Skip to main content
. 2023 Sep 18;14(9):478. doi: 10.3390/jfb14090478

Table 6.

Spontaneous platelet aggregation in citrated whole blood after contact with the tested biomaterials assessed by scattering laser light forward and sideways (a.u.). The results shown represent the mean values and standard deviations (MEAN ± SD) of the percentage of aggregated platelets from four independent experiments with three replicates of each sample (n = 12). For negative and positive controls n = 48. Explanations of other symbols are in the text. Statistical significance levels are marked with asterisks as follows: *** p < 0.001, ns—not significant.

A
Grinding
B
Polishing
C
Rosette Like
D
Crater Like
Significance
AISI 316L
{1}
1.14 ± 0.25 1.24 ± 0.26 1.53 ± 0.57 1.12 ± 0.37 ns
CoCrMo
{2}
1.29 ± 0.40 1.30 ± 0.18 1.68 ± 0.46 1.32 ± 0.59 ns
Ti6Al4V
{3}
1.31 ± 0.39 1.16 ± 0.19 1.49 ± 0.52 1.18 ± 0.44 ns
Ti6Al7Nb
{4}
1.28 ± 0.34 1.29 ± 0.21 1.64 ± 0.65 1.23 ± 0.57 ns
negative control {5} 1.34 ± 0.44 *** {5}–{6}
positive control {6} 1.88 ± 0.77
significance ns ns ns ns