Table 1.

Representative uncertainty budget for a microdrop deposition. These data correspond to the third data point in Figure 4.

Term value u
mf(td) 31.8863 mg 0.0012 mg
mr(td) 26.1471 mg 0.0042 mg
mf(tf) 31.9088 mg 0.0040 mg
me 24.9354 mg 0.010 mg
bf 0.99890 0.00013
ubal 0.0006 mg
Ad 831.60 Bq 0.59 Bq
 
Am0 (Eq. 4) 144.27 0.17 Bq/mg
Am0 (Eq. 5) 144.17 0.18 Bq/mg