Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2023 Jul 5;1228:123815. doi: 10.1016/j.jchromb.2023.123815

Table 3: Precision and accuracy of quality control samples (n=6).

Analyte Nominal concentration (ng/ml) Accuracy (Bias (%)) Precision (CV (%))
Day 1 Day 2 Day 3 Day 1 Day 2 Day 3 Inter-day (n=18)
PGP 0.01 158.10 −18.57 35.10 11.76 41.40 27.14 64.11
0.07 17.89 −2.73 7.03 7.91 16.75 18.68 19.89
0.25 −22.68 −17.92 −16.76 8.07 8.05 14.61 16.15
2.5 1.64 12.32 −0.60 5.39 12.39 10.87 12.26
25 6.20 13.12 3.00 9.59 8.98 10.67 11.39
50 13.42 14.10 0.32 6.47 8.66 10.81 10.53
AcPGP 0.01 −15.41 1.50 51.30 84.20 111.32 48.56 61.67
0.07 15.77 −7.96 12.54 18.19 18.09 16.97 19.54
0.25 6.00 −1.72 −2.20 6.80 11.58 6.96 12.27
2.5 8.00 −3.56 −1.56 5.39 7.96 1.98 5.45
25 10.36 −3.36 −5.64 2.00 4.61 4.76 6.56
50 9.64 −4.38 −4.90 3.46 1.76 6.21 4.64