Skip to main content
. 2023 Oct 5;42(11):1355–1363. doi: 10.1007/s10096-023-04666-x

Table 3.

Practical aspects considering usability

Property Biotyper sirius VITEK MS PRIME
Usability
  Dimensions of device

Tabletop

500 × 1070 × 710 mm

(W × H × D)

Tabletop

710 × 1100 × 700 mm

(W × H × D)

  Weight of device 75 kg 146 kg
  Control PC Separate On-board PC
  Noise emission  < 60 dB  < 65 dB
  Ready-to-use matrix solution No Yes
  Calibration BTS E. coli ATCC 8739
  Ease of target preparation High Moderate
  Disposable target Yes Yes
  Reusable target Yes No
  Capacity per target (spots) 96 48
  Loading capacity (spots) 1 × 96 16 × 48
  Random access No Yes
Database, IVD extensions, and additional kits
  Database

4194 species

(BMT IVD reference library 12.0)

Mycobacteria and filamentous fungi not included

1316 species

(VITEK IVD 3.2 knowledge base)

Mycobacteria and filamentous fungi included

  IVD software extensions

MBT Mycobacteria IVD Module

MBT HT Filamentous Fungi IVD Module

MBT Subtyping IVD Module

MBT HT Sepsityper IVD Module

No
  Additional kits

MBT Sepsityper IVD Kit

MBT STAR-Carba IVD Assay

MBT STAR-Cepha IVD Assay

VITEK MS Blood Culture Kit (RUO)

VITEK MS Mould Kit (IVD)

VITEK MS Mycobacteria/Nocardia Kit (IVD)

W width, H height, D depth