Table 2.
Tolerable values (log KPotAC+) of the designed AC-PM-MWCNTs and AC-PM-MWCNTs-Al2O3 nanocomposite sensors estimated by separate and mixed solution techniques using 1.0 × 10−3 mol L−1AC and a foreign material (1.0 × 10−3 and 1.0 × 10−1 mol L−1) solution for the two methods, respectively.
Interferents |
AC-PM-MWCNTs sensor |
AC-PM-MWCNTs-Al2O3 nanocomposite sensor |
||
---|---|---|---|---|
SSM | MSM | SSM | MSM | |
Mg2+ | −2.6 | −1.4 | −3.4 | −2.9 |
Ca2+ | −2.4 | −2.4 | −4.7 | −4.3 |
Ti4+ | −2.7 | −3.4 | −4.2 | −4.6 |
Na+ | −2.2 | −7.1 | −3.2 | −3.5 |
Cu2+ | −2.1 | −9.7 | −3.8 | −2.8 |
K+ | −2.6 | −2.7 | −4.6 | −4.2 |
Cr3+ | −2.0 | −2.1 | −3.3 | −3.9 |
Al3+ | −2.6 | −2.6 | −3.2 | −2.7 |
Zn2+ | −2.4 | −2.5 | −3.8 | −3.8 |
Lactose | −2.7 | −2.7 | −3.5 | −4.2 |
Sucrose | −2.1 | −2.1 | −4.1 | −3.6 |
Histidine | −2.6 | −2.7 | −4.2 | −5.2 |
Glycine | −2.9 | −2.5 | −3.3 | −3.4 |
Magnesium | −2.4 | −2.3 | −3.6 | −3.1 |
stearate | −2.3 | −2.4 | −3.1 | −4.1 |