Table 1.

AI-based approaches for attenuation correction

Application Approach Advance Quantitative Comparison vs. Acquired CT Reference
Brain perfusion SPECT CNN Enable correction for SPECT only scanners PSNR 62.2 dB, SSIM 0.9995 Sakaguchi et al., 202112
Brain PET (FDG) CNN, UNet Enable correction for PET only scanners Dice coefficients of 0.80 ± 0.02 for air, 0.94 ± 0.01 for soft tissue, and 0.75 ± 0.03 for bone, MAE of 111 ± 16 HU, PET reconstruction bias <1% Liu et al., 201813
Brain PET/MR (FDG) CNN Synthesize CT from T1 MRI for PET/MRI Dice coefficient of 0.936 ± 0.011 for soft tissue, and 0.803 ± 0.021 for bone, PET reconstruction bias −0.7% ± 1.1% Liu et al., 201714
Brain PET/MR (FDG) CNN, GAN Synthesize CT from T1 MRI for PET/MRI Dice coefficient of 0.77 ± 0.07 for bone, PET reconstruction bias <4% Arabi et al., 201915
Brain PET/MR (FDG) CNN, GAN Synthesize CT from T1 MRI for PET/MRI Dice coefficient of 0.74 ± 0.05 for bone, PET reconstruction bias <0.025 SUV Gong et al., 202016
Brain PET/MR (FDG) CNN, UNet Synthesize CT from T1 MRI for PET/MRI Dice coefficient of 0.81 ± 0.03 for bone, PET RMSE 253.5, PET reconstruction bias <1% Blanc-Durand et al., 201917
Brain PET/MR (FDG) CNN, UNet Synthesize CT from T1 and ZTE MRI for PET/MRI Dice coefficient of 0.86 ± 0.03 for bone, PET reconstruction bias <0.02 SUV Gong et al., 201818
Brain PET/MR (FDG) CNN, UNet Synthesize CT from UTE MRI for PET/MRI Dice coefficient of 0.96 ± 0.006 for soft tissue, 0.88 ± 0.01 for bone, PET reconstruction bias <1% Jang et al., 201819
Brain PET/MR (FET) CNN, UNet Synthesize CT from UTE MRI for PET/MRI PET reconstruction bias <1% Ladefoged et al., 201920
Myocardial perfusion SPECT CNN Enable correction for SPECT only scanners MAE 3.60% ± 0.85% in whole body, 1.33% ± 3.80% in the left ventricle Shi et al., 202021
Myocardial perfusion SPECT CNN Enable correction for SPECT only scanners Mean segmental errors 0.49 ± 4.35%, mean absolute segmental errors 3.31 ± 2.87% Yang et al., 202122
Myocardial perfusion SPECT GAN Enable correction for SPECT only scanners RMSE 0.141±0.0768, PSNR 36.382±3.74 dB, SSIM 0.995±0.0043 Torkaman et al., 202123
Pelvis PET/MR (68Ga-PSMA) CNN, GAN Synthesize CT from T1 MRI for PET/MRI PET reconstruction bias <2.2% Pozaruk et al., 202124
Pelvis PET/MR (FDG) CNN Synthesize CT from T1 and T2 MRI for PET/MRI Dice coefficient of 0.98 ± 0.01 for soft tissue, 0.79 ± 0.03 for cortical bone, PET reconstruction bias 4.9% Bradshaw et al., 201825
Pelvis PET/MR (FDG) CNN, UNet Synthesize CT from T1 MRI for PET/MRI PET reconstruction bias <4% Torrado-Carvajal et al., 201926
Whole body PET (FDG) CNN, GAN Enable correction for PET only scanners PET reconstruction bias −0.8 ± 8.6% Armanious et al., 202027