Comparison of Safety of Leadless Pacemakers and Transvenous Pacemakers: A Meta-Analysis

Mahesh Gangannapalle 1, Obinna Monday 2, Anurag Rawat 3, Ugonna A Nwoko 4, Arun Kumar Mandal 5, Maham Babur 6, Tayyaba J Khan 7, Sujith K Palleti 8,
Editors: Alexander Muacevic, John R Adler
PMCID: PMC10568652  PMID: 37842352

Abstract

Pacemakers have been accessible for six decades, and clearly defined criteria for pacemaker implantation have been established. Within the contemporary clinical practice, two dependable pacing platforms exist leadless pacemakers and transvenous pacemakers. The aim of this meta-analysis is to compare the safety of leadless pacemakers to transvenous pacemakers. This meta-analysis adhered to the guidelines outlined in the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 framework. A comprehensive and systematic search was conducted across various databases including Scopus, Cochrane Library, and EMBASE, spanning from inception to August 15, 2023. The primary outcomes assessed in this meta-analysis were total complications, all-cause mortality, and device-related complications. Furthermore, secondary outcomes evaluated encompassed the need for reintervention, occurrences of pneumothorax, pericardial effusion, endocarditis, hemothorax, and hematoma. Total 17 studies were included in this meta-analysis. The findings of this study showed that patients with leadless pacemakers had a lower risk of total complications, device-related complications, pneumothorax, and endocarditis. The risk of reintervention was significantly lower in the leadless pacemaker group. However, compared to a transvenous pacemaker, the risk of pericardial effusion was significantly higher in the leadless pacemaker group. It is important to acknowledge the limitations arising from the lack of extensive long-term follow-up data for leadless pacemakers. As technology evolves, continued research will be essential in uncovering the full spectrum of prolonged complications associated with these devices.

Keywords: meta-analysis, complications, safety, transvenous pacemakers, leadless pacemakers

Introduction and background

Pacemakers have been accessible for six decades, and clearly defined criteria for pacemaker implantation have been established based on evidence from randomized controlled trials (RCTs) [1,2]. Within contemporary clinical practice, two dependable pacing platforms exist: leadless pacemakers and transvenous pacemakers [3]. Despite significant technological advancements in implantable electronic cardiac devices, the implantation procedure for conventional transvenous pacemakers - entailing a subcutaneous pulse generator and transvenous leads - has endured without alteration. The subcutaneous pocket and transvenous leads are the primary culprits behind complications arising from transvenous pacemaker implantation, affecting approximately 3%-12% of patients [4,5]. To mitigate the incidence of these issues, the concept of leadless pacemakers was introduced in the 1970s [6]. Leadless intracardiac pacemakers, deployed through a minimally invasive femoral vein approach, are fully implanted within the right ventricle and are currently utilized in patients requiring single-chamber ventricular pacing [7].

Approval for the use of leadless pacemakers by the United States Food and Drug Administration (FDA) was initially granted in April 2016 [8]. In recent times, leadless pacemakers have gained considerable popularity. These self-contained devices integrate a pulse generator, battery, and electrode into a single unit, addressing certain limitations of conventional transvenous pacemakers [9]. Positioned entirely within the right ventricle, initial clinical trials have demonstrated the excellent performance of leadless pacemakers in terms of device electrical parameters for up to a year. These early trials also highlighted lower rates of lead and pocket-related complications compared to traditional pacemakers, with reported rates ranging from 3.4% to 6.5% [9,10].

Notably, recent findings have suggested a potential safety advantage of leadless pacemakers over transvenous pacemakers, although these studies encompassed relatively small patient cohorts. As a result, this meta-analysis seeks to enhance precision by consolidating data from studies directly comparing leadless pacemakers to transvenous pacemakers, aiming to establish a robust evaluation of safety outcomes. The aim of this meta-analysis was to compare the safety of leadless pacemakers to transvenous pacemakers.

Review

Methodology

This meta-analysis adhered to the guidelines outlined in the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 framework.

Literature Search and Inclusion Criteria

A comprehensive and systematic search was conducted across various databases including Scopus, Cochrane Library, and EMBASE, spanning from inception to August 15, 2023. The search utilized specific key terms such as “leadless cardiac pacing,” “leadless pacing,” “Micra transcatheter pacing,” “leadless cardiac pacemaker,” and “leadless pacemaker,” in conjunction with terms like “standard pacemaker,” “permanent pacemaker,” “conventional pacemaker,” and “traditional pacemaker.” Additionally, a manual search of references in the included studies was conducted to identify potential supplementary research. The acquired records from electronic databases were imported into EndNote X9 software, and duplicate entries were eliminated. The screening process encompassed an initial evaluation of titles and abstracts, followed by a full-text examination to ascertain adherence to eligibility criteria.

Eligibility Criteria

This meta-analysis incorporated any RCT or non-RCT (non-RCT) that juxtaposed leadless pacemakers with transvenous pacemakers while reporting relevant outcomes. Inclusion was confined to studies published in the English language. Notably, meta-analyses, reviews, editorials, and expert opinions were excluded. The primary outcomes assessed included total complications, all-cause mortality, and device-related complications. Furthermore, secondary outcomes evaluated encompassed the need for reintervention, occurrences of pneumothorax, pericardial effusion, endocarditis, hemothorax, and hematoma.

Data Extraction and Quality Appraisal

Data extraction entailed gathering pertinent details from included studies, such as the first author, publication year, study design, geographic location, sample size, follow-up duration, participant characteristics, and outcomes. This process was independently conducted by two authors utilizing a pre-designed data extraction form in Excel. Any discrepancies that arose during data extraction were resolved through consensus. The quality of included studies was evaluated using the New Castle Ottawa Scale (NCOS), which assesses studies across categories such as group selection, group comparability, and exposure or outcome assessment. The cumulative score from these categories indicated the overall study quality.

Data Analysis

Data analysis was executed using RevMan Version 5.4.1 (The Cochrane Collaboration) and STATA 16.0 (Stata Corp). Risk ratios (RR) were presented alongside their corresponding 95% confidence intervals (CI). A significance threshold of P-value < 0.05 was employed. Heterogeneity was assessed through I-square values, where an I-square exceeding 50% was deemed indicative of significant heterogeneity. In cases of notable heterogeneity, RR calculations were performed using a random-effects model; conversely, a fixed-effects model was employed for analysis when heterogeneity was insignificant. We performed meta-regression to understand to understand the potential sources of heterogeneity observed across the included studies in our meta-analysis taking total complications (primary outcome) as a dependent variable. Meta-regression allows us to explore how various covariates or factors may influence the effect sizes or outcomes reported in the studies.

Results

The online database search yielded 1,255 citations. After screening abstracts and titles, we excluded 1,186 studies. We obtained full text of 31 studies and detailed assessment was performed based on pre-defined inclusion and exclusion criteria. Finally, 20 studies were included in this meta-analysis. Figure 1 shows the PRISMA flowchart of study selection. Table 1 presents the characteristics of the included studies. Table 2 presents a quality assessment of the included studies. Three domains were assessed including selection, comparability and exposure, and outcome ascertainment. Each of these domains is further divided into specific criteria, and studies are awarded points for meeting these criteria.

Table 1. Characteristics of included studies.

RC: Retrospective cohort, PC: Prospective cohort

Author Name Year Region Study Design Groups Sample Size Mean Age (Years) Males (%)
Bodin et al. [11] 2022 France RC Leadless 1344 70.7 57.9
Tranvenous 1344 83.4 60.2
Boveda et al. [12] 2023 United States RC Leadless 5514 77.7 56.8
Tranvenous 6422 80.6 59.4
Cantillon et al. [13] 2018 United States RC Leadless 718 75.6 62.3
Tranvenous 1436 76.1 63
Chami et al. [14] 2022 United States RC Leadless 6219 79.5 55.9
Tranvenous 10212 82 56.8
Martinez-Sande et al. [15] 2020 Spain PC Leadless 245 83.6 27.3
Tranvenous 198 79.2 62.1
Moore et al. [16] 2019 United States RC Leadless 10 82.8 60
Tranvenous 23 84 52
Okuyama et al. [17] 2020 Japan RC Leadless 10 86.5 20%
Tranvenous 14 83 21
Pagan et al. [18] 2020 United States RC Leadless 183 89.5 51.9
Tranvenous 119 89.9 40.3
Palmisano et al. [19] 2021 Italy PC Leadless 77 77.5 71.4
Tranvenous 77 78.3 63.6
Palmisano et al. [20] 2023 Italy RC Leadless 442 73.2 68.3
Tranvenous 442 74 68.8
Reynolds et al. [7] 2016 United States PC Leadless 725 75.9 58.8
Tranvenous 2667 71.1 55.1
Sanchez et al. [21] 2020 United States RC Leadless 67 73 54
Tranvenous 131 74 27
Sasaki et al. [22] 2022 Japan RC Leadless 58 81 38
Tranvenous 58 82 40
Tachibana et al. [23] 2020 Japan RC Leadless 35 90.1 44.4
Tranvenous 27 90.9 40
Vaibhav et al. [24] 2018 United States RC Leadless 90 80.5 63
Tranvenous 90 78.2 63
Yarlagadda et al. [25] 2018 United States RC Leadless 60 74 48
Tranvenous 67 74 24
Zucchelli et al. [3] 2021 Italy RC Leadless 100 77.5 77
Tranvenous 100 78.8 67

Table 2. Quality assessment of included studies using New Castle Ottawa Scale.

Selection: Out of 3, Comparability: Out of 2, Exposure or outcome assessment: Out of 4

Author Name Selection Comparison Outcome Overall
Bodin et al. [11] 2 1 3 Fair
Boveda et al. [12] 3 2 4 Good
Cantillon et al. [13] 2 2 3 Good
Chami et al. [14] 2 2 3 Good
Martinez-Sande et al. [15] 2 2 4 Good
Moore et al. [16] 2 2 4 Good
Okuyama et al. [17] 3 1 3 Good
Pagan et al. [18] 3 2 3 Good
Palmisano et al. [19] 2 2 4 Good
Palmisano et al. [20] 3 2 3 Good
Reynolds et al. [7] 3 1 3 Good
Sanchez et al. [21] 3 2 3 Good
Sasaki et al. [22] 3 2 4 Good
Tachibana et al. [23] 2 1 3 Fair
Vaibhav et al. [24] 2 1 2 Fair
Yarlagadda et al. [25] 2 2 3 Good
Zucchelli et al. [3] 3 2 4 Good

Figure 1. PRISMA flowchart showing the process of study selection.

Figure 1

Primary End Points

Total complications: Thirteen studies compared the risk of total complications between leadless and transvenous pacemaker groups. As shown in Figure 2, the risk of all complications was significantly lower in the leadless pacemaker group compared to the transvenous pacemaker group (RR: 0.67, 95% CI: 0.47-0.94). Significant heterogeneity was reported among the study results.

Figure 2. Comparison of total complications between leadless and transvenous pacemaker groups.

Figure 2

Sources: References [3,7,12-18,20,23-25]

Mortality: Ten studies were included in the pooled analysis of the risk of mortality. A pooled analysis of 10 studies showed that the risk of all-cause mortality was higher in the transvenous pacemaker group compared to the leadless pacemaker group. However, the difference was statistically insignificant (RR: 0.80, 95% CI: 0.63-1.03) as shown in Figure 3. Significant heterogeneity was reported among the study results.

Figure 3. Comparison of mortality between leadless and transvenous pacemaker groups.

Figure 3

Sources: References [3,11,12,14-17,19,24,25]

Device-related complications: Eleven studies were included in the pooled analysis of device-related complications. As shown in Figure 4, the risk of device-related complications was significantly higher in the leadless pacemaker group compared to the transvenous pacemaker group (RR: 0.49, 95% CI: 0.43-0.57). No significant heterogeneity was reported among the study results.

Figure 4. Comparison of total device-related complications between leadless and transvenous pacemaker groups.

Figure 4

Sources: References [3,7,12-15,17,18,22,23,25]

Secondary Outcomes

The outcomes of secondary importance are presented in Table 3. A noteworthy observation is the notable disparity in the risk of certain complications between the leadless and transvenous pacemaker groups. Specifically, patients in the leadless group exhibited significantly reduced risks of pneumothorax and endocarditis compared to those in the transvenous group. Conversely, the risk of pericardial effusion was found to be substantially greater in the leadless pacemaker group compared to the alternative study group. No statistically significant differences were detected between the two groups concerning the risks of hemothorax and hematoma.

Table 3. Comparison of secondary outcome between leadless and transvenous pacemakers.

RR: Risk ratio; CI: Confidence interval

RR>1 signifies risk of outcome is greater in leadless group, while RR<1 signifies risk of outcome is greater in transvenous group

Outcome RR 95% CI I-square
Pneumothorax 0.22 0.07-0.71 17%
Pericardial effusion 2.39 1.37-4.17 0%
Endocarditis 0.18 0.06-0.55 0%
Required reintervention 0.55 0.47-0.66 0%
Hemothorax 0.94 0.65-1.36 0%
Hematoma 0.58 0.21-1.66 0%

Meta-regression

Table 4 shows the results of meta-regression taking total complications as the dependent variable. Hypertension is statistically significant in predicting total complications (p = 0.021). The coefficient's value further suggests a positive relationship between hypertension and total complications. No significant impact of diabetes, heart failure history, gender, and age is there in predicting total complications.

Table 4. Meta-regression taking total complications as a dependent variable.

The statistical test used to assess the significance of the relationships between predictor variables and the total complications is the Wald Test.

Factor Coefficient P-value
Diabetes 0.000085 0.213
Gender -0.000033 0.089
Hypertension 0.000079 0.021
Heart failure -0.001271 0.307
Age 0.00011 0.161

Discussion

In this meta-analysis, we compared the safety of a transvenous pacemaker with a leadless pacemaker. Patients with leadless pacemakers had a lower risk of total complications, device-related complications, pneumothorax, and endocarditis. The risk of reintervention was significantly lower in the leadless pacemaker group. However, compared to a transvenous pacemaker, the risk of pericardial effusion was significantly higher in the leadless pacemaker group. The findings of this meta-analysis confirm the value of leadless pacemakers, which is in line with ESC guidelines [26]. The findings of this meta-analysis are also in line with previous meta-analyses comparing the safety of leadless pacemakers and transvenous pacemakers [27]. Transvenous pacemakers are equipped with leads and necessitate a pocket for housing the pacemaker generator. This characteristic renders them susceptible to complications such as pocket hematoma and pocket infection [7].

Comparatively, a higher incidence of endocarditis was observed in cases involving transvenous pacemakers in contrast to leadless pacemakers, often necessitating the removal of the lead [3] [24]. The absence of implanted leads could confer benefits in mitigating the risk of bloodstream infection and endocarditis, given their reduced potential for biofilm accumulation, thrombus formation, disturbances in flow dynamics, and interference with heart valves. Unlike their transvenous counterparts, leadless pacemakers eliminate the need for a subcutaneous pocket; their smaller surface area and intracardiac placement offer inherent advantages in terms of infection and endocarditis prevention [3].

In the context of this comprehensive review, it is evident that the risk of pericardial effusion tends to be higher in leadless pacemakers when compared to their transvenous counterparts. Out of the eight studies encompassing this particular outcome, five studies documented a notably greater incidence of pericardial effusion in patients with leadless pacemakers. It is important to note, however, that statistical significance in this regard was observed in only one of these studies [13]. Pooling the data from these studies, the collective incidence of pericardial effusion within the leadless pacemaker group was found to be 1.12%, in contrast to the 0.45% incidence observed in the transvenous pacemaker group. While these figures suggest a disparity, it is worth considering that the absolute difference in incidence is not substantially pronounced. Therefore, to draw more conclusive insights into this outcome, additional comprehensive studies are warranted. Given the novelty of leadless pacemaker implantation, it is plausible that a learning curve associated with this relatively new technology might contribute to the observed increase in cardiac injury, including pericardial effusion. The process of implanting leadless pacemakers likely demands a certain level of procedural mastery that could influence complication rates. Furthermore, it is conceivable that local irritation stemming from the tines that anchor the leadless pacemaker to the myocardium could potentially lead to pericardial inflammation, thus contributing to the observed trend of increased pericardial effusion [28].

A crucial aspect that will shed further light on this matter is the long-term follow-up of patients. The leadless pacemaker offers advantages over transvenous pacemakers due to its reduced complications and enhanced safety and effectiveness. However, previous clinical studies on transvenous lead placements have shown that placing the pacemaker in the mid-septum, as opposed to the apex, significantly reduces the risk of pericardial tamponade and myocardial perforation. Similarly, research involving the midseptal implantation of the Micra leadless pacemaker demonstrates increased safety, a narrower paced QRS duration, and decreased susceptibility to cardiac perforation compared to placing it in the right ventricular apex [29]. The mid-septum of the right ventricle is an ideal location for Micra leadless pacemakers due to its lower risk of myocardial perforation, attributed to the presence of more myocardial trabeculae and columnar carneae compared to the right ventricular apex and outflow tract [30]. This holds particularly true at the center of the septum. To enhance the precision of implantation, a study involving the ventriculography-guided deployment of Micra pacemakers within the septum was conducted. This technique proved to be significantly more accurate (93.8%) than the standard method of implantation (48.1%). Additionally, while maintaining adequate pacing parameters, the use of right ventriculography substantially reduced fluoroscopy exposure time and radiation dose [29].

In this study, we merged data from current comparative research to assess dissimilarities in complications between conventional pacemakers and leadless pacemakers. Our analysis demonstrated that leadless pacemakers exhibit an edge in terms of complications related to leads and the pocket site. However, regarding other complications, disagreements between leadless pacemakers and traditional pacemakers persist. Concurrently, it is important to acknowledge that due to the absence of extensive follow-up data for leadless pacemakers, our grasp of their long-term safety concerning complications remains undisclosed. This underscores a significant knowledge gap concerning prolonged complications associated with these modern devices.

The present meta-analysis has certain limitations. All included studies were observational in nature and therefore confounding variables and other biases could not be excluded. Secondly, most of the included studies used Micra as the leadless pacemaker and therefore we cannot extrapolate these findings to AVEIR leadless pacemakers. Lastly, due to a lack of patient-level data, we were not able to perform subgroup analysis. In the future, it is essential to conduct additional studies to gain a deeper understanding of the populations that will benefit most from leadless and transvenous pacemakers.

Conclusions

In conclusion, this meta-analysis offers valuable insights into the safety profiles of leadless pacemakers compared to transvenous pacemakers. The endpoint analysis strongly supports the notion that leadless pacemakers present a lower risk of total complications, device-related complications, pneumothorax, and endocarditis when compared to transvenous pacemakers. Additionally, the observed reduction in the risk of reintervention within the leadless pacemaker group underscores its potential clinical benefits. It is important to acknowledge the limitations arising from the lack of extensive long-term follow-up data for leadless pacemakers. As technology evolves, continued research will be essential in uncovering the full spectrum of prolonged complications associated with these devices.

The authors have declared that no competing interests exist.

Author Contributions

Acquisition, analysis, or interpretation of data:  Sujith K. Palleti, Maham Babur, Anurag Rawat, Ugonna A. Nwoko, Tayyaba J. Khan

Drafting of the manuscript:  Sujith K. Palleti, Maham Babur, Anurag Rawat, Mahesh Gangannapalle, Ugonna A. Nwoko, Obinna Monday, Arun Kumar Mandal, Tayyaba J. Khan

Concept and design:  Mahesh Gangannapalle, Obinna Monday, Arun Kumar Mandal

Critical review of the manuscript for important intellectual content:  Obinna Monday

References

  • 1.Guidelines for heart pacemaker therapy (Article in German) Lemke B, Nowak B, Pfeiffer D. Z Kardiol. 2005;94:704–720. doi: 10.1007/s00392-005-0269-3. [DOI] [PubMed] [Google Scholar]
  • 2.Indications for pacemakers, implantable cardioverter-defibrillator and cardiac resynchronization devices. Samii SM. Med Clin North Am. 2015;99:795–804. doi: 10.1016/j.mcna.2015.02.008. [DOI] [PubMed] [Google Scholar]
  • 3.Comparison between leadless and transvenous single chamber pacemaker therapy in a referral centre for lead extraction. Zucchelli G, Tolve S, Barletta V, et al. J Interv Card Electrophysiol. 2021;61:395–404. doi: 10.1007/s10840-020-00832-9. [DOI] [PubMed] [Google Scholar]
  • 4.Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Kirkfeldt RE, Johansen JB, Nohr EA, Jørgensen OD, Nielsen JC. Eur Heart J. 2014;35:1186–1194. doi: 10.1093/eurheartj/eht511. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Udo EO, Zuithoff NP, van Hemel NM, de Cock CC, Hendriks T, Doevendans PA, Moons KG. Heart Rhythm. 2012;9:728–735. doi: 10.1016/j.hrthm.2011.12.014. [DOI] [PubMed] [Google Scholar]
  • 6.Totally self-contained intracardiac pacemaker. Spickler JW, Rasor NS, Kezdi P, Misra SN, Robins KE, LeBoeuf C. J Electrocardiol. 1970;3:325–331. doi: 10.1016/s0022-0736(70)80059-0. [DOI] [PubMed] [Google Scholar]
  • 7.A leadless intracardiac transcatheter pacing system. Reynolds D, Duray GZ, Omar R, et al. N Engl J Med. 2016;374:533–541. doi: 10.1056/NEJMoa1511643. [DOI] [PubMed] [Google Scholar]
  • 8.FDA approves first leadless pacemaker to treat heart rhythm disorders. [ Aug; 2023 ]. 2016. https://www.fda.gov/newsevents/newsroom/pressannouncements/ucm494417.html https://www.fda.gov/newsevents/newsroom/pressannouncements/ucm494417.html
  • 9.Percutaneous implantation of an entirely intracardiac leadless pacemaker. Reddy VY, Exner DV, Cantillon DJ, et al. N Engl J Med. 2015;373:1125–1135. doi: 10.1056/NEJMoa1507192. [DOI] [PubMed] [Google Scholar]
  • 10.Chronic performance of a leadless cardiac pacemaker: 1-year follow-up of the LEADLESS trial. Knops RE, Tjong FV, Neuzil P, et al. J Am Coll Cardiol. 2015;65:1497–1504. doi: 10.1016/j.jacc.2015.02.022. [DOI] [PubMed] [Google Scholar]
  • 11.Leadless or conventional transvenous ventricular permanent pacemakers: a nationwide matched control study. Bodin A, Clementy N, Bisson A, Pierre B, Herbert J, Babuty D, Fauchier L. J Am Heart Assoc. 2022;11:0. doi: 10.1161/JAHA.122.025339. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Two-year outcomes of leadless vs. transvenous single-chamber ventricular pacemaker in high-risk subgroups. Boveda S, Higuera L, Longacre C, Wolff C, Wherry K, Stromberg K, El-Chami MF. Europace. 2023;25:1041–1050. doi: 10.1093/europace/euad016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Comparative study of acute and mid-term complications with leadless and transvenous cardiac pacemakers. Cantillon DJ, Dukkipati SR, Ip JH, et al. Heart Rhythm. 2018;15:1023–1030. doi: 10.1016/j.hrthm.2018.04.022. [DOI] [PubMed] [Google Scholar]
  • 14.Leadless vs. transvenous single-chamber ventricular pacing in the Micra CED study: 2-year follow-up. El-Chami MF, Bockstedt L, Longacre C, et al. Eur Heart J. 2022;43:1207–1215. doi: 10.1093/eurheartj/ehab767. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Conventional single-chamber pacemakers versus transcatheter pacing systems in a "real world" cohort of patients: a comparative prospective single-center study. Martinez-Sande JL, Garcia-Seara J, Gonzalez-Melchor L, Rodriguez-Mañero M, Baluja A, Fernandez-Lopez XA, Juanatey JRG. Indian Pacing Electrophysiol J. 2021;21:89–94. doi: 10.1016/j.ipej.2020.12.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Leadless pacemaker implantation: a feasible and reasonable option in transcatheter heart valve replacement patients. Moore SK, Chau KH, Chaudhary S, et al. Pacing Clin Electrophysiol. 2019;42:542–547. doi: 10.1111/pace.13648. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Comparison in clinical outcomes between leadless and conventional transvenous pacemaker following transcatheter aortic valve implantation. Okuyama K, Izumo M, Sasaki K, et al. https://www.hmpgloballearningnetwork.com/site/jic/articles/comparison-clinical-outcomes-between-leadless-and-conventional-transvenous-pacemaker-following-transcatheter-aortic-valve-implantation. J Invasive Cardiol. 2020;32:400–404. doi: 10.25270/jic/20.00076. [DOI] [PubMed] [Google Scholar]
  • 18.Safety of leadless pacemaker implantation in the very elderly. Pagan E, Gabriels J, Khodak A, Chang D, Beldner S, Epstein LM, Willner J. Heart Rhythm. 2020;17:2023–2028. doi: 10.1016/j.hrthm.2020.05.022. [DOI] [PubMed] [Google Scholar]
  • 19.Leadless pacemaker versus transvenous single-chamber pacemaker therapy: peri-procedural aspects, utilization of medical resources and patient acceptance. Palmisano P, Guido A, Panico V, et al. Expert Rev Med Devices. 2021;18:483–491. doi: 10.1080/17434440.2021.1921573. [DOI] [PubMed] [Google Scholar]
  • 20.Rate and nature of complications with leadless transcatheter pacemakers compared with transvenous pacemakers: results from an Italian multicentre large population analysis. Palmisano P, Facchin D, Ziacchi M, et al. Europace. 2023;25:112–120. doi: 10.1093/europace/euac112. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Incidence of pacing-induced cardiomyopathy in pacemaker-dependent patients is lower with leadless pacemakers compared to transvenous pacemakers. Sanchez R, Nadkarni A, Buck B, et al. J Cardiovasc Electrophysiol. 2021;32:477–483. doi: 10.1111/jce.14814. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Clinical outcomes of non-atrial fbrillation bradyarrhythmias treated with a ventricular demand leadless pacemaker compared with an atrioventricular synchronous transvenous pacemaker— a propensity score-matched analysis. Sasaki K, Togashi D, Nakajima I, Suchi T, Nakayama Y, Harada T, Akashi YJ. Circ J. 2022;86:1283–1291. doi: 10.1253/circj.CJ-21-0889. [DOI] [PubMed] [Google Scholar]
  • 23.The feasibility of leadless pacemaker implantation for super elderly patients. Tachibana M, Banba K, Matsumoto K, Ohara M. Pacing Clin Electrophysiol. 2020;43:374–381. doi: 10.1111/pace.13894. [DOI] [PubMed] [Google Scholar]
  • 24.Real-world experience with leadless cardiac pacing. Vaidya VR, Dai M, Asirvatham SJ, et al. Pacing Clin Electrophysiol. 2019;42:366–373. doi: 10.1111/pace.13601. [DOI] [PubMed] [Google Scholar]
  • 25.Safety and feasibility of leadless pacemaker in patients undergoing atrioventricular node ablation for atrial fibrillation. Yarlagadda B, Turagam MK, Dar T, et al. Heart Rhythm. 2018;15:994–1000. doi: 10.1016/j.hrthm.2018.02.025. [DOI] [PubMed] [Google Scholar]
  • 26.2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Glikson M, Nielsen JC, Kronborg MB, et al. Eur Heart J. 2021;42:3427–3520. doi: 10.1093/eurheartj/ehab699. [DOI] [PubMed] [Google Scholar]
  • 27.Comparative assessment of safety with leadless pacemakers compared to transvenous pacemakers: a systemic review and meta-analysis [PREPRINT] Shtembari J, Shrestha DB, Awal S, et al. J Interv Card Electrophysiol. 2023 doi: 10.1007/s10840-023-01550-8. [DOI] [PubMed] [Google Scholar]
  • 28.The learning curve associated with the implantation of the Nanostim leadless pacemaker. Tjong FV, Beurskens NE, Neuzil P, et al. J Interv Card Electrophysiol. 2018;53:239–247. doi: 10.1007/s10840-018-0438-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Right ventriculography improves the accuracy of leadless pacemaker implantation in right ventricular mid-septum. Li Y, Xing Q, Xiaokereti J, et al. J Interv Card Electrophysiol. 2023;66:941–949. doi: 10.1007/s10840-022-01399-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Assessing safety of leadless pacemaker (MICRA) at various implantation sites and its impact on paced QRS in Indian population. Sharma P, Singh Guleria V, Bharadwaj P, Datta R. Indian Heart J. 2020;72:376–382. doi: 10.1016/j.ihj.2020.08.001. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Cureus are provided here courtesy of Cureus Inc.

RESOURCES