Dear Editors,
Since the early 20th century, protection from the deleterious effects of ionizing radiation in diagnostic imaging examinations has been a much-discussed topic. At present, there is a consensus among safety agencies about the need to try, whenever possible, to reduce the levels of radiation to which professionals and patients are exposed, based on the ALARA (as low as reasonably achievable), ALADA (as low as diagnostically acceptable) and ALADAIP (as low as diagnostically acceptable being indication-oriented and patient-specific) principles.1,2,3
The last decade has been marked by growth in the sale and use of portable hand-held X-ray devices in dental offices and universities. Since the use of these devices makes it difficult for the operator to maintain a distance of 2 meters between the radiation source and patient, it is necessary to clarify the radioprotection norms. Given the lack of clear, concise, and direct information on this topic, this Letter to the Editor is extremely relevant.
The development of the portable hand-held dental X-ray device has led to several advantages, including a reduction in size and weight, which allows for easy transportation to any required location. This makes it an appealing option for use in a surgical center during operations, in forensic dentistry, community work, and home care.4 However, the use of portable hand-held X-ray devices in routine dental care is not recommended due to the secondary radiation dose that the operator may receive during radiographic examinations.4,5,6,7,8,9,10,11,12,13,14
Studies have been conducted to develop protocols that ensure excellent operator safety, reducing or avoiding exposure to radiation. Protective measures have proven effective in reducing the radiation dose to the operator, thereby enhancing protection against secondary radiation during radiographic exams conducted with portable hand-held dental X-ray devices. These measures include 1) operating the portable device with arms fully extended to maintain distance from the body, 2) using a backscatter shield on the cylinder, 3) employing a longer cylinder, 4) wearing protective aprons, 5) using lead gloves, and 6) utilizing a rectangular collimator. These precautions are particularly crucial when the operator is in close proximity to both the radiation source and the patient.4,5,6,7,8,9,10,11,12,13,14,15 Additionally, staff members and the public should maintain a distance of 2 meters from both the patient and the radiation source, and avoid standing in the path of the central X-ray beam.
It is important to emphasize that radioprotection measures will only minimize the effective dose received by the professional if the operator holds the device during the examination. Only when portable hand-held dental X-ray devices are used on a stand and operated from a protected area (either 2 meters away or behind a barrier), similar to the usage of conventional radiographic devices, can operators be fully safeguarded against secondary radiation. Consequently, manufacturers should endeavor to create supports for portable devices that enable the operator to emit radiation from a safe distance.
Even if the equivalent dose in the head and extremity regions of the portable device operator does not exceed 0.6 mSv/year and 20 mSv/year respectively, which are within the limits deemed acceptable by the International Commission on Radiological Protection (ICRP),4,5,6,7,8,9,10,11,12,13,14 it is necessary to consider the linear non-threshold (LNT) theory. The LNT theory posits a linear relationship between the dose and the risk of inducing stochastic effects, even at low doses, implying that there is no safe dose limit. Despite the dose intensities falling within the low range, there are several reasons to apply the LNT theory. First, a policy is required to establish exposure limits for individuals in the low-dose range, including those undergoing diagnostic imaging procedures and occupational exposure. Second, current epidemiological evidence neither confirms nor rules out the existence of a safe dose in oral radiology. Therefore, the LNT theory, which is scientifically plausible, should be considered even at low radiation doses. Most radiation protection organizations concur that it is prudent to assume the risk is proportional to the dose, even in the case of diagnostic exposure.16
In conclusion, it is essential to use protective measures to nullify or reduce the secondary radiation exposure received by the operator of the portable hand-held dental X-ray device. Adhering to the principles of optimization (ALARA) and dose limitation, the portable hand-held dental X-ray device should only be utilized under the circumstances outlined in the third paragraph of this letter, when the use of a conventional device is not feasible. Specifically, a safe scenario where the operator is positioned 2 meters away from the patient, at an angle of 90°-135° from the central X-ray beam. Companies that distribute these devices bear the responsibility for disseminating this information and instructing professionals on radioprotection measures for the operator, staff, and the public.
Footnotes
Conflicts of Interest: None
References
- 1.Lurie AG. Doses, benefits, safety, and risks in oral and maxillofacial diagnostic imaging. Health Phys. 2019;116:163–169. doi: 10.1097/HP.0000000000001030. [DOI] [PubMed] [Google Scholar]
- 2.Yeung AW, Jacobs R, Bornstein MM. Novel low-dose protocols using cone beam computed tomography in dental medicine: a review focusing on indications, limitations, and future possibilities. Clin Oral Investig. 2019;23:2573–2581. doi: 10.1007/s00784-019-02907-y. [DOI] [PubMed] [Google Scholar]
- 3.Oenning AC, Pauwels R, Stratis A, De Faria Vasconcelos K, Tijskens E, De Grauwe A, et al. Halve the dose while maintaining image quality in pediatric Cone Beam CT. Sci Rep. 2019;9:5521. doi: 10.1038/s41598-019-41949-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Makdissi J, Pawar RR, Johnson B, Chong BS. The effects of device position on the operator's radiation dose when using a handheld portable X-ray device. Dentomaxillofac Radiol. 2016;45:20150245. doi: 10.1259/dmfr.20150245. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Goren AD, Bonvento M, Biernacki J, Colosi DC. Radiation exposure with the NOMAD portable X-ray system. Dentomaxillofac Radiol. 2008;37:109–112. doi: 10.1259/dmfr/33303181. [DOI] [PubMed] [Google Scholar]
- 6.Danforth RA, Herschaft EE, Leonowich JA. Operator exposure to scatter radiation from a portable hand-held dental radiation emitting device (Aribex NOMAD) while making 915 intraoral dental radiographs. J Forensic Sci. 2009;54:415–421. doi: 10.1111/j.1556-4029.2008.00960.x. [DOI] [PubMed] [Google Scholar]
- 7.Cho JY, Han WJ. The reduction methods of operator’s radiation dose for portable dental X-ray machines. Restor Dent Endod. 2012;37:160–164. doi: 10.5395/rde.2012.37.3.160. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.McGiff TJ, Danforth RA, Herschaft EE. Maintaining radiation exposures as low as reasonably achievable (ALARA) for dental personnel operating portable hand-held X-ray equipment. Health Phys. 2012;103(2 Suppl 2):S179–S185. doi: 10.1097/HP.0b013e318259fa29. [DOI] [PubMed] [Google Scholar]
- 9.Hosseini Pooya SM, Hafezi L, Manafi F, Talaeipour AR. Assessment of the radiological safety of a Genoray portable dental X-ray unit. Dentomaxillofac Radiol. 2015;44:20140255. doi: 10.1259/dmfr.20140255. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Iwawaki A, Otaka Y, Asami R, Ozawa T, Izawa M, Saka H. The study of protection of operators and surrounding workers at the time of using portable intraoral X-ray units. Leg Med (Tokyo) 2018;33:66–71. doi: 10.1016/j.legalmed.2018.05.007. [DOI] [PubMed] [Google Scholar]
- 11.Zenóbio EG, Zenóbio MA, Azevedo CD, Nogueira MDS, Almeida CD, Manzi FR. Assessment of image quality and exposure parameters of an intraoral portable X-rays device. Dentomaxillofac Radiol. 2019;48:20180329. doi: 10.1259/dmfr.20180329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Otaka Y, Harata Y, Izawa M, Iwawaki A, Ishii T, Saka H, et al. Reduction of operator exposure by rectangular collimation in portable intraoral radiography. Radiol Phys Technol. 2020;13:312–320. doi: 10.1007/s12194-020-00573-4. [DOI] [PubMed] [Google Scholar]
- 13.Iwawaki A, Otaka Y, Asami R, Ishii T, Kito S, Tamatsu Y, et al. Comparison of air dose and operator exposure from portable X-ray units. Leg Med (Tokyo) 2020;47:101787. doi: 10.1016/j.legalmed.2020.101787. [DOI] [PubMed] [Google Scholar]
- 14.Leadbeatter J, Diffey J. Evaluation of radiation exposure to operators of portable hand-held dental X-ray units. Phys Eng Sci Med. 2021;44:377–385. doi: 10.1007/s13246-021-00985-5. [DOI] [PubMed] [Google Scholar]
- 15.Otaka Y, Harata Y, Izawa M, Iwawaki A, Asami R, Saka H, et al. Efficacy of shields against the backscatter radiation of portable X-ray units. Radiat Saf Manag. 2018;17:1–12. [Google Scholar]
- 16.Mossman KL. The LNT debate in radiation protection: science vs. policy. Dose Response. 2012;10:190–202. doi: 10.2203/dose-response.11-017.Mossman. [DOI] [PMC free article] [PubMed] [Google Scholar]