Limit of detection analyses on serially diluted DNA in triplicate.

Sample Copies/µL Copies/Reaction PrRT Positive Reactions INT Positive Reactions
8E5 cell line DNA 515.8 5157.6 3 3
103.2 1031.5 3 3
51.6 515.8 3 3
25.8 257.9 3 3
12.9 128.9 3 3
6.4 64.5 3 3
3.2 32.2 3 3
1.6 16.1 2 * 2 *
0.8 8.1 3 3
Study
sample
12.7 127.0 3 3
6.35 63.5 2 * 3
3.18 31.8 3 3
1.59 15.9 3 3
0.79 7.9 3 3
0.4 4.0 3 3
0.2 2.0 0 1

* Serial dilutions at low DNA concentrations risk exclusion of target DNA from some aliquots.