ABSTRACT
Assuring that cell therapy products are safe before releasing them for use in patients is critical. Currently, compendial sterility testing for bacteria and fungi can take 7–14 days. The goal of this work was to develop a rapid untargeted approach for the sensitive detection of microbial contaminants at low abundance from low volume samples during the manufacturing process of cell therapies. We developed a long-read sequencing methodology using Oxford Nanopore Technologies MinION platform with 16S and 18S amplicon sequencing to detect USP <71> organisms and other microbial species. Reads are classified metagenomically to predict the microbial species. We used an extreme gradient boosting machine learning algorithm (XGBoost) to first assess if a sample is contaminated, and second, determine whether the predicted contaminant is correctly classified or misclassified. The model was used to make a final decision on the sterility status of the input sample. An optimized experimental and bioinformatics pipeline starting from spiked species through to sequenced reads allowed for the detection of microbial samples at 10 colony-forming units (CFU)/mL using metagenomic classification. Machine learning can be coupled with long-read sequencing to detect and identify sample sterility status and microbial species present in T-cell cultures, including the USP <71> organisms to 10 CFU/mL.
IMPORTANCE
This research presents a novel method for rapidly and accurately detecting microbial contaminants in cell therapy products, which is essential for ensuring patient safety. Traditional testing methods are time-consuming, taking 7–14 days, while our approach can significantly reduce this time. By combining advanced long-read nanopore sequencing techniques and machine learning, we can effectively identify the presence and types of microbial contaminants at low abundance levels. This breakthrough has the potential to improve the safety and efficiency of cell therapy manufacturing, leading to better patient outcomes and a more streamlined production process.
KEYWORDS: T-cells, adventitious agents, machine learning, sterility
INTRODUCTION
Cell therapies are increasingly prevalent in the treatment of incurable diseases. For example, chimeric antigen receptor T-cells (CAR-T) are used for the treatment of hematologic malignancies (1). Ongoing work with human pluripotent stem cells (hPSCs) is targeted to treat Parkinson’s and age-related macular degeneration (AMD) (2), while mesenchymal stromal cells (MSCs) are being developed for immunomodulatory treatments (3). Compendial sterility methods based on microbial growth are laborious and slow, and faster methods are required to guide clinical management (4). Rapid testing methodologies could be an important tool for decreasing the time that a patient must wait from initial leukapheresis to application of the cell therapy. Depending on a patient’s current health status, the patient may not be able to afford delays in the application of a potentially lifesaving therapy. Thus, a reduction of release testing time will ensure the timely and safe delivery of lifesaving cell therapies leading to improved patient outcomes.
Current good manufacturing practice for microbial safety has been developed from experience in recombinant protein manufacturing, where standard practices include three pillars of safety: (i) identifying appropriate sterile raw materials, (ii) testing of cell banks and in-process microbe testing for materials, and (iii) inclusion of process steps to inactivate and remove undetected microbes (5). Cell therapy manufacturers are currently only able to use the first two pillars, as the product cannot be terminally sterilized. For testing, as such, the main approach to microbial safety is with compendial sterility tests. For example, test samples are inoculated into multiple growth media that support proliferation of aerobic and anaerobic organisms, as well as using the plate-count method and membrane filtration (6). Validation of compendial methods includes testing with USP <71> organisms. For a list of USP <71> organisms used in this study, see Table 1. The procedures for ensuring that the tests give valid results are codified in the United States Pharmacopoeia/European Pharmacopoeia (USP/EP). These methodologies can detect contamination events but cannot determine the species identity, which requires additional time-consuming follow-up by the manufacturer for failure investigation. The USP <71> tests suffer from false negatives at low (<30 CFU) contaminant concentration when organisms fail to grow to the point of visible detection in the allocated time (4). Alternative sterility testing approaches have been developed with lower limits of detection, such as the BacT/ALERT system, which is an automated system that works by colorimetric change of CO2 level evolution monitored every 15 min. The BacT/ALERT system has a limit of detection of <10 CFU for sample volumes of 0.5–10 mL (7, 8).
TABLE 1.
Culture conditions a
| Organism | Medium | Temperature | Culture time | Subculture time | Respiration |
|---|---|---|---|---|---|
| Cutibacterium acnes (ATCC-6919) | BHIB | 33ºC | 5 days | 2 days | Anaerobic |
| Klebsiella pneumoniae (KP1) | LB | 37ºC | O/N | 2 h | Aerobic |
| Escherichia coli (K12) | LB | 37ºC | O/N | 2 h | Aerobic |
| Pseudomonas aeruginosa PAO1 (ATCC BAA47) | LB | 37ºC | O/N | 2 h | Aerobic |
| Candida albicans (ATCC 10231) | YPD | RT | 2 days | 2 h | Aerobic |
| USP <71> organisms | |||||
| Staphylococcus aureus subsp. aureus (ATCC 6538) | LB | 37ºC | O/N | 2 h | Aerobic |
| Pseudomonas aeruginosa (ATCC 9027) | LB | 37ºC | O/N | 2 h | Aerobic |
| Bacillus subtilis subsp. spizizenii (ATCC 6633) | LB/ BHIB | 33ºC | O/N | 2 h | Aerobic |
| Clostridium sporogenes (ATCC 19404) | MRCB | 37ºC | 2 days | 2 h | Anaerobic |
| Candida albicans (ATCC 10231) | YPD | RT | 2 days | 2 h | Aerobic |
| Aspergillus brasiliensis (ATCC 16404) b | Aerobic | ||||
MRCB, modified reinforced clostridial broth; YPD, yeast extract–peptone–dextrose; BHIB, brain heart infusion broth; LB, Luria-Bertani broth; RT, room temperature; O/N, overnight.
Aspergillus brasiliensis was purchased as a pellet from ATCC.
The performance of both standard compendial tests and BacT/ALERT system can be further improved upon. A thorough analysis by England and colleagues reported an average 40 h time to detection for the BacT/ALERT compared to 53 h for compendial testing. Within the acceptable runtime (<144 h), 100/118 (84.7%) tested isolates were detected by the compendial USP <71> methods. When running the sterility tests using the BacT/ALERT alongside supplemental Sabouraud dextrose agar (SDA) plates were incubated up to 360 h (15 dys), 100% of fungi were detected, while USP <71> detected 95.8% of contamination events. The authors report that the majority of fungal isolates were detected within 144 h for the manual USP <71> methods; however, with automated systems fungal detection could take longer to resolve and were only detected by manual inspection after 360 h of incubation. Additionally, the authors report that some bacterial species could not be detected without the infusion of fresh human blood into the culture, for example, Haemophilus influenzae and Cutibacterium acnes using the BacT/ALERT after 270 h. Furthermore, isolates grown at low inocula <30 CFU proved difficult to detect within the accepted time frame (<96 h, bacteria; <144 h, fungi) (4). In summary, within the accepted time frame for bacteria (<96 h), 83.1% of isolates were detected with the USP <71> compendial approach and 87% for the 32.5°C BacT/ALERT. Meanwhile for fungal isolates (<144 h), 87.8% of isolates were detected with the USP <71> compendial approach and 63.4% for the 32.5°C BacT/ALERT (4). Given the described time constraints, as well as reduced sensitivity to filamentous fungal species, we propose an alternative methodology to detect and identify microbial contaminants through long-read MinION sequencing.
In the case of autologous CAR-T therapy products, each manufacturing lot is prepared for a single patient. Each lot of a cell therapy must pass through manufacturing release testing. This poses a limitation with regard to available test material—using a minimal volume for sterility assessment is important to maximize cells available for the patient (with guidelines described in USP <1071>) (9), while minimizing additional manufacturing costs. The products will have in-process and final product testing for sterility, endotoxin levels, mycoplasma, and replication competent virus (10). As CAR-T treatments become more readily available, rapid throughput and multiplexing will become increasingly necessary for the analysis of large numbers of scaled-out autologous cell therapy samples.
We propose an approach that identifies contaminants by the presence of either bacterial or fungal DNA. Long-read MinION sequencing offers a rapid, simple-setup, real-time reads (contaminant identification before sequencing completion) and low-cost approach to DNA sequencing (11, 12). Compared to Illumina platforms, which can take days to weeks to complete sequencing and bioinformatics analyses, the nanopore sequencing can provide results in less than 24 h (13, 14). Another advantage with the long-read MinION sequencer is greater taxonomic resolution than amplicon sequencing using the Illumina MiSeq system at the species level (15). This positions an amplicon sequencing approach using the long-read MinION combined with a full-size metagenome database favorably for managing rapid release testing.
Samples processed for T-cell therapy release tests will present at low sample volume and low contaminant DNA concentration. The long MinION offers long-read sequencing with two potential routes for identification of contaminants: direct DNA and amplicon-sequencing. Direct DNA sequencing comes with the constraint of high background host DNA that will be sequenced; this can make detection of the relatively lower concentration microbial DNA harder. Autologous cell therapies will present as low volume, low contaminant concentration samples, thus by necessity we took the PCR-based amplicon approach, which has the additional benefit of reducing microbe to host noise before downstream computational screening for host reads. In the case of bacteria, the highly conserved prokaryote 16S ribosomal RNA region is widely used in the metagenomics field with either the amplicon or shotgun metagenomics approach to perform microbial identification at the species level (16 – 18), while the equally highly conserved eukaryote 18S–28S operon is used for fungal classification.
RESULTS
Design of the DNA read analysis pipeline
The overview of the pipeline is described in Fig. 1A, which processes reads derived from the spiked samples and derives a decision as to whether the sample is contaminated or not. The methodology for data acquisition, model training, and validation, followed by the decision matrix are described in Fig. 1B. In this analytical sterility study, we detected low levels of microbial contamination (10 CFU/mL) with high specificity and accuracy. The workflow we developed is compatible with low sample volume and rapid turnaround that meets patient needs and preserves sample shelf-life.
Fig 1.
Pipeline workflow. (A) Machine learning pipeline overview: microbial contaminants were prepared, sequenced, and the reads processed. The metagenomic classification data, overall run read quality data, predicted species quality data, and time to next read data were combined into a single table of features. A decision tree gradient boosting classifier algorithm XGBoost was deployed to assess contaminant sterility status, for more information see Machine learning pipeline in Materials and Methods or github. (B) Bacteria (gram positive or gram negative) or fungus (yeast) are spiked into PBS-washed cultured T-cells. The process is repeated threefold with and without T-cells using cells from a different passage and separately cultured microbes. DNA is extracted using mechanical lysis, buffers, and magnetic beads. DNA is amplified using targeted rDNA primers for the 16S region and 18S–28S region. (C) Sequencing analysis pipeline: sequenced base called reads were cleaned and host reads removed. Remaining reads were classified against the combined viral, fungal, and bacterial database using Centrifuge and High-Speed BLAST. Classified reads along with other data were provided to the machine learning pipeline for sample contamination status analysis.
Samples were comprised of negative samples (T-cell only, medium-only, and cell-free medium) and target organisms (fungal and bacterial species) (Table 1). The table contains information about the added species, barcode used, sequencing time duration, and the nanopore kit used. Negative samples were prepared alongside inoculated samples, which were run in both direct and amplicon-sequencing runs. Only the amplicon-sequencing data is presented here.
Inclusion of low-quality reads improved correctly classified read count
Our goal was to detect microbes at low concentrations (≤10 CFU). Detection therefore required generation of additional genetic information or a means of improving the pool of reads for assessment. We did not screen by phred quality score (Q) as we sought to identify contaminant read species identity. Consequently, there were a median of 6.9% additional reads across all samples. Mean read length was compared between high-quality (Q ≥ 7) and high-quality plus low-quality reads (Q > 0) for correctly classified true positive reads (Fig. 2A). For 16S amplicon-enriched species, the read length did not change greatly between the two groups, at around 1450 bp, while 18S–28S had smaller mean fragment size (Fig. 2A).
Fig 2.
Assessment of incorporating low-quality reads in sequence classification. (A) Additional low-quality reads incorporated into data analysis (N = 3) for 10 CFU/mL samples. Mean read length for correctly classified (true positives) reads with and without use of the low-quality reads were depicted in blue and black, respectively. Incorrectly assessed reads (misclassified) were depicted in cyan and gray. (B) Summary table of high-quality reality reads compared to using all reads for the microbes alone and microbes spiked into T-cells at 10 CFU/mL. Read numbers are for reads assigned to the correct spiked organism and represent a subset of all sequenced samples.
We calculated summary statistics for a subset of low concentration (10–100 CFU/mL) contaminated samples compared to microbe-only samples. Generally, we observed more microbe reads in the microbe-only cultures compared to samples containing host reads too; this was independent of low-quality read inclusion (Fig. 2B). For example, by incorporating the low-quality reads, Candida albicans was detectable in two of the three samples contaminated at 10 CFU, while when using high-quality reads alone, C. albicans was detected in only one of the analyzed samples. A side-by-side comparison of high-quality vs any-quality reads revealed that by including the low-quality reads, one additional true positive sample was correctly identified; the overall read count increased for correctly predicted species. However, there was a concomitant increase in the number of misclassified species. As such, it is important to consider the use case before making the choice to include this additional source of reads.
Detection sensitivity and time to detection
Initially, we sought to understand the limit of detection for a single species, Pseudomonas aeruginosa PAO1. The rationale for selecting a limit of detection of 10 CFU/mL is based on the USP <1071> chapter discussing infectious dose. The document discusses that a contaminant detection of 100 CFU/mL would catch all viable organisms, as such we set 100 CFU/mL as our target limit of detection, with the aim of detecting at lower concentrations as a form of stringent validation (9). T-cells were spiked with serial 10-fold dilutions of P. aeruginosa to the lowest input of 10 and 100 CFU/mL. Our approach examined the utility for both direct shotgun sequencing and amplicon sequencing. Extracted DNA was processed for both direct and amplicon sequencing. We consistently achieved 10 CFU/mL detection (from 1 mL spike sample) with the amplicon approach (Table 2), while the limit of detection for direct sequencing was 1000 CFU/mL (data not shown). We then proceeded to contaminant detection for microbial cultures and T-cells for intentional contamination with C. acnes, Klebsiella pneumoniae, Escherichia coli, P. aeruginosa, or the USP <71> species C. albicans, Staphylococcus aureus subsp. aureus, P. aeruginosa, Bacillus subtilis subsp. spizizenii, and Clostridium sporogenes (Fig. 3A and B; Tables S1 to S6; Table S5 available at https://figshare.com/s/5ecffffa2578fd87977f for the Centrifuge summary data and Table S6 available at https://figshare.com/s/439ea9a88baf513c7b39 for the HS-BLASTn summary data). In all cases, we consistently detected the contaminant to 100 CFU/mL (Fig. 3A) and 10 CFU/mL (Fig. 3B) in microbially contaminated T-cell cultures. One organism that proved consistently difficult to amplify was K. pneumoniae, resulting in fewer reads per sample than other species. Others have reported difficulty with taxonomic resolution in Klebsiella when amplifying the 16S rRNA regions of V1V2, V3, and V6V7 with Nanopore. They observed misclassification to closely related genera and low count of total mapped reads (13).
TABLE 2.
Limit of detection and sensitivity for contaminant detection a
| Contaminant species | Limit of detection | Sensitivity (species detected in sample/total samples) |
|---|---|---|
| Pseudomonas aeruginosa (PAO1) | 10 CFU | 9/10 |
| Klebsiella pneumoniae (KP1) | 10 CFU | 3/7 |
| Cutibacterium acnes | 10 CFU | 4/6 |
| USP <71> organisms | ||
| Staphylococcus aureus | 10 CFU | 5/6 |
| Bacillus subtilis | 10 CFU | 6/6 |
| Clostridium sporogenes | 10 CFU | 6/6 |
| Pseudomonas aeruginosa (9027) | 10 CFU | 4/6 |
| Candida albicans | 10 CFU | 11/12 |
In the sensitivity column, the first value denotes total samples from amplicon sequencing split for combined spiked T-cell contaminant and microbial culture of that contaminant. Second value shows only the T-cell contaminant amplicon sequencing value.
Fig 3.
Microbes spiked into T-cell cultures. All samples were amplified with either 16S primers for bacterial species, or 18S–28S primers for fungal species. Samples were prepared for analysis as microbial cultures as well as simulated microbial contamination by the addition of microbes to activated T-cells at 100 CFU/mL (A) and 10 CFU/mL (B), pure culture spikes (gray) compared to contaminants spiked into T-cells (blue). Species tested were K. pneumoniae, P. aeruginosa PAO1, C. acnes, and the USP <71> species; C. albicans, B. subtilis, Clostridium sporogenes, S. aureus, P. aeruginosa 9027. Error bars are biological replicates (N = 3).
Our primary goal was to identify sample contamination as rapidly and accurately as possible. We were able to detect and identify contaminants within 24 h, including DNA extraction (2 h), PCR (2 h), sequencing (12 h), and the bioinformatics analysis (1 h). If the sequencing was shortened to 2 h, the time to detection was approximately 8 h per sample. However, this might not provide enough time to sequence enough reads dependent on the species present (in species that amplify well, e.g., B. subtilis, we observed enough reads within 2 h). Multiple biological samples can be processed simultaneously with multiplexed barcoding, allowing up to 12 samples to run on the same flow cell. This can be replicated fivefold for technical replicates per USP <1223> guidance (19), which can be run in parallel with the primer and culture medium negative controls. To further improve the time to detection, we evaluated different computational processing approaches. The read trimming and metagenomic classification were very time intensive. However, multiprocessing resulted in an average fourfold improvement in speed. A computer system with higher specifications than used here would see an even greater improvement in time to result with more simultaneous processes.
Machine learning pipeline and sample sterility status
The methods described in the pipeline allow for rapid and sensitive detection of low concentrations of contaminations when the contaminant is known. In addition to the proposed machine learning approach, all experiments would have plates cultured alongside the sequencing analysis. This would allow organisms that are detected as borderline to be captured by a secondary, albeit slower methodology. To detect unknown contaminants using an unbiased approach, we used machine learning to enable decision making regarding the sterility status of a sample. The metagenomic classifiers yield multiple potential species identities for perceived contaminants, which we initially attempted to deplete by using filters based on total minimum read count per species. Two separate pairs of models were generated for both data from Centrifuge and HS-BLASTn. The Centrifuge models assessing sterility status and whether a contaminant was correctly predicted are shown in Fig. 4A through C. Model analysis for the HS-BLASTn model is depicted on Fig. 4D through F. The classification report demonstrates a model that can identify sample contaminant status, while the model examining contaminant identity can find contaminant species (Fig. 4A and B, 4D and E). The model predictions from sample status (sterile: true negative or contaminated: true positive) and correctness of contaminant classification (correctly classified vs misclassified contaminant) were combined and a decision matrix was used to decide if a sample was sterile, contaminated, potentially contaminated, or showed signs of most likely being sterile with background noise signal (Fig. 4C and E). For Centrifuge, 65.0% of negative controls were correctly labeled as sterile, 64.2% of 10 CFU/mL samples were correctly assessed as contaminated, and 77.3% of 100/mL CFU samples were identified as contaminated. Overall, while similar, the BLAST models perform better than Centrifuge on the final assessment for sterility status, with reduced ambiguity in a final decision on whether a negative control sample is sterile (75.0%) or whether a spiked sample was contaminated (10 CFU/mL; 83.9%, 100 CFU/mL; 96.0%). Overall, the 95% confidence intervals for the combined Centrifuge models were 80.67% ± 1.06% and 76.75% ± 1.03% for HS-BLASTn.
Fig 4.
Machine learning XGBoost model performance, sample, and prediction contaminant status. (A–C) Model performance statistics from data from centrifuge metagenomic classifier used to generate two XGBoost classification models. (D–F) Model performance statistics from data from high-speed BLASTn metagenomic classifier used to generate two XGBoost classification models. (A and D) Confusion matrix and classification report for model assessing sample sterility status. (B and E) Confusion matrix and classification report for model assessing whether a predicted contaminant is a correctly classified contaminant or misclassified. (C and F) All spikes and negative control model predictions were assessed for prediction accuracy regarding whether the sample assayed is sterile. Black bars depict samples assigned as likely contaminated, blue bars depict samples identified as sterile, while gray depicts samples where the algorithm had difficulty assigning a decision of either sterile or contaminated. CFU: colony-forming units. Sample status is defined as sterile: true negative or contaminated: true positive. Correct contaminant classification is defined as a true positive contaminant vs a misclassified contaminant.
DISCUSSION
We have demonstrated that amplicon sequencing can be used for low sample volume (<1 mL), low concentration (≤10 CFU) microbe detection in intentionally contaminated T-cell cultures. Our goal was to detect a contaminant with high specificity and in a short turnaround time to enable rapid sterility release testing for cell therapy products. This method is designed to report the presence of very low concentration contaminant organisms and does not yet allow for accurate organism quantification. We designed our sterility detection to be faster than current commercial detection times for low concentration contamination events of <100 CFU/mL to within <24 h, which compares with the historical detection times of 7–14 days using BacT/ALERT (20, 21) and more recent detection times of 40 h (4, 9), which is in line with FDA sterility guidelines (22, 23). For bacterial and fungal detection, we used an amplicon approach that significantly increased the contaminant signal from DNA extraction. Use of the Nanopore 16S amplicon kit allowed us to generate consistent full-length 16S fragments. Previously, full-length amplicons have been shown to be comparable in reliably identifying genera on Nanopore and MiSeq technologies, while Nanopore can operate at lower run costs (50 USD/sample) (14). However, when generating similar fragments using the 18S–28S primers, we did not observe similar read length consistency. This is likely due to use of the transposase technology in the Rapid Barcoding Sequencing Kit (SQK-RBK004) that cleaves the PCR product for barcode ligation. The problem could be solved by using the Nanopore ligation kit (SQK-LSK110); however, this approach increased sample preparation time by up to 5 h. This limited our ability to sequence the entire region as a single strand, however as we have demonstrated, we are still able to detect C. albicans to 10 CFU.
During the sequencing analysis, we observed background genomic material in the sterile media controls. We believe these are derived from DNA contaminants in the DNA extraction, PCR, and library preparation kits. The presence of a kitome has been widely described in the literature, including from the DNeasy PowerSoil Kit used in this work (24, 25). This is problematic because at low microbial concentration the signal to noise ratio becomes elevated, making identifying correctly classified contaminants difficult and increasing read misclassification. False positive predictions will potentially disrupt the ability to deliver the cell therapy product to a patient in the needed time frame because of additional time required to carry out failure investigation to assess actual lot sterility status, in addition to incurring further manufacturing costs. The issue of high noise at low concentration reinforces the need to use negative controls to identify the contamination background as well as control the number of PCR cycles, as previous studies have shown that 20 cycles were too few while 40 cycles will amplify the noise (26). Amplification efficiency was a potential problem we observed in certain species. We assayed 25, 27, 29, and 36 cycles and found 29 to be the optimal value for our use case in terms of reads sequenced for total cycle count, though others reported 30 as optimal (14). We observe a strong correlation between CFU and OD, which was corroborated by digital PCR. We conclude that the dilutions accurately achieved the target cell concentration and that the read count variation most likely occurs during library preparation or sequencing itself. We propose further investigation with the new 10.4.1. chemistry to better understand this observation.
One of the key limitations of this approach is the use of DNA for detection of adventitious agents. This does not provide a definitive answer to the viability of the detected contaminant and is part of the reason a deeper understanding of the kitome is required. Alternatively, sequencing of RNA after reverse transcription and rDNA amplification would provide a means to test for contamination directly. We are investigating background species that are detected during the sequencing, they are characterized by weak signal and may be on the limit of perception due to the low read count derived from an already very low starting sample concentration and volume. Extensive testing and validation of potential background contaminant species could reduce noise detection, we propose performing multiple technical replicates per sample, as well as running multiplexed positive and negative standard controls in every sterility test.
To optimize sensitivity and pipeline runtime, we examined the choice of input species to build the metagenomic databases. This is mostly an issue with the NCBI fungal refseq, which at the time of database construction was confined to only 12 complete genomes. To expand the range of species, we incorporated incomplete genomes for both the BLASTn and Centrifuge databases. However, the bacterial reference database is very large and contains many genome replicates for common bacterial genera Pseudomonas (613) and Escherichia (1100), as well as species including P. aeruginosa (219) and E. coli (1078). As such, to reduce the database size, we retained only genomes with a “complete genome” and removed those labeled as “chromosome” or “plasmid.” This helped improve runtime and reduce database size differences between the fungal and bacterial databases. The only downside we observed by shrinking the bacterial database was the loss of species-level inference as observed with C. sporogenes, where C. butyricum was indicated instead. For future organisms, a living database could be constructed and periodically updated. Reference genome agnostic approaches could be used for unknown contaminants, such as read binning using the DNABERT approach described subsequently (27). After inclusion of C. sporogenes, we continued to observe that C. butyricum was preferentially predicted with 42% more reads on average for HS-BLASTn (and not identified by Centrifuge).
We subtracted host sequences using the human genome to reduce noise in the analysis pipeline and improve pipeline efficacy. With this approach, 98% of total reads were identified as host reads and removed, illustrating the challenge of detecting low-level bacterial or fungal contaminants in samples of cell culture material. Our study focused on amplicon-sequencing, which enriched for target sequences and reduced the ratio of host reads to target reads. We implemented host read removal to improve sensitivity and enable a non-amplicon approach going forward. Nevertheless, host read removal is useful for improving microbial identification, as has been described previously in viral contaminant detection (28). Previously, host read removal has been shown to improve viral contig assembly while reducing the number of assembled contigs so as to allow use of metagenomic classifiers such as BLASTn (29). In addition to removing host reads, removal of background reads by sequencing healthy and sterile T-cells, then subtracting the reads from the final pool could further improve sensitivity by reducing misclassification events (29). The 16S and 18S–28S amplicon sequencing approach is more successful than direct metagenome sequencing, especially at contaminant concentrations ≤1,000 CFU/mL. Previous studies have demonstrated that amplicon-based approaches can introduce inaccuracy and misclassification, for example, 0.93% of reads are misclassified by EPI2ME at the genus level (16) and that there is a 2.09% misclassification for the NCBI 16S reference database at genus level (16). For cell therapy sterility, the improved contaminant detection at the cost of a small increase to misclassification is an acceptable outcome.
Two key limitations for model development are the volume of available sequencing analysis data and the high sample-to-sample variance in the data distribution. For example, we observed sources of variation from read count, read quality, and experimental runtime. The implications for the observed variance include difficulty in applying the machine learning model to new samples that may contain data with novel distributions. Solving the problem of noisy, unreliable data requires a robust and flexible model that can classify despite lack of knowledge on the data distribution for the evaluation samples. This includes selecting for features with minimal missing data and applying, as well as deploying data augmentation techniques such as addition of gaussian noise, generating squares, and taking log10 of the numerical data. Finally, we have applied regularization techniques during model development including optimizing L1 (lasso regression), L2 (ridge regression), and the gamma hyperparameters to mitigate overfitting. To improve issues with generating more similar and reliable samples that follow a specific data distribution would require further refinement of the DNA extraction and library prep, in addition to controlling for missing data points during feature selection. However, this would come at the cost of having a model that is less able to generalize to other cell types. We would like to highlight that the accuracy values we achieved of 95% should be taken as theoretical best-case scenario for the developed model. Further testing incorporating double-blinded and real-world samples will be necessary to examine performance. We would expect performance to be reduced under these circumstances as the new data will have changed distributions and may be imbalanced. We therefore stress that it is important to run positive and negative controls alongside all PCR runs during multiplexing.
Our analysis has primarily focused on generating many instances of contaminated samples. During production for cell therapy manufacturing, we envisage sample contamination events as rare. Consequently, future model optimization must be considered through this data imbalance, as observed in fraud detection for better anomaly prediction (30 – 32). Alternatively, it might be possible to replace the metagenome classifier and binary classifiers with a single step natural-language processing (NLP)-based approach using, for example, DNABERT (27). A NLP-based approach would complement the metagenomics analysis that we have already designed. Furthermore, it would give us a potential solution for the identification of unknown contaminated samples independent of whether they have been previously characterized.
In conclusion, we have developed a rapid (<24 h), reliable, sensitive, and specific long-read sequencing pipeline for the detection of microbes in T-cell therapies alongside a large data set of microbial organisms’ (USP <71> focus) low concentration samples in pure culture and spiked with T-cells. This is partly achieved by using machine learning to provide an unbiased, untargeted approach that permitted automated decision making of the sterility status of cell therapy products. Our approach has demonstrated that we can achieve high sensitivity and detect contaminated samples on par or superior to accepted methods in superior time frames.
MATERIALS AND METHODS
Cell culture
Healthy human primary T-cells (human PBMCs were obtained from a single commercial leukopak [Cat#260240.01, Lonza]) and one set of T-cells was derived from StemCell technologies (70025) were cultured in AIMV (Cat#12055091, Gibco) + 2% AB Human Serum (Cat#H4522, Sigma-Aldrich) + 100 U/mL IL-2 (Cat#130-097-748, Miltenyi Biotech) and activated using ImmunoCult Human CD3/CD28 T-Cell Activator (Cat#10971, StemCell Technologies). T-cells were cultured for 14–21 days. Cell counting was performed using a TC20 Automated Cell Counter (Biorad) with Trypan blue stain.
Microbial species were grown and cultured as described in Table 1. We used and developed standard curves alongside CFU plate counting to accurately estimate the number of bacterial cells present within a culture at a given timepoint using the 600 nm wavelength optical density (OD) values (33). CFUs were counted before the addition of microbial cells to sterile cultured T-cells. Briefly, 100 µL culture was serially diluted, 10 µL of each serial dilution was plated on agar, and CFUs were counted the following day. For low concentration samples, following overnight culture, a subculture was prepared at 0.05 OD and incubated for 1–3 h (the time was determined using the standard curve to predict the log phase of growth). Next the culture was diluted in LB to 0.1 OD 600 nm and to 0.01 OD 600 nm. Subsequently, 0.01 OD was serially diluted in phosphate buffered saline (PBS) eight- to ninefold to obtain 10 CFU/mL. At the last set of serial dilutions, 1 mL was plated onto LB agar plates for CFU counting. Only once the growth and counting of the colonies was complete were the spikes prepared. Other methodologies that could be used to confirm the final concentrations for future work include digital PCR, which we observed as a highly capable tool for measuring spiking dilutions post-rDNA PCR.
Preparation of spiking samples
Spiked samples were prepared using activated T-cells and bacterial or fungal species. Each sample used 500,000 T-cells. Bacterial and fungal cultures at a range of 10–100 CFU/mL were prepared as described previously and added directly into T-cells in PBS and incubated for 5 min.
DNA extraction and amplification
For sample preparation, a uniform approach to DNA extraction was taken to maximize DNA extraction from all possible microbes. Using the manufacturer’s protocols, the DNeasy PowerSoil Pro Kit (Cat#47014, Qiagen) was the most consistent and reliable product. The kit disrupts cell walls and membranes alike through bead beating (TissueLyser II [Cat#85300, Qiagen]). The samples were quality controlled using the NanoDrop 2000/2000 c Spectrophotometers (Cat#ND-2000, Thermofisher Scientific) to assess RNA and protein contaminants (260/280 [1.8–2.0], 260/230 [2.0–2.2]) and Qubit 4 Fluorometer (Cat#Q33238, Thermofisher Scientific) for DNA concentration.
Primers to amplify full-length 18S–28S rRNA genes were adapted from reference 34 to amplify the entire operon: 18S NS1 short F, CAGTAGTCATATGCTTGTC, and 28S RCA95m R, CTATGTTTTAATTAGACAGTCAG (34). All 16S primers described were those available in the Oxford Nanopore 16S-barcoding kit (SQK-RAB204). PCR conditions were as follows: initial denaturation 1 min at 95°C for 1 cycle, 20 s denaturation at 95°C for 29 cycles, 30 s annealing at 55°C for 29 cycles, an extension time of 2 min at 65°C for 29 cycles, followed by a final 5 min extension at 65°C, LongAmp Taq 2X Master Mix (e.g., NEB M0287).
Long-read sequencing library preparation
We used a multiplexing approach to maximize the sample throughput. This uses the existing 16S primers that have barcodes associated with them that can later be demultiplexed. In the case of the 18S–28S amplicons, no pre-existing kit exists. Thus, we applied barcodes from the rapid barcoding kit (SQK-RBK004) to the 18S–28S amplified samples, which could later be demultiplexed. Briefly, amplified DNA had RB01-12 fragmentation primers attached and were concentrated as per the manufacturer’s protocol. The Oxford Nanopore MinION sequencer with a MinIT device for base calling. The FLO-MIN106 flow cell was used to run the DNA samples for between 2 and 24 h. Reads were generated and basecalled (the electrical impedance signal was converted to a kmer string) by the Nanopore device were processed using the basecaller Guppy v3.2.10.
Pipeline tools
Metagenomic classification databases were generated from NCBI bacteria, virus, and fungal databases, these were combined into a single database. Due to runtime issues with the bacterial database using BLAST, we developed an abridged database “filter-bacteria” that contained a reduced number of entries for highly prevalent organisms within the NCBI database, for example, E. coli. This reduced the database to one-sixth its original size and was combined with the viral and fungal sequences to make a viral-fungal-bacterial database. To augment the fungal database for HS-BLAST, additional genomes were included from NCBI that were labeled as incomplete genomes. The database location and information were saved in a .txt file. Three metagenomic classifiers used in the pipeline were: Centrifuge ( 35), high-speed (HS-) BLASTn (36), and Krakenuniq (37). The DNA reads from sequencing were used as the raw data for the metagenomic classifiers. Using “centrifuge-download,” the NCBI refseq libraries for viruses (24/06/2020), fungi (25/05/2021), and bacteria (21/09/2020) were acquired and “centrifuge-build” was applied to generate the database using the NCBI taxonomy structure. Krakenuniq was an additional classifier used for removing the maximum number of host reads, in order to prevent the host reads being a majority. Porechop 0.2.4 (https://github.com/rrwick/Porechop) trimmed adapters from the Nanopore reads. Qcat 1.1.0 demultiplexed the barcoded reads, as well as the updated guppy algorithm (v6.1.5).
Pipeline
The bioinformatics pipeline (Fig. 1A and B) links together multiple tools, in addition to hosting custom scripts to check analysis progress. If the sample was multiplexed, it was demultiplexed before read trimming by porechop. We retained low-quality reads from the “fastq_fail” folder within the pipeline because the aim was to detect low concentration contaminants.
If the user specifies a known host species and reference genome (e.g., human), the pipeline will align host reads using HS-BLAST, Krakenuniq, and Centrifuge. The read IDs were used to remove these host reads from the read pool, which accelerates the workflow and reduces noise related to the host reads. Host read removal was completed in tandem using multiprocessing for each metagenomic classifier. Statistics on the number of host reads, classified and unclassified reads, were retained. Host-depleted reads were then processed using the metagenomic classifiers against the combined fungal, viral, and bacterial database. HS-BLAST and Centrifuge output were used to generate descriptive summary statistics from the troubleshooting files using the pandas describe function. Summary statistics for each predicted contaminant were generated from the Guppy file “sequencing_summary.txt” by aggregating the individual read quality scores that were processed using the describe function.
After analysis using the pipeline discussed earlier, Nanoplot (38) was used to generate and monitor run statistics. A wrapper was created to generate unique sequencing summary files for barcoded samples by splitting the reads on their barcode IDs. The “NanoStats.txt” file was cleaned up for later use in the machine learning section of the pipeline. Run environment specification were as follows: Python version: 3.8.3, Bash version: GNU bash, version 4.4.20(1) release (x86_64-pc-linux-gnu).
Machine learning pipeline
Initial pre-processing for metagenomic classification data from the combined viral, fungal, and bacterial database uses an independent filter for each metagenomic classifier. The pre-processing step is necessary because it filters out many of the low-quality predicted species made by both HS-BLAST and Centrifuge. The HS-BLAST filter uses a maximum percent identity value greater than 83%. This reduced misclassified species by 33.19% (incorrectly predicted species removed from downstream analysis), while retaining (166/167) of samples, resulting in the loss of one sample from the analysis. Similarly, the Centrifuge filter uses a minimum metagenomic classification mean score greater than 900. The filter removes misclassified species by 38.64%, while retaining 97.45% (153/157) of samples. The Nanoplot output included experiment-level summary information were subsequently concatenated to the data. Summary statistics using the describe() function were generated for the metagenome classification data. Predicted species that have NaN values (Not a Number) for read count and mean read quality were removed from the data set.
The data were pre-split based on two layers of train_test_split() 75:25 split ratio. The split ratio was selected due to our relatively small limited data set size, previous publications recommended splits that are generally between 60:40 and 80:20 training to testing (39). Previous publications provide guidance as follows: there is no clear guidance and recommendations such as 80:20 are based off rule-of-thumb decisions by practitioners (40). In larger data sets, splits can be much more extreme at 99.5:0.4:0.1 split (training/development/test). The first split generated the training and testing data set alongside an unseen and (for model assessment) unused evaluation data set. The training and testing were then split again into training data set and test data set. Features were selected using Featurewiz (https://github.com/AutoViML/featurewiz), which was used to generate a list of important features that accounted for highly correlated features, as well as less insightful features and provided features for the generation of a high-performing model.
Binary classifiers were designed to answer two questions: (1) is the sample contaminated? and (2) is the predicted contaminant correctly classified and does it match the spiked species? For (1), the labels and encoded labels were as follows: True_positive: 1, True_negative: 0 and (2) False_positive: 0, True_positive: 1. XGBoost classifier models were developed and implemented. The data preparation was similar for each. Gridsearch was used to identify the ideal model starting parameters. Performance was assessed with cross-validation testing (cv = 5, scoring = accuracy), confusion matrices, and generation of a classification report. Models were described subsequently:
Is the sample contaminated? Defined as either sterile or not sterile.
Is the predicted contaminant correctly classified? Defined as expected contaminant from the spike or any other predicted species.
The model predictions were then compared to the original labels and compared for accuracy. This step was completed for both questions (1) and (2). The last step for the training and testing of the machine learning pipeline were to take the predictions and make an assessment using a decision matrix as to whether a given sample is sterile or contaminated based on the available data.
To evaluate the machine learning model, the trained machine learning model applied to an unseen data pool that shared the same distribution as the complete data set: the evaluation data. The standard scalers were reloaded from the training step and used for standardization. The features identified by Featurewiz were also imported and reused when running the model with new data.
ACKNOWLEDGMENTS
This research is supported by the National Research Foundation, Prime Minister’s Office, Singapore, under its Campus for Research Excellence and Technological Enterprise (CREATE) program, through Singapore MIT Alliance for Research and Technology (SMART): Critical Analytics for Manufacturing Personalised-Medicine (CAMP) Inter-Disciplinary Research Group. Financial support from National Research Foundation and Ministry of Education Singapore under its Research Centre of Excellence Program was provided through Singapore Centre for Environmental Life Sciences Engineering. Graphical abstract and parts of Fig. 1A through C were created with BioRender.com.
J.P.B.S.: Conceptualization, Methodology, Validation, Software, Writing, Visualization, Formal Analysis. M.N.: Validation, Investigation. D.B.L.T.: Investigation. W.-X.S.: Investigation. E.L.: Methodology, Validation. K.-W.C.: Investigation. M.C.: Investigation. K.T.: Investigation. P.W.B.: Writing—Review & Editing. J.M.W.: Writing—Review & Editing. R.B.H.W.: Methodology, Writing—Review & Editing, Supervision. S.A.R.: Conceptualization, Methodology, Writing—Review & Editing, Supervision. S.L.S.: Conceptualization, Methodology, Writing—Review & Editing, Supervision.
The authors have declared no competing interests.
Contributor Information
Stacy L. Springs, Email: ssprings@mit.edu.
Wujian Ke, Southern Medical University, Guangzhou, China .
DATA AVAILABILITY
Pipeline can be found at https://github.com/Electrocyte/adventitious-pipeline/. The following link provides the python files used to build the model: https://github.com/Electrocyte/publication-figures. The Sequence Read Archive (SRA) deposit ID is PRJNA869859.
SUPPLEMENTAL MATERIAL
The following material is available online at https://doi.org/10.1128/spectrum.01350-23.
A table with all processed samples.
This table contains the results from running Nanoplot and gives sequencing-run raw information.
This table includes data on number of reads sequenced before removal of human host reads, as well as classified contaminant reads.
A table showing results from pycaret of comparing various machine learning models that guided our choice of a final model.
ASM does not own the copyrights to Supplemental Material that may be linked to, or accessed through, an article. The authors have granted ASM a non-exclusive, world-wide license to publish the Supplemental Material files. Please contact the corresponding author directly for reuse.
REFERENCES
- 1. Wei J, Han X, Bo J, Han W. 2019. Target selection for CAR-T therapy. J Hematol Oncol 12:62. doi: 10.1186/s13045-019-0758-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Nath SC, Harper L, Rancourt DE. 2020. Cell-based therapy manufacturing in stirred suspension bioreactor: thoughts for cGMP compliance. Front Bioeng Biotechnol 8:599674. doi: 10.3389/fbioe.2020.599674 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. García-Bernal D, García-Arranz M, Yáñez RM, Hervás-Salcedo R, Cortés A, Fernández-García M, Hernando-Rodríguez M, Quintana-Bustamante Ó, Bueren JA, García-Olmo D, Moraleda JM, Segovia JC, Zapata AG. 2021. The current status of mesenchymal stromal cells: controversies, unresolved issues and some promising solutions to improve their therapeutic efficacy. Front Cell Dev Biol 9:650664. doi: 10.3389/fcell.2021.650664 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. England MR, Stock F, Gebo JET, Frank KM, Lau AF. 2019. Comprehensive evaluation of compendial USP <71>, BacT/Alert Dual-T, and Bactec FX for detection of product sterility testing contaminants. J Clin Microbiol 57:e01548-18. doi: 10.1128/JCM.01548-18 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Barone PW, Wiebe ME, Leung JC, Hussein ITM, Keumurian FJ, Bouressa J, Brussel A, Chen D, Chong M, Dehghani H, Gerentes L, Gilbert J, Gold D, Kiss R, Kreil TR, Labatut R, Li Y, Müllberg J, Mallet L, Menzel C, Moody M, Monpoeho S, Murphy M, Plavsic M, Roth NJ, Roush D, Ruffing M, Schicho R, Snyder R, Stark D, Zhang C, Wolfrum J, Sinskey AJ, Springs SL. 2020. Viral contamination in biologic manufacture and implications for emerging therapies. Nat Biotechnol 38:563–572. doi: 10.1038/s41587-020-0507-2 [DOI] [PubMed] [Google Scholar]
- 6. Pacific Biolabs. 2018. Compendial Microbiology. Available from: https://pacificbiolabs.com/compendial-microbiology [Google Scholar]
- 7. Abela MA, Fenning S, Maguire KA, Morris KG. 2018. Bacterial contamination of platelet components not detected by BacT/ALERT. Transfus Med 28:65–70. doi: 10.1111/tme.12458 [DOI] [PubMed] [Google Scholar]
- 8. Totty H, Ullery M, Spontak J, Viray J, Adamik M, Katzin B, Dunne WM, Deol P. 2017. A controlled comparison of the Bact/ALERT 3d and VIRTUO microbial detection systems. Eur J Clin Microbiol Infect Dis 36:1795–1800. doi: 10.1007/s10096-017-2994-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. U.P.F . 2019. 〈1071〉 rapid microbial tests for release of sterile short-life products: a risk-based approach. General Information Chapter 44. doi: 10.31003/USPNF_M12457_02_01 [DOI] [Google Scholar]
- 10. Stroncek DF, Jin P, Ren J, Feng J, Castiello L, Civini S, Wang E, Marincola FM, Sabatino M. 2010. Quality assessment of cellular therapies: the emerging role of molecular assays. Korean J Hematol 45:14–22. doi: 10.5045/kjh.2010.45.1.14 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Gauthier NPG, Nelson C, Bonsall MB, Locher K, Charles M, MacDonald C, Krajden M, Chorlton SD, Manges AR. 2021. Nanopore metagenomic sequencing for detection and characterization of SARS-CoV-2 in clinical samples. PLoS One 16:e0259712. doi: 10.1371/journal.pone.0259712 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Sanderson ND, Street TL, Foster D, Swann J, Atkins BL, Brent AJ, McNally MA, Oakley S, Taylor A, Peto TEA, Crook DW, Eyre DW. 2018. Real-time analysis of nanopore-based metagenomic sequencing from infected orthopaedic devices. BMC Genomics 19:714. doi: 10.1186/s12864-018-5094-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Pecman A, Adams I, Gutiérrez-Aguirre I, Fox A, Boonham N, Ravnikar M, Kutnjak D. 2022. Systematic comparison of nanopore and illumina sequencing for the detection of plant viruses and viroids using total RNA sequencing approach. Front Microbiol 13:883921. doi: 10.3389/fmicb.2022.883921 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Matsuo Y, Komiya S, Yasumizu Y, Yasuoka Y, Mizushima K, Takagi T, Kryukov K, Fukuda A, Morimoto Y, Naito Y, Okada H, Bono H, Nakagawa S, Hirota K. 2021. Full-length 16S rRNA gene amplicon analysis of human gut microbiota using MinION nanopore sequencing confers species-level resolution. BMC Microbiol 21:35. doi: 10.1186/s12866-021-02094-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Nygaard AB, Tunsjø HS, Meisal R, Charnock C. 2020. A preliminary study on the potential of Nanopore MinION and illumina MiSeq 16S rRNA gene sequencing to characterize building-dust microbiomes. Sci Rep 10:3209. doi: 10.1038/s41598-020-59771-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Winand R, Bogaerts B, Hoffman S, Lefevre L, Delvoye M, Braekel JV, Fu Q, Roosens NH, Keersmaecker SCD, Vanneste K. 2019. Targeting the 16S rRNA gene for bacterial identification in complex mixed samples: comparative evaluation of second (illumina) and third (oxford nanopore technologies) generation sequencing technologies. Int J Mol Sci 21:298. doi: 10.3390/ijms21010298 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Kosecka-Strojek M, Sabat AJ, Akkerboom V, Kooistra-Smid A, Miedzobrodzki J, Friedrich AW. 2019. Development of a reference data set for assigning Streptococcus and Enterococcus species based on next generation sequencing of the 16S-23S rRNA region. Antimicrob Resist Infect Control 8:178. doi: 10.1186/s13756-019-0622-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Rodríguez-Pérez H, Ciuffreda L, Flores C. 2021. Nanoclust: a species-level analysis of 16S rRNA nanopore sequencing data. Bioinformatics 37:1600–1601. doi: 10.1093/bioinformatics/btaa900 [DOI] [PubMed] [Google Scholar]
- 19. Overview of United States and European Pharmacopoeia chapters dedicated to the validation of alternative Microbiological methods, . 2000. In . http://www.drugfuture.com/pharmacopoeia/usp32/pub/data/v32270/usp32nf27s0_c1223.html.
- 20. Thorpe TC, Wilson ML, Turner JE, DiGuiseppi JL, Willert M, Mirrett S, Reller LB. 1990. BacT/alert: an automated colorimetric microbial detection system. J Clin Microbiol 28:1608–1612. doi: 10.1128/jcm.28.7.1608-1612.1990 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Yue J, Chen C, Jing X, Ma Q, Li B, Zhang X. 2020. Development of next-generation sequencing-based sterility test. Microbiology. doi: 10.1101/2020.12.27.424516 [DOI]
- 22. Onaghise, O. (2020). Pharmaceutical Microbiology Manual. 92. [Google Scholar]
- 23. Guidance for industry - sterile drug products produced by aseptic processing current good manufacturing practice. 2004. https://www.gmp-compliance.org/guidelines/gmp-guideline/fda-guidance-for-industry-sterile-drug-products-produced-by-aseptic-processing-current-good-manufacturing-practice-september-200.
- 24. Salter SJ, Cox MJ, Turek EM, Calus ST, Cookson WO, Moffatt MF, Turner P, Parkhill J, Loman NJ, Walker AW. 2014. Reagent and laboratory contamination can critically impact sequence-based microbiome analyses. BMC Biol 12:87. doi: 10.1186/s12915-014-0087-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Martí JM. 2019. Recentrifuge: robust comparative analysis and contamination removal for metagenomics. PLoS Comput Biol 15:e1007131. doi: 10.1371/journal.pcbi.1007131 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Kim D, Hofstaedter CE, Zhao C, Mattei L, Tanes C, Clarke E, Lauder A, Sherrill-Mix S, Chehoud C, Kelsen J, Conrad M, Collman RG, Baldassano R, Bushman FD, Bittinger K. 2017. Optimizing methods and dodging pitfalls in microbiome research. Microbiome 5:1–14. doi: 10.1186/s40168-017-0267-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Ji Y, Zhou Z, Liu H, Davuluri RV. 2021. DNABERT: pre-trained bidirectional encoder representations from transformers model for DNA-language in genome. Bioinformatics 37:2112–2120. doi: 10.1093/bioinformatics/btab083 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Ames SK, Gardner SN, Marti JM, Slezak TR, Gokhale MB, Allen JE. 2015. Using populations of human and microbial genomes for organism detection in metagenomes. Genome Res 25:1056–1067. doi: 10.1101/gr.184879.114 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Daly GM, Leggett RM, Rowe W, Stubbs S, Wilkinson M, Ramirez-Gonzalez RH, Caccamo M, Bernal W, Heeney JL. 2015. Host subtraction, filtering and assembly validations for novel viral discovery using next generation sequencing data. PLoS One 10:e0129059. doi: 10.1371/journal.pone.0129059 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Pourhabibi T, Ong K-L, Kam BH, Boo YL. 2020. Fraud detection: a systematic literature review of graph-based anomaly detection approaches. Decision Support Systems 133:113303. doi: 10.1016/j.dss.2020.113303 [DOI] [Google Scholar]
- 31. Tiwari P, Mehta S, Sakhuja N, Kumar J, Singh AK. 2021. Credit card fraud detection using machine learning: a study. doi: 10.48550/arXiv.2108.10005 [DOI]
- 32. Li Z, Zhao Y, Botta N, Ionescu C, Hu X. 2020. “COPOD: Copula-based Outlier detection” 2020 IEEE International Conference on Data Mining (ICDM); Sorrento, Italy: p 1118–1123. doi: 10.1109/ICDM50108.2020.00135 [DOI] [Google Scholar]
- 33. Stevenson K, McVey AF, Clark IBN, Swain PS, Pilizota T. 2016. General calibration of microbial growth in microplate readers. Sci Rep 6:38828. doi: 10.1038/srep38828 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34. Karst SM, Ziels RM, Kirkegaard RH, Sørensen EA, McDonald D, Zhu Q, Knight R, Albertsen M. 2021. High-accuracy long-read amplicon sequences using unique molecular identifiers with nanopore or pacbio sequencing. Nat Methods 18:165–169. doi: 10.1038/s41592-020-01041-y [DOI] [PubMed] [Google Scholar]
- 35. Kim D, Song L, Breitwieser FP, Salzberg SL. 2016. Centrifuge: rapid and sensitive classification of metagenomic sequences. Genome Res 26:1721–1729. doi: 10.1101/gr.210641.116 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Chen Y, Ye W, Zhang Y, Xu Y. 2015. High speed BLASTN: an accelerated megablast search tool. Nucleic Acids Res 43:7762–7768. doi: 10.1093/nar/gkv784 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Breitwieser FP, Baker DN, Salzberg SL. 2018. KrakenUniq: confident and fast metagenomics classification using unique k-mer counts. Genome Biol 19:198. doi: 10.1186/s13059-018-1568-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. De Coster W, D’Hert S, Schultz DT, Cruts M, Van Broeckhoven C. 2018. NanoPack: visualizing and processing long-read sequencing data. Bioinformatics 34:2666–2669. doi: 10.1093/bioinformatics/bty149 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Xu Y, Goodacre R. 2018. On splitting training and validation set: a comparative study of cross-validation, bootstrap and systematic sampling for estimating the generalization performance of supervised learning. J Anal Test 2:249–262. doi: 10.1007/s41664-018-0068-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Joseph VR. 2022. Optimal ratio for data splitting. Statistical Analysis 15:531–538. doi: 10.1002/sam.11583 [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
A table with all processed samples.
This table contains the results from running Nanoplot and gives sequencing-run raw information.
This table includes data on number of reads sequenced before removal of human host reads, as well as classified contaminant reads.
A table showing results from pycaret of comparing various machine learning models that guided our choice of a final model.
Data Availability Statement
Pipeline can be found at https://github.com/Electrocyte/adventitious-pipeline/. The following link provides the python files used to build the model: https://github.com/Electrocyte/publication-figures. The Sequence Read Archive (SRA) deposit ID is PRJNA869859.