Skip to main content
. 2023 Oct 3;16:343. doi: 10.1186/s13071-023-05942-7

Table 2.

Results obtained by diagnostic methods used in this study in comparison with microscopy as reference standard

Total Microscopy
Positive (N = 67) Negative (N = 133)
RDT +  98 62 (92.5%) 36 (27.1%)
RDT− 102 5 (7.5%) 97 (72.9%)
Field-cLAMP +  112 66 (98.5%) 46 (34.6%)
Field-cLAMP− 88 1 (1.5%) 87 (65.4%)
Lab-cLAMP +  91 60 (89.6%) 31 (23.3%)
Lab-cLAMP− 109 7 (10.4%) 102 (76.7%)
Lab-qLAMP +  86 58 (86.6%) 28 (21.1%)
Lab-qLAMP− 114 9 (13.4%) 105 (78.9%)

RDT + and RDT-, positive and negative results by Rapid Diagnostic Tests; Field-cLAMP + and Field-cLAMP-, positive and negative results detected by colorimetric field LAMP assay; Lab-cLAMP + and Lab-cLAMP-, positive and negative results obtained by colorimetric laboratory LAMP assay; Lab-qLAMP + and Lab-qLAMP-, positive and negative results amplified by laboratory real-time LAMP assay