Published in final edited form as: Ultrasound Med Biol. 2023 Aug 28;49(11):2388–2397. doi: 10.1016/j.ultrasmedbio.2023.08.002

Table 1.

Acoustic parameters for the three insonation protocols. The peak negative pressure is reported as the maximum value at the external surface of the EKOS catheter. The order of schemes for protocol 3 was 1, 2, 3, 4, 3, 2.

Protocol# Scheme# Electrical drive
pulse power (W)
Peak
negative
pressure
(MPa)
Pulse
duration
(ms)
PRF
(Hz)
Duty
cycle (%)
1 - 18 0.95 3 50 15
2 - 18 0.95 15 10 15
3 1 9 0.67 8.0 21 16.93
2 15 0.87 5.0 27 13.50
3 30 1.20 6.9 21 14.41
4 47 1.47 4.0 27 10.80