[Preprint]. 2023 Oct 11:2023.10.10.23296833. [Version 2] doi: 10.1101/2023.10.10.23296833

Table 2:

Sensitivity and specificity of swab testing by Xpert Ultra and MTB Ultima, based on all samples tested.

Cohort Sensitivity Specificity
Xpert Ultra MTB Ultima Xpert Ultra MTB Ultima
N Percent (95% CI) N Percent (95% CI) N Percent (95% CI) N Percent (95% CI)
1 71/100 71.0 (61.1, 79.6) 58/89 65.2 (54.3, 75.0) -* -* -* -*
2 113/144 78.5 (70.9, 84.9) 103/136 75.7 (67.7, 82.7) 72/72 100.0 (95.0, 100.0) 62/64 96.9 (89.2, 99.6
1+2 184/244** 75.4 (69.5, 80.7) 161/225** 71.6 (65.2, 77.4) 72/72** 100.0 (95.0, 100.0) 62/64** 96.9 (89.2, 99.6
*

Cohort 1 consisted entirely of TB-positive participants by sputum MRS

**

Totals fell below the total enrollment of 320 participants due to invalid results, which differed between the two test platforms