GC-MS/MS detection parameters

Items Parameters
Injection volume 1 µL
Front Inlet Mode 5:1
Carrier Gas Helium
Column DB-FFAP (30m x 0.25mm x 0.25µm)
Column Flow 1.2 mL/min
Oven Temperature Ramp held at 50°C for 1 min, raised to 220°C at a rate of 18°C/min and kept for 5 min.
Front Injection Temperature 250°C
Transfer Line Temperature 230°C
Ion Source Temperature 230°C
Quad Temperature 150°C