Results of [68Ga]Ga-EMP100 stability study in final vial

Test Method Specifications Batch 1 Batch 2 Batch 3
Aspect Visual Clear, colourless solution Yes Yes Yes
pH value pH-strips 4.0–8.0 6.0 6.0 6.0
Stability over 3 h (RCP %) Radio HPLC [68Ga]Ga-EMP100 > 95,00% 98.63% 98.87% 99.52%
Free [68Ga]GaCl3 ≤ 5% 1.37% 1.13% 0.48%
Radio TLC AcNH4 1 M Methanol (V/V) [68Ga]Ga-EMP100 > 95% 99.85% 99% 99.91%
[68Ga]Ga-colloidal ≤ 5% 0. 15% 1.00% 0.09%
RCP (overall) [68Ga]Ga-EMP100 (overall) > 95.00% 99.08% 97.87% 99.43%
Sterility Inoculation 14 days (Ph Eur 20601) Sterile Sterile Sterile Sterile