TABLE 3.

An overview of the design requirements and their results.

# Design requirements Goal Result [median (range)]
1 Instrument is detachable from driver All shafts and end-effector All shafts and end-effector are detachable
2 MI is detachable from MU Remove IM from the sterile site IM is removable and separated by the SBI
3 Adaptive instrument clutching Shaft size of 3–5 mm The SKC’s can clutch 7 mm diameter and 5 mm instruments
4 Low instrument exchange time 15 s IM exchange possible in 11.0 (6.3–14.6) seconds
5 No arm reorientation required Independent from arm IM exchange is independent from arm
6 Instant instrument control 2 s IM and sensors operational in 0.98 (0.07–2.38) seconds
7 Precise instrument control 0.5° accuracy and 50 ms latency 0.152° articulation and 25 ms latency
8 Focus on cost reduction Production price of 200€ Efficient design cost less than 200€ to produce
9 Reusable system through cleanability Internal surfaces exposable All internal surfaces of IN, IM and SBI are exposable
10 Low weight 500 g or less Total system at 550 g, IM at 350 g