Sensitivity, specificity, and area under the curve values identified by cutoff points of different OCT-based corneal thickness parameter sets to differentiate eyes with suspect keratoconus from normal corneas and keratoconus from normal ones.

Pachymetry (μm) Cutoff value AUC Sensitivity (%) Specificity (%)
Normal vs. SKC Normal vs. KC Normal vs. SKC Normal vs. KC Normal vs. SKC Normal vs. KC Normal vs. SKC Normal vs. KC
Min CT (0_2) ≤526 ≤478 0.75 0.88 88.37 82.86 57.4 83.6
Avg CT (0_2) ≤535 ≤497 0.74 0.85 86.05 82.35 54.1 78.7
Max CT (0_2) ≤557 ≤526 0.75 0.8 88.37 76.47 50.8 73.8
Min-Max CT (0_2) ≤−26 ≤−32 0.68 0.94 53.49 94.12 86.9 93.4
Min CT (2_5) ≤537 ≤493 0.75 0.88 95.35 91.18 50.8 77.1
Avg CT (2_5) ≤581 ≤542 0.73 0.78 100 91.18 36.1 60.7
Max CT (2_5) ≤613 ≤608 0.72 0.72 90.7 91.18 49.2 50.8
Min-Max CT (2_5) ≤−77 ≤−77 0.61 0.89 41.86 91.18 41.9 82
S-I CT (2_5) >1 >15 0.58 0.77 95.35 90.62 23 59
SN-IT CT (2_5) >25 >35 0.66 0.83 74.42 78.12 50.8 82
CCT ≤533 ≤491 0.76 0.87 86.05 82.35 55.7 83.6

Min = minimum; CT = corneal thickness; Avg = average; Max = maximum; Min-Max = minimum-maximum; S-I = superior-inferior; SN-IT = superonasal-inferotemporal; CCT = central corneal thickness; AUC = area under the receiver operating characteristic curve; SKC = suspect keratoconus; KC = keratoconus.