Skip to main content
. 2023 Oct 9;8(5):911–922. doi: 10.1089/can.2021.0165

Table 5.

Linear Regression Statistics from Extreme Gradient Boost Model That Includes All 576 Prediction and Cross-Validation Data Points for Total Cannabidiol and Total Tetrahydrocannabinol Measured by Gas Chromatography-Flame Ionization

Technique Parameter Intercept SE Slope LOD % (w/w) LOQ % (w/w) Adjusted R2 n
A-TEEM CBD+CBDA 0.019 0.006 1 0.019 0.059 0.9954 576
THC+THCA −0.016 0.006 1 0.020 0.061 0.9958 576
GC-FID CBD+CBDA −2.872×10−5 NA 5.424×10−4 0.01 0.025 1 NA
THC+THCA −2.877×10−6 NA 5.546×10−4 0.01 0.025 1 NA

The slope for A-TEEM data was constrained to unity. Defined are the y-intercept (intercept), SE of the y-intercept (SE), LOD=3.3×SE, LOQ=3.3×LOD, the adjusted R2, and number of files (n) that combines calibration and validation.

A-TEEM, absorbance-transmittance excitation emission matrix; CBDA, cannbidiolic acid; GC, gas chromatography; LOD, limit of detection; LOQ, limit of quantification; NA, not applicable; SE, standard error; THCA, tetrahydrocannabinolic acid.