Skip to main content
. 2023 Aug 4;151:e151. doi: 10.1017/S0950268823000845

Table 2.

Comparison of measles IgM RDT result interpretations based on relative reflectance measured using the ESEQuant Lateral Flow Reader with the Siemens Enzygnost Anti-Measles Virus/IgM EIA

Measles IgM RDT result interpretation based on ESEQuant LF Reader measurements Total
Positive (Test Line ≥ 60 mV) Negative (Test Line < 60 mV)
Measles IgM EIA results Positive 73 5 78
Equivocal 1 0 1
Negative 1 45 46
Total 75 50 125

A measles IgM RDT was interpreted as valid when the ESEQuant LF Reader reflectance peak height measurement at the Control Line was ≥60 mV. A relative reflectance peak height measurement of ≥60 mV at the Test Line was interpreted as positive for valid measles IgM RDTs. EIA, enzyme immunoassay; ESEQuant LF Reader, ESEQuant Lateral Flow Reader; mV, millivolts; RDT, rapid diagnostic test.