Table 5.

ECG QRS characteristics (duration, amplitude, and morphology) and correlation coefficients comparison between Smartwatch and standard 12-leads ECG.

Variables Lead SW-ECG 12-Leads ECG Correlation coefficient* p-value
Duration (milliseconds) DI 84.13 ± 18.35 101.92 ± 21.02 97.3 % <0.05
V1 96.43 ± 21.29 102.04 ± 19.32 98.1 % <0.05
V3 96.99 ± 21.13 102.74 ± 18.52 97.6 % <0.05
V6 85.13 ± 22.41 103.04 ± 20.41 96.3 % <0.05
Amplitude (millivolts) DI 7.9 ± 7.83 9,72 ± 6.22 68.7 % <0.05
V1 2.36 ± 1.79 8.48 ± 4.22 69.6 % <0.05
V3 4.00 ± 2.84 9.08 ± 3.92 77.6 % <0.05
V6 11.06 ± 8.63 9.88 ± 5.02 80 % <0.05
Morphology (Narrow/Wide) DI N = 124/W = 13 N = 110/W = 27 84.3 % <0.05
V1 N = 110/W = 26 N = 108/W = 29 68.9 % <0.05
V3 N = 108/W = 28 N = 105/W = 26 66.6 % <0.05
V6 N = 114/W = 22 N = 106/W = 27 53 % <0.05
N: Narrow/W: Wide.

A strong correlation between the segment ST of the SW and standard ECG was found, except for DI lead, where the correlation coefficient was 55 %, as reported in Table 6.