Table 1.
CVVH |
Total effluent rate = total UF rate (sum of pre-filter and post-filter replacement fluid rate, ml/h) + fluid removal rate (ml/h) |
CVVHD |
Total effluent rate = dialysate rate (ml/h) + fluid removal rate (ml/h) |
CVVHDF |
Total effluent rate = total UF rate (sum of pre-filter and post-filter replacement fluid rate, ml/h) + dialysate rate (ml/h) + fluid removal rate (ml/h) |
Dilution factor for pre-dilution |
Plasma flow rate (ml/h)/[plasma flow rate (ml/h) + pre-filter replacement fluid rate (ml/h)] |
Plasma flow rate (ml/h) = blood flow rate (ml/min) × 60 (min/h) x (1 – HCT) |
RRT renal replacement therapy, CVVH continuous venovenous hemofiltration, CVVHD continuous venovenous hemodialysis, CVVHDF continuous venovenous hemodiafiltration, UF ultrafiltration, HCT current patient hematocrit