, the following data processing protocol was applied.
Fig. 2.

Pipeline used to process the longitudinal in vivo bilateral acquisitions. For each subject, within each time point map was first obtained from the MP-DAM images and smoothed with a 2D median filter. The FC was automatically segmented from the qDESS images and subdivided into 6 sub-regions: anterior/central/posterior for the medial/lateral sides. The map was then co-registered to qDESS and maps were then obtained with and without correction and average quantitative values computed within each FC sub-region. Finally, for visualization, the 3D segmented volume was projected onto a 2D space
The map was computed from the MP-DAM acquisition using an in-house script developed in MATLAB (version 20a, Mathworks, Natick, MA), which can be found in the public repository that contains the data of this work (see data availability statement). A customized 2D median filter with intensity thresholding was applied to smooth the map. Thus, within the kernel, only pixels for which their intensity was above a set threshold value were considered in the computation of the median. This aimed to avoid bias in the median due to pixels of very low SNR, such as those in the bone/cartilage interface (see arrow in Fig. 2 panel 1). Since the DAM is magnitude based, in such region the estimation of is inaccurate. Specifically, at each step of the convolution, the median was evaluated on the map, but the decision of which pixels to include in the computation was performed on the raw MP-DAM images belonging to the first flip angle acquisition. The threshold was automatically adjusted for each acquisition, as it was set to the 75-th percentile of the image intensity values. The selection of the percentile threshold was done empirically and aimed to assure the exclusion of the pixels belonging to the cartilage-bone region.
The femoral cartilage (FC) was automatically segmented from the qDESS acquisition using the open-source Deep Open-Source Medical Image Analysis (DOSMA) framework [29, 30](https://github.com/ad12/DOSMA), and further subdivided into 6 sub-regions: anterior/central/posterior for the medial/lateral sides.
The map was then registered to the first echo of qDESS through a rigid transformation using Elastix [31, 32], so that and maps were co-registered for each acquisition.
maps were computed on a pixel-by-pixel basis by analytically inverting the qDESS signal model [16] expressed in Eq. 1, both using the nominal FA and the actual FA computed by multiplying the nominal FA with the ratio estimated from the auxiliary double-angle acquisition. In performing the computation, and were fixed to 1200 ms and , respectively.
The average values, with and without correction, and values within each FC sub-region were then computed and used to perform statistical analysis.
For visualization purpose only, the 3D maps were projected onto a 2D space (Fig. 2 panel 6). To visualize the between-knee asymmetry. Pixel-wise subtraction was evaluated by mirroring and registering the right 2D projected FC to the left 2D FC projection for direct comparison.
For all the phantom and unilateral in vivo knee measurements, the procedure used to obtain and maps was the same as described above with the only exception that the customized 2D smoothing filter was not applied for the phantom experiments. For the phantom, the for computation was fixed to 1600 ms, which had been previously estimated using a variable flip angle mapping method.
Statistical analysis
For the phantoms and in vivo knee experiments, the dependence of values, obtained with and without applying the correction, on was assessed by fitting a linear model to the experimental data. The linear coefficient of the model, a, also referred to as slope, was used to assess the strength of the linear dependence between and , and the fit p-value was used to test the statistical significance of the linear correlation.
For the in vivo longitudinal bilateral acquisitions, a repeated measurement Analysis of Variance (RMANOVA) was used to determine the within-subject standard deviation (SD), also called measurement error SD, of three parameters across time points and in each FC sub-region. The parameters were: without correction, with correction and . The coefficient of repeatability (RC) and the coefficient of variation (CV) were also computed to characterize longitudinal stability. The rationale for considering each FC sub-region separately was to eliminate variance related to differences in values across FC sub-regions. To assess the dependence of on , the following multivariate linear regression model was used:
| (2) |
The statistical significance of each coefficient was tested with a significance level of , and the Bonferroni correction was used to set the significance threshold for the p value. Since 5 predictors were considered, the significant threshold was set to 0.05/5, which is equal to 0.01. The model was fitted to values obtained with and without correction.
Finally, Bland–Altman (BA) analysis was used to assess left–right asymmetry with and without using the correction. Lin’s concordance correlation coefficient and the coefficient of variation of the root mean squared error (CVRMSE) were also evaluated as global similarity metrics.
All the statistical analyses were run using in-house scripts developed in Python.
Results
Phantom validation
The Phantom maps, obtained with and without applying the correction, are shown in Fig. 3, along with the map and the comparison among distributions within each pair of vials (Fig. 3B). values are dimensionless as they are expressed as the ratio between the actual and nominal FA. The map showed a variation in the across the phantom within the range [0.8, 1.1], which corresponds roughly to ± 20% FA variations with respect to the nominal value. The spatial variation of was such that for each pair of vials, one of the two vials experienced a lower than the other. The map computed without applying the correction showed that in each pair of identical vials, an asymmetry in the values between the two vials was observed. This can be better appreciated by looking at the distributions (first row of Fig. 3B). When the map is computed by applying the proposed correction, the differences between paired vials in the phantom were greatly reduced. This can be seen in the appearance of the 2D maps and by observing the overlapping histograms of the distributions (second row of Fig. 3B).
Fig. 3.

A Sketch of the phantom (left) along with its map, and map (bottom right) computed from qDESS both with and without correcting for B1 (top right) that consisted of roughly ± 20% FA variations. values are dimensionless, being the ratio between the actual and nominal flip angles. The phantom was designed to have 3 pairs of vials with different relaxation times. B Comparisons between the distributions of the paired samples without (top row) and with (bottom row) corrections demonstrate that using correction reduces errors in the paired calculations
The correlation plots between and values for each vial of the phantom, obtained with various TGs, are shown in Fig. 4. The linear coefficient and the associated fit p-value are reported for each correlation plot. When was computed without correcting for , the linear coefficients were greater than or equal to 21 ms (21–50 ms), with p values (p) less than or equal to 0.001. When the correction was applied, the slopes of the linear fit were reduced, and in almost all vials, the correlation was not statistically significant (p > 0.05). For vials 2 and 6, the correlations were still statistically significant after applying the correction (p < 0.05), but the slopes of the linear fit were substantially reduced compared to the case in which the correction was not applied: the slopes decreased from 41 ms to 8.5 ms and from 50 ms to 8.4 ms for vials 2 and 6, respectively.
Fig. 4.

Correlation plots between and for each phantom vial acquired varying the TG. The first column reports the correlation plots obtained using values computed without applying the correction, whereas in the second column values are computed using the correction. Points represent the average parameter values within the vial, while error bars represent the standard deviation of parameter values within the vial
Validation in vivo
The correlation plots between and values in FC sub-regions obtained with varying TG weights are shown in Fig. 5. For each FC sub-region, the correlation plot obtained using values computed without applying the correction is drawn in blue, whereas red is used for values computed using the correction. The points represent the average parameter values within the sub-region, while the error bars represent the standard deviation. The linear coefficient and the fit p-value are reported for each correlation plot.
Fig. 5.

Correlation plots between and in each FC sub-regions acquired varying the TG. For each FC sub-region, the correlation plot obtained using values computed without applying the correction is drawn in blue, whereas red is used for values computed using the correction. Points represent the average parameter values within the sub-region, while error bars represent the standard deviation of parameter values within the sub-region
Measured values varied from 0.75 to 1.3. Without applying the correction, for all sub-regions, a statistically significant linear correlation between and values was observed (p ≤ 0.002) and a slope greater than or equal to 18 ms. When the correction was applied, the linear coefficients were greatly reduced for all the sub-regions. Furthermore, the correlations were not statistically significant in all the FC sub-regions (p > 0.05), with the only exception of the lateral posterior region (p = 0.01).
In vivo bilateral knee acquisitions
Longitudinal variation
Subjects #5 and 6 did not complete the study and only data from the first 6 weeks were available for these subjects. Furthermore, for subject #3, the MP-DAM data did not correctly reconstruct at the scanner for the first time point, which was then discarded from the analysis. Hence, the RMANOVA test was run over 8 total time points (from week #2 to week #6) for which data from all subjects were available.
For each parameter, the RMANOVA test was unable to reject the null hypothesis with a significance value of 0.05. The summary statistics of the repeated measurement analysis are reported for the left and right knee in Table 2. The value of each statistic is reported as the averaged value across different sub-regions, and the standard deviation across sub-regions is also reported. Overall, was relatively stable during time points, with a CV of 4.1% both in the left and right knee. Without the correction, the average within-subject SD was 1.8 ms and 1.3 ms for the left and right knees, respectively, and the corresponding RCs were 5.1 ms and 3.6 ms. The CV was 6% in both knees. When the correction was applied, the average within-subject SD was 1.9 ms and 1.5 ms for the left and right knees, respectively. The RC was 5.3 ms and 4.1 ms in the left and right knee, respectively. The CV was 6% in both knees.
Table 2.
Summary statics for the longitudinal in vivo simultaneous bilateral knee acquisitions on six healthy volunteers
| Statistic (avg ± sd) | Left knee |
Right knee |
||||
|---|---|---|---|---|---|---|
|
|
|
|||||
| W/O Corr | W/Corr | W/O Corr | W/Corr | |||
| Within-subject SD | (1.8 ± 0.4) ms | (1.9 ± 0.5) ms | 0.04 ± 0.01 | (1.3 ± 0.3) ms | (1.5 ± 0.1) ms | 0.03 ± 0.01 |
| RC | (5.1 ± 2.5) ms | (5.3 ± 3.1) ms | 0.1 ± 0.1 | (3.6 ± 0.8) ms | (4.1 ± 0.4) ms | 0.1 ± 0.1 |
| CV (%) | 6.0 ± 1.1 | 6.0 ± 1.2 | 4.1 ± 0.6 | 4.4 ± 0.6 | 4.6 ± 0.4 | 4.1 ± 0.7 |
The value of each statistic is reported as the mean value averaged across the six FC sub-regions, and the standard deviation across FC sub-regions is also reported
Dependence of on
The linear coefficients for the multivariate linear regression model fitted to the values obtained with and without correction are reported in Table 3. The p value, R-squared , and its relative importance on the total model are reported for each linear coefficient. Without applying the correction, the linear regression model explained 26% of the variance (total ). Out of the 5 explanatory variables, had the most relative importance, explaining 64.4% of the total . The associated linear coefficient was equal to (24.3 ± 1.6) ms with a p value of , showing a positive linear correlation between and . Of the other explanatory variables, the FC Sub-region and Knee had the highest relative importance: 28.8% and 5.3%, respectively. The remaining variables were progressively negligible, and the linear coefficient associated with the variable Time was not statically significant (p value > 0.01).
Table 3.
Summary of the multivariate linear regression model expressed in Eq. 2 run on data with and without applying the correction
| Variable | p value | rel. imp. (%) | ||
|---|---|---|---|---|
| Without correction | ||||
| Intercept | 9.7 ± 1.5 | 1.3×10−10 | – | – |
| Subject | 0.20 ± 0.07 | 4×10−3 | 0.003 | 1.2 |
| Knee | 1.1 ± 0.3 | 6×10−5 | 0.014 | 5.3 |
| Sub-region | −0.78 ± 0.07 | 5×10−28 | 0.075 | 28.8 |
| Time | −0.02 ± 0.03 | 0.5 | 0.0003 | 0.1 |
| 24.3 ± 1.6 | 7×10−47 | 0.17 | 64.6 | |
| Total: | 0.26 | 100 | ||
| With correction | ||||
| Intercept | 30.2 ± 1.7 | 4×10−62 | – | – |
| Subject | 0.26 ± 0.08 | 6×10−4 | 0.01 | 6.6 |
| Knee | 1.03 ± 0.29 | 4×10−4 | 0.008 | 5.2 |
| Sub-region | −0.90 ± 0.08 | 3×10−29 | 0.13 | 84.3 |
| Time | −0.04 ± 0.04 | 0.3 | 0.001 | 0.9 |
| 4.1 ± 1.8 | 0.02 | 0.005 | 2.9 | |
| Total | 0.15 | 100 | ||
For each variable of the model, the linear coefficient is reported along with its p value, , and the relative importance of such on the total of the model. Following the Bonferroni correction, the threshold for statistical significance was set to 0.01
When the correction was applied, the total explanatory power of the regression model was reduced to 15% (total ). Specifically, the linear coefficient associated with was reduced to (4.1 ± 1.8) ms, explaining only the 2.9% of the total and it was not statically significant (p value > 0.01). The most important explanatory variable was the FC Sub-region, explaining 84.3% of the total , with a linear coefficient equal to −0.9 ± 0.08. The remaining variables were progressively negligible, and the linear coefficient associated with the variable Time was not statically significant (p value > 0.01).
left–right knee asymmetry and correlation between and
An example of left–right asymmetry for one time point of subject #5 is displayed in Fig. 6. Each row displays: the parameter of interest ( and in the top and bottom panels, respectively); the 2D projections of FC for the left and the mirrored right knee; the pixel-wise difference; and the BA plots between the left and right parameter average values in the six FC sub-regions. In Fig. 6, differed by up to 0.3 between regions in the left and right knees, as highlighted by the 2D projections. Left–right symmetry (bottom panel of Fig. 6) increased following a correction, evident through visual inspection of the unrolled maps and the BA analysis. In particular, when the correction was not applied, the mean bias, CV, and were equal to −4.3 ms, 12.3 %, and 0.31, respectively. When the correction was applied, the mean bias, CV, and were equal to 0.5 ms, 7.0 %, and 0.82, respectively.
Fig. 6.

Example of comparison between the left and right maps of FC where differs substantially between the left and right knee. Each row displays the quantitative 2D projected and unrolled parameter ( and in the top and bottom panels, respectively) for the left and the mirrored right knee. The pixel-wise difference and the BA plots between left and right parameter values in the six FC sub-regions are also reported
The BA plots regarding left–right asymmetry for and , computed with and without correction, considering all the subjects and time points are reported in Fig. 7. On average ratio differed by about 0.1 between the left and right knees, with LOA equal to ± 0.17 and an RMSECV of 7%. Poor agreement between left and right values was observed as summarized by Lin’s concordance coefficient being equal to 0.19. Regarding , the left–right symmetry improved when the was applied compared to the case in which the correction was not applied. Specifically, the RMSECV was reduced from 5 to 4.4% and Lin’s concordance coefficient was increased from 0.69 to 0.74. The limits of agreement were ± 5.2 ms when correction was not applied, and ± 5.4 ms when correction was applied.
Fig. 7.

BA plots between left and right values for (right), without correction (center), and with correction. All subjects and time points are reported in the plot, and each color represents one of the FC sub-regions. For each plot, the limits of agreement (LOA), the RMSECV and Lin’s concordance coefficient are reported
Discussion
qDESS is becoming a widely utilized tool for studying knee OA due to its capability of providing accurate relaxation time measurements and SNR-efficient morphological bilateral imaging in 5 min or less. As for many other mapping methods, qDESS mapping is sensitive to inhomogeneities, which becomes increasingly important to account for when performing simultaneous bilateral mapping. Therefore, (i) we proposed a pixel-wise correction method for qDESS mapping that exploits an auxiliary map, and (ii) investigated to what extent the correction impacts qDESS mapping in simultaneous bilateral knee imaging.
Our experiments with the phantom demonstrated the errors in calculations that occur with variations. Without correction, qDESS relaxation times for sample pairs varied over the field of view, with a mean variation in the order of 10 ms. This variation was removed when the correction was applied. Similarly, phantom and in vivo experiments run with varying TG showed that the proposed correction was able to considerably mitigate the dependence of qDESS on . Furthermore, these experiments demonstrated that the proposed method was able to correct measurements for a wide range of values (0.68–1.34).
The results regarding the in vivo longitudinal bilateral knee acquisitions gave several insights on the impact of on qDESS mapping in vivo. First, Regarding longitudinal repeatability, there was a negligible difference between the case in which the correction was applied compared to the case in which correction was not applied both for the right and left knee. In both cases, the average CVs were considered acceptable considering that they were within the repeatability CVs reported in the literature for qDESS mapping [33]. The longitudinal study was carried out using the same scanner, and each subject was acquired under similar conditions. As a consequence, within each subject, was relatively stable across time as highlighted by the high repeatability metrics reported in Table 2. Although not surprising, these results suggest that correction might be less concerning when dealing with longitudinal studies where the measurement conditions do not change across time points. Nonetheless, the multivariate linear regression analysis showed that the proposed correction method greatly mitigated the dependence of on . The interpretation of this result is that even if did not vary significantly across timepoints, it appeared to vary across subjects. Thus, correction may be more important for cross-sectional studies to reduce a systematic source of bias and variability in measurements. Furthermore, the results suggested that asymmetries in relaxation times between the left and right knees of asymptomatic volunteers were partially related to inhomogeneities. It is important to note that there is no clear understanding of whether or not values in left and right knees should be symmetrical in healthy volunteers. There is a lack of studies in the literature addressing this scientific question. The main finding of our study was that left–right asymmetries in appeared to contribute to left–right asymmetries in . For example, when considering the results reported in Fig. 6, values were higher in the left knee compared with the right knee. As a consequence, when the correction was not applied values in the left knee were also higher than values in the right knee. Applying the correction removed this bias due to , and left–right asymmetry was reduced. The key point is that our work clearly suggests that correction mitigated variations driven by the sensitivity of the mapping method to and did not reflect actual biological differences between knees. This finding is particularly relevant for bilateral scanning and applications at ultra-high fields, but can also help ensure accuracy for unilateral longitudinal scans. Our study showed that errors due to inhomogeneities are in the order of 5 ms or more with a 20% variation in , as also demonstrated by Bloch simulations in Fig. 1B. These errors are relevant as studies investigating the longitudinal variation of in cartilage in knee OA over a period of two years have found variations in the order of 1.5–2 ms [34, 35]. Similarly, longitudinal studies of in knee cartilage after ACL reconstruction over a period of 24 months have observed average differences of about 5 ms between at 6 months after surgery and at 24 months [36]. Thus, taking into account inhomogeneities in qDESS mapping might be beneficial for obtaining meaningful results from cross-sectional and longitudinal studies, as well as studies investigating the use of the contralateral knee as a reference.
Although only healthy subjects were considered in this study, the obtained results on the effect of , and correction, in qDESS mapping are independent of the underlying knee condition. In fact, preliminary investigations on the application of the same protocol to two subjects that had undergone ACL surgery in one of their knees showed that multivariate linear regression analysis gave results that were consistent with those obtained in healthy volunteers (see Table S1 of supplementary material). Overall, our experiments suggest that implementing a correction should be considered when designing studies investigating with qDESS. Future studies can further investigate the impact of on qDESS mapping when performing multi-center and multi-vendor studies since inhomogeneities might show a higher variance than that observed in our study. Furthermore, this work may provide a protocol to effectively investigate the scientific value of using the contralateral (healthy) knee as an intrinsic reference for studying knee pathologies using bilateral qDESS mapping.
There are some limitations to our methodology. The proposed correction requires an additional acquisition to obtain the auxiliary map. However, mapping can be performed relatively quickly, requiring just about a minute and a half in our bilateral studies. When used in tandem with qDESS, the SNR efficiency of qDESS still allows for reduced scan times compared to non-model based mapping methods, such as MESE which normally require more than 5 min per knee and would still require a correction. The qDESS time efficiency is particularly relevant with simultaneous bilateral acquisitions. Further, additional post-processing is needed to extract corrected relaxation times. However, the additional accuracy in estimation may be critical to obtain meaningful information from cross-sectional and longitudinal studies. It is worth noting that in estimating through Eq. 1, the values of and were assumed to be constant across the cartilage tissue and subjects. This in principle could introduce biases whenever this condition is not true. In practice, of cartilage is usually considered to be constant across cartilage also in other techniques such as delayed Gadolinium-Enhanced MRI [37, 38]. Furthermore, simulations have shown that for the typical value assumed for cartilage (1200 ms), if the actual is 20% lower or higher than the assumed value, the error in estimation is only 4% [16]. Thus, the method has low sensitivity to variations. The sensitivity of to diffusion is a function of the diffusion coefficient and the spoiling gradient area. In our case, the gradient area was set to the minimum value to achieve full spoiling. Thus, the sequence has minimal sensitivity to diffusion [16] and it is not thought to introduce significant errors in the estimation of . However, further studies could study whether in presence of pathological changes the diffusion coefficient of cartilage increases to the point of introducing biases in the estimation of with qDESS. Lastly, it is worth mentioning that in our study, we used receive-only coils, and the transmission was performed with the body coil. Thus, when using transmit–receive coils, the size effect of variations in might be different. Nonetheless, the outlined methodology will work independently from the setup utilized, provided that an accurate map can be acquired.
Conclusions
A correction for qDESS mapping was proposed to improve the accuracy of the method. The correction exploited an auxiliary map to compute the actual FA to be input into the model used to estimate . Validation of the methodology was performed in phantom and in vivo. Further, in vivo longitudinal bilateral knee acquisitions demonstrated that the correction mitigated variations that were driven by inhomogeneities rather than reflecting biological differences. The proposed method may improve the robustness of bilateral qDESS mapping, allowing more accurate and efficient evaluation of OA pathophysiology through longitudinal and cross-sectional studies.
Supplementary Material
Acknowledgements
This work was supported by GE Healthcare and NIH Grants R01-AR077604, R01-EB002524, R01-AR074492, R21EB030180, K24-AR062068, and R00-EB022634.
Footnotes
Supplementary Information The online version contains supplementary material available at https://doi.org/10.1007/s10334-023-01094-y.
Declarations
Conflict of interest The authors declare no conflicts of interest.
Ethical approval This study received Institutional Review Board approval.
Informed consent Written informed consent was obtained from the volunteer.
Data availability
The MRI data that support the findings of this study are openly available in Zeonodo at https://doi.org/10.5281/zenodo.7478225.
References
- 1.Liu F, Chaudhary R, Hurley SA, Munoz Del Rio A, Alexander AL, Samsonov A et al. (2014) Rapid multicomponent T2 analysis of the articular cartilage of the human knee joint at 3.0T. J Magn Reson Imaging 39(5):1191–1197. 10.1002/jmri.24290 [DOI] [PubMed] [Google Scholar]
- 2.Ling W, Regatte RR, Navon G, Jerschow A (2008) Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST). Proc Natl Acad Sci 105(7):2266–2270. 10.1073/PNAS.0707666105 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Monu UD, Jordan CD, Samuelson BL, Hargreaves BA, Gold GE, McWalter EJ (2017) Cluster analysis of quantitative MRI T2 and T1ρ relaxation times of cartilage identifies differences between healthy and ACL-injured individuals at 3T. Osteoarthr Cartil 25(4):513–520. 10.1016/j.joca.2016.09.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Jungmann PM, Brucker PU, Baum T, Link TM, Foerschner F, Minzlaff P et al. (2015) Bilateral cartilage T2 mapping 9 years after Mega-OATS implantation at the knee: a quantitative 3T MRI study. Osteoarthr Cartil 23(12):2119–2128. 10.1016/j.joca.2015.06.013 [DOI] [PubMed] [Google Scholar]
- 5.Barbieri M, Desai AD, Kogan F, Mazzoli V, Rubin E, Castellani G et al. (2019) Rapid Quantitative Bilateral knee imaging with fully automated femoral cartilage analysis In: Proc. of ISMRM 27th Annual Meeting and Exhibition. Montreal, Canada: Proc. of ISMRM 27th Annual Meeting and Exhibition; 11–16 [Google Scholar]
- 6.Black M, Younk K, Chaudhari A, Kogan F, Sveinsson B, McWalter E et al. (2020) T2 cluster asymmetry can detect differences in cartilage of ACL-injured subjects 3 months post-surgery In: Proc. of 2020 International Workshop on Osteoarthritis Imaging. Salzburg, Austria: Proc. of 2020 International Workshop on Osteoarthritis Imaging [Google Scholar]
- 7.Casula V, Tajik BE, Kvist J, Frobell R, Haapea M, Nieminen MT et al. (2022) Quantitative evaluation of the tibiofemoral joint cartilage by T2 mapping in patients with acute anterior cruciate ligament injury vs contralateral knees: results from the subacute phase using data from the NACOX study cohort. Osteoarthr Cartil 30(7):987–997. 10.1016/j.joca.2022.02.623 [DOI] [PubMed] [Google Scholar]
- 8.Bruder H, Fischer H, Graumann R, Deimling M (1988) A new steady-state imaging sequence for simultaneous acquisition of two MR images with clearly different contrasts. Magn Reson Med 7(1):35–42. 10.1002/MRM.1910070105 [DOI] [PubMed] [Google Scholar]
- 9.Redpath TW, Jones RA (1988) FADE-A new fast imaging sequence. Magn Reson Med 6(2):224–234. 10.1002/mrm.1910060211 [DOI] [PubMed] [Google Scholar]
- 10.Lee SY, Cho ZH (1988) Fast SSFP gradient echo sequence for simultaneous acquisitions of FID and echo signals. Magn Reson Med 8(2):142–150. 10.1002/mrm.1910080204 [DOI] [PubMed] [Google Scholar]
- 11.Eckstein F, Westhoff J, Sittek H, Maag KP, Haubner M, Faber S et al. (2013) In vivo reproducibility of three-dimensional cartilage volume and thickness measurements with MR imaging. Am J Roentgenol 170(3):593–597. 10.2214/AJR.170.3.9490936 [DOI] [PubMed] [Google Scholar]
- 12.Moriya S, Miki Y, Yokobayashi T, Ishikawa M (2009) Three-dimensional double-echo steady-state (3D-DESS) magnetic resonance imaging of the knee: contrast optimization by adjusting flip angle. Acta Radiologica 50(5):507–511. 10.1080/02841850902849444 [DOI] [PubMed] [Google Scholar]
- 13.Hardy PA, Recht MP, Piraino D, Thomasson D (1996) Optimization of a dual echo in the steady state (DESS) free-precession sequence for imaging cartilage. J Magn Reson Imaging 6(2):329–335. 10.1002/JMRI.1880060212 [DOI] [PubMed] [Google Scholar]
- 14.Welsch GH, Scheffler K, Mamisch TC, Hughes T, Millington S, Deimling M et al. (2009) Rapid estimation of cartilage T2 based on double echo at steady state (DESS) with 3 Tesla. Magn Reson Med 62(2):544–549. 10.1002/MRM.22036 [DOI] [PubMed] [Google Scholar]
- 15.Staroswiecki E, Granlund KL, Alley MT, Gold GE, Hargreaves BA (2012) Simultaneous estimation of T2 and apparent diffusion coefficient in human articular cartilage in vivo with a modified three-dimensional double echo steady state (DESS) sequence at 3 T. Magn Reson Med 67(4):1086–1096. 10.1002/MRM.23090 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Sveinsson B, Chaudhari AS, Gold GE, Hargreaves BA (2017) A simple analytic method for estimating T2 in the knee from DESS. Magn Reson Imaging 38:63–70. 10.1016/j.mri.2016.12.018 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Chaudhari AS, Black MS, Eijgenraam S, Wirth W, Maschek S, Sveinsson B et al. (2018) Five-minute knee MRI for simultaneous morphometry and T2 relaxometry of cartilage and meniscus and for semiquantitative radiological assessment using double-echo in steady-state at 3T. J Magn Reson Imaging 47(5):1328–1341. 10.1002/JMRI.25883 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Hennig J (1991) Echoes-how to generate, recognize, use or avoid them in MR-imaging sequences. Part I: fundamental and not so fundamental properties of spin echoes. Concepts Magn Reson 3(3):125–143. 10.1002/cmr.1820030302 [DOI] [Google Scholar]
- 19.Hennig J (1991) Echoes-how to generate, recognize, use or avoid them in MR-imaging sequences. Part II: Echoes in imaging sequences. Concepts Magn Reson 3(4):179–192. 10.1002/cmr.1820030402 [DOI] [Google Scholar]
- 20.Weigel M (2015) Extended phase graphs: dephasing, RF pulses, and echoes—pure and simple. J Magn Reson Imaging 41(2):266–295. 10.1002/jmri.24619 [DOI] [PubMed] [Google Scholar]
- 21.Kogan F, Levine E, Chaudhari AS, Monu UD, Epperson K, Oei EHG et al. (2018) Simultaneous bilateral-knee MR imaging. Magn Reson Med 80(2):529–537. 10.1002/MRM.27045 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Chaudhari AS, Stevens KJ, Sveinsson B, Wood JP, Beaulieu CF, Oei EHG et al. (2019) Combined 5-minute double-echo in steady-state with separated echoes and 2-minute proton-density-weighted 2D FSE sequence for comprehensive whole-joint knee MRI assessment. J Magn Reson Imaging 49(7):e183–e194. 10.1002/jmri.26582 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Chaudhari AS, Grissom MJ, Fang Z, Sveinsson B, Lee JH, Gold GE et al. (2021) Diagnostic accuracy of quantitative multicontrast 5-minute knee MRI using prospective artificial intelligence image quality enhancement. Am J Roentgenol 216(6):1614–1625. 10.2214/AJR.20.24172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Prasloski T, Mädler B, Xiang QS, MacKay A, Jones C (2012) Applications of stimulated echo correction to multicomponent T2 analysis. Magn Reson Med 67(6):1803–1814. 10.1002/MRM.23157 [DOI] [PubMed] [Google Scholar]
- 25.Stollberger R, Wach P (1996) Imaging of the active B1 field in vivo. Magn Reson Med 35(2):246–251. 10.1002/MRM.1910350217 [DOI] [PubMed] [Google Scholar]
- 26.Cunningham CH, Pauly JM, Nayak KS (2006) Saturated double-angle method for rapid B1+ mapping. Magn Reson Med 55(6):1326–1333. 10.1002/MRM.20896 [DOI] [PubMed] [Google Scholar]
- 27.Chung S, Kim D, Breton E, Axel L (2010) Rapid B1+ mapping using a pre-conditioning RF pulse with TurboFLASH readout. Magn Reson Med 64(2):439. 10.1002/MRM.22423 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Watkins LE, Rubin EB, Mazzoli V, Uhlrich SD, Desai AD, Black M et al. (2020) Rapid volumetric gagCEST imaging of knee articular cartilage at 3 T: evaluation of improved dynamic range and an osteoarthritic population. NMR Biomed 33(8):e4310. 10.1002/nbm.4310 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Desai AD, Barbieri M, Mazzoli V, Rubin E, Black M, Watkins L et al. (2019) DOSMA: a deep-learning, open-source framework for musculoskeletal MRI analysis In: Proc. of ISMRM 27th Annual Meeting and Exhibition. Montreal, Canada: Proc. of ISMRM 27th Annual Meeting and Exhibition; 11–16 [Google Scholar]
- 30.Desai AD, Gold GE, Hargreaves BA, Chaudhari AS (2019. feb) Technical considerations for semantic segmentation in MRI using Convolutional Neural Networks. arXiv:1902.01977v1 [eessIV]. arXiv:1902.01977 [Google Scholar]
- 31.Klein S, Staring M, Murphy K, Viergever MA, Pluim JPW (2010) Elastix: a toolbox for intensity-based medical image registration. IEEE Trans Med Imaging 29(1):196–205. 10.1109/TMI.2009.2035616 [DOI] [PubMed] [Google Scholar]
- 32.Shamonin DP, Bron EE, Lelieveldt BPF, Smits M, Klein S, Staring M (2014) Fast parallel image registration on CPU and GPU for diagnostic classification of Alzheimer’s disease. Front Neuro-inform 0(JAN):50. 10.3389/FNINF.2013.00050 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Jordan CD, McWalter EJ, Monu UD, Watkins RD, Chen W, Bangerter NK et al. (2014) Variability of CubeQuant T1ρ, quantitative DESS T2, and cones sodium MRI in knee cartilage. Osteoarthr Cartil 22(10):1559–1567. 10.1016/j.joca.2014.06.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Pan J, Pialat JB, Joseph T, Kuo D, Joseph GB, Nevitt MC et al. (2011) Knee cartilage T2 characteristics and evolution in relation to morphologic abnormalities detected at 3-T MR imaging: a longitudinal study of the normal control cohort from the osteoarthritis initiative. Radiology 261(2):507–515. 10.1148/radiol.11102234 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Kretzschmar M, Heilmeier U, Yu A, Joseph GB, Liu F, Solka M et al. (2016) Longitudinal analysis of cartilage T2 relaxation times and joint degeneration in African American and Caucasian American women over an observation period of 6 years- Data from the Osteoarthritis Initiative. Osteoarthritis Cartilage 24(8):1384. 10.1016/J.JOCA.2016.03.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Lansdown DA, Xiao W, Zhang AL, Allen CR, Feeley BT, Li X et al. (2020) Quantitative imaging of anterior cruciate ligament (ACL) graft demonstrates longitudinal compositional changes and relationships with clinical outcomes at 2 years after ACL reconstruction. J Orthop Res 38(6):1289–1295. 10.1002/JOR.24572 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Williams A, Mikulis B, Krishnan N, Gray M, McKenzie C, Burstein D (2007) Suitability of T1Gd as the “dGEMRIC index” at 1.5T and 3.0T. Magn Reson Med 58(4):830–834. 10.1002/MRM.21376 [DOI] [PubMed] [Google Scholar]
- 38.Williams A, Gillis A, McKenzie C, Po B, Sharma L, Micheli L et al. (2012) Glycosaminoglycan distribution in cartilage as determined by delayed gadolinium-enhanced MRI of cartilage (dGEMRIC). Potential Clin Appl 182(1):167–172. 10.2214/AJR.182.1.1820167 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The MRI data that support the findings of this study are openly available in Zeonodo at https://doi.org/10.5281/zenodo.7478225.