Table 15.
Operational risks, and measures to mitigate risk.
Hazard | Potential cause | Mitigation |
---|---|---|
Injury from spring decompression | Uncontrolled release of compressed spring and plunger during loading. | Follow assembly instructions, ensure the escapement and one-way bearing lock the screw, and ensure the grub screws are installed securely in the plunger. |
Injury from needle | Contact with the exposed needle of the infusion set. | Follow manufacturer’s instructions for handling infusion sets and cover exposed needle with a sheath when the pump is not in use. |
Injury from sharp edges on manufactured components | Burrs or rough surfaces on pump housings could cause injury when loading and handling the pump. | De-burr all components after manufacture and ensure there are no exposed sharp edges. |