Table 3.

Phantom ADC bias, repeatability metrics, and SNR for each DWI sequence. ADC bias was calculated using the central phantom vial. The ADC bias and within-ROI %CV are expressed as the mean ± standard deviation of the values measured in each of the four replicate images. Tolerance values from the QIBA Profile are included in the last column for comparison.

EPI (MR-linac) SPLICE (MR-linac) TSE (MR-linac) EPI (MR sim) BLADE (MR sim) RESOLVE (MR sim) QIBA tolerance values
ADC bias (10−3 mm2/s) 0.026 ± 0.004 −0.057 ± 0.003 −0.123 ± 0.002 −0.015 ± 0.002 −0.001 ± 0.010 −0.024 ± 0.010 ≤ ±0.040
%CV (within-ROI) 2.55 ± 0.71 3.14 ± 0.11 3.62 ± 0.30 1.76 ± 0.12 1.85 ± 0.12 8.30 ± 3.04 ≤ 2
RC (×10−3 mm2/s) 0.0107 0.0082 0.0041 0.0067 0.0269 0.0279 ≤ 0.0150
Repeatability wCV (%) 0.33 0.28 0.15 0.22 0.86 0.91 N/A
RDC (×10−3 mm2/s) 0.0305 0.0254 0.0027 0.0055 0.0416 0.0049 ≤ 0.0650
Reproducibility wCV (%) 0.96 0.85 0.10 0.18 1.36 0.16 N/A
SNR b = 0 s/mm2 87.3
(78.6 – 96.1)
180.5
(168.0 – 193.0)
161.3
(146.7 – 175.9)
171.0
(160.8 – 181.2)
171.9
(147.5 – 196.3)
130.2
(112.3 – 148.0)
≥ 50 ± 5
SNR b = 150 s/mm2 118.9
(102.8 – 135.0)
193.1
(183.6 – 202.6)
266.2
(242.2 – 290.1)
N/A N/A N/A N/A
SNR b = 500 s/mm2 or b = 800 s/mm2 106.7
(94.6 – 118.8)
(b = 500)
184.0
(174.8 – 193.3)
(b = 500)
261.6
(241.6 – 281.7)
(b = 500)
232.2
(222.8 – 241.7)
(b = 500)
104.1
(89.4 – 118.7)
(b = 800)
97.3
(83.5 – 111.1)
(b = 800)
N/A