Table 4.
The intrarater reliability of the standing position measurements obtained by the Smartphone Surgimap Application and Surgimap System Software.
| Readings from Smartphone Surgimap Application (R1, R2, R3) versus reading from Surgimap System Software (R) | ICC | 95% CI |
SEM | MDC | ||
|---|---|---|---|---|---|---|
| Lower | Upper | |||||
| Group A Standing | R1 Vs R | .978 | .959 | .988 | 0.320 | 1.569 |
| R2 Vs R | .985 | .973 | .992 | 0.261 | 1.417 | |
| R3 Vs R | .972 | .948 | .984 | 0.351 | 1.642 | |
| Group B Standing | R1 Vs R | .986 | .974 | .992 | 0.297 | 1.510 |
| R2 Vs R | .990 | .981 | .994 | 0.253 | 1.395 | |
| R3 Vs R | .980 | .964 | .989 | 0.348 | 1.636 | |
ICC: intraclass correlation coefficient; CI: confidence interval; SEM: standard error of measurement; MDC: minimum detectable change.