TABLE 4.
Clinical agreement analysis of fibrinolysis
ROTEM delta | |||
---|---|---|---|
EXTEM ML>15%a (fibrinolysis+) | EXTEM ML≤15%a (fibrinolysis−) | Total | |
QStat Cartridge | |||
CSL<90% (fibrinolysis+) | 61 | 27 | 88 |
CSL≥90% (fibrinolysis−) | 8 | 267 | 275 |
Total | 69 | 294 | 363 |
Positive agreement | |||
88.4% (78.4%, 94.9%) | — | — | — |
Negative agreement | |||
90.8% (86.9%, 93.9%) | — | — | — |
Overall agreement | |||
90.3% (86.9%, 93.2%) | — | — | — |
Note: Agreement is expressed as percentage (95% CI).
Measured at 60 minutes after clot initiation.
Abbreviation: CSL, clot stability to lysis.