Intra-day and inter-day precision and accuracy of almonertinib, apixaban and rivaroxaban in rat plasma.

Analytes Concentration (ng/mL) Intra-day (n = 6) Inter-day (n = 18)
Mean ± SD RSD (%) RE (%) Mean ± SD RSD (%) RE (%)
almonertinib 0.5 0.53 ± 0.04 6.7 6.47 0.51 ± 0.05 9.4 1.83
1 0.98 ± 0.05 5.4 −2.43 0.98 ± 0.08 8.1 −2.05
80 77.70 ± 3.27 4.2 −2.88 79.90 ± 4.51 5.6 −0.13
150 152.50 ± 4.51 3.0 1.67 148.06 ± 7.37 5.0 −1.30
apixaban 1 1.03 ± 0.05 4.6 2.95 0.99 ± 0.06 5.6 −0.21
2 2.03 ± 0.05 2.7 1.67 2.04 ± 0.07 3.3 1.83
400 407.17 ± 7.41 1.8 1.79 410.72 ± 15.56 3.8 −3.50
800 808.50 ± 19.38 2.4 2.13 806.83 ± 22.60 2.8 −1.50
rivaroxaban 5 4.76 ± 0.11 2.3 −4.80 4.93 ± 0.22 4.4 1.29
10 10.47 ± 0.37 3.6 4.68 10.80 ± 0.42 3.9 8.01
800 769.00 ± 22.86 3.0 −3.88 779.22 ± 32.08 4.1 −2.60
1,500 1,486.67 ± 52.41 3.5 −0.89 1,446.11 ± 85.42 5.9 −3.59