TABLE 3.
An overview of the design requirements and their results.
| # | Design requirements | Goal | Result [median (range)] |
|---|---|---|---|
| 1 | Instrument is detachable from driver | All shafts and end-effector | All shafts and end-effector are detachable |
| 2 | MI is detachable from MU | Remove IM from the sterile site | IM is removable and separated by the SBI |
| 3 | Adaptive instrument clutching | Shaft size of 3–5 mm | The SKC’s can clutch 7 mm diameter and 5 mm instruments |
| 4 | Low instrument exchange time | 15 s | IM exchange possible in 11.0 (6.3–14.6) seconds |
| 5 | No arm reorientation required | Independent from arm | IM exchange is independent from arm |
| 6 | Instant instrument control | 2 s | IM and sensors operational in 0.98 (0.07–2.38) seconds |
| 7 | Precise instrument control | 0.5° accuracy and 50 ms latency | 0.152° articulation and 25 ms latency |
| 8 | Focus on cost reduction | Production price of 200€ | Efficient design cost less than 200€ to produce |
| 9 | Reusable system through cleanability | Internal surfaces exposable | All internal surfaces of IN, IM and SBI are exposable |
| 10 | Low weight | 500 g or less | Total system at 550 g, IM at 350 g |