Table 2 |.
Troubleshooting table
Step | Problem | Possible reason | Solution |
---|---|---|---|
10 (Procedure 1), 11 (Procedure 2) | Blank fluorescence reading | The instrument does not have the right filters | Replace the filters to 435Ex/485Em, 450Ex/480Em or similar |
The reading optics do not match the type of plate The plate sealer is not transparent at the needed wavelength | Clear-bottom plates are compatible with top and bottom optics, but black-bottom plates are compatible only with top reading Use the recommended plate seal (cat. no. 4311971) or use a clear- bottom plate |
||
16 (Procedure 1), 17 (Procedure 2) | Spontaneous self-aggregation | PIPES buffer | Prepare new PIPES and use NaOH only for pH adjustment. If too much NaOH is added, start again and prepare a new solution. Do not add HCl or other acids |
Undissolved PIPES | Pass the buffer through 0.22-μm filters to remove undissolved PIPES | ||
Hydrated PIPES powder | Buy new reagent and maintain it inside a desiccator cabinet | ||
Substrate contains pre-aggregated αSyn | Re-filtrate the substrate by using a 100-kDa-cutoff filter | ||
A shaking mode other than orbital shaking was selected | Program the shaker for orbital shaking. Protocols that use double orbital shaking use different shaking speeds and overall assay conditions | ||
Poor-quality substrate | Confirm the protein concentration of the stock. Underestimation could lead to a higher concentration of the protein in the assay than desired Buy or purify a new batch of substrate | ||
Substrate dimerization | Make sure that the plasmid encodes tyrosine at codon 136 by using TAT nucleotides70 Analyze the substrate by using DTT or 2-Mercaptoethanol (BME). If dimers were formed by disulfide bonds, after DTT, the protein should appear monomeric. If this is the problem, the plasmid needs to be modified as above If the dimers remain (confirmed by western blot), get a new batch of rec-αSyn |