Evidentiary table (selected key studies only): non-invasive cerebral oxygenation monitoring (NIRS)
| Reference | Patient number | Study design | Patient group | Technique assessment | End-point | Findings | Quality of evidence |
|---|---|---|---|---|---|---|---|
| Buchner, 2000 | 31 | Prosp Obs | SAH, TBI | NIRS With PbtO2 | Data quality, factors influencing signal, parameter correlation | 50–80 % good quality data, signal influenced by optode wetting, galea hematoma, subdural air; partial correlation of NIRS with PbtO2 | Low |
| Kirkpatrick, 1998 | 130 | Prosp Obs | Carotid endarterectomy | NIRS with EEG, TCD CBFV | Ischemia thresholds | 80 % good quality data | Low |
| Davie, 2012 | 12 | Prosp Obs | Healthy volunteers | NIRS | Extracerebral signal influence | Head cuff inflation reveals 7–17 % extracranial signal contribution in three commercial NIRS monitors | Low |
| Yoshitani, 2007 | 103 | Prosp Obs | Cardiac/Neuro-ICU | NIRS | Factors influencing the signal | NIRS signal (rSO2) influenced by skull thickness, CSF layer, hemoglobin | Moderate |
| Brawanski, 2002 | 12 | Prosp Obs | TBI, SAH | NIRS with PbtO2 | Inter-monitoring correlation | Good correlation | Low |
| Rothoerl, 2002 | 13 | Prosp Obs | TBI, SAH | NIRS with PbtO2 | Inter-monitoring correlation | Good correlation | Low |
| McLeod, 2003 | 8 | Prosp Obs | TBI | NIRS with PbtO2, SjvO2 | Inter-monitoring correlation | Good correlation | Low |
| Ter Minassian, 1999 | 9 | Prosp Obs | TBI | NIRS with SjvO2 | Inter-monitoring correlation | Poor correlation | Low |
| Buunk, 1998 | 10 | Prosp CS | Cardiac arrest | NIRS with SjvO2 | Inter-monitoring correlation | Poor correlation | Low |
| Weerakkody, 2012 | 40 | Prosp Obs | TBI | NIRS with ICP/CPP | Inter-monitoring correlation | Good correlation | Low |
| Zweifel, 2010a | 40 | Prosp Obs | TBI | NIRS with ICP/CPP PRx | Inter-monitoring correlation | Good correlation | Moderate |
| Zweifel, 2010b | 27 | Prosp Obs | SAH | NIRS with TCD CBFV/ MAP Mx | Inter-monitoring correlation | Good correlation | Moderate |
| Rothoerl, 2003 | 9 | Prosp Obs | TBI | NIRS with Xe133 perfusion | Inter-monitoring correlation | Poor correlation | Low |
| Terborg, 2004 | 25 | Prosp CaseCont | Hem AIS | NIRS with MRI perfusion | Inter-monitoring correlation | Good correlation | Moderate |
| Frisch, 2012 | 5 | Case series | Card arrest | NIRS with PetCO2 | Inter-monitoring correlation | Poor correlation | Very low |
| Bhatia, 2007 | 32 | Prosp Obs | SAH | NIRS with DSA | Inter-monitoring correlation | Good correlation | Low |
| Taussky, 2012 | 6 | Retrosp CS | SAH, AIS, ICH | NIRS with CT perfusion | Inter-monitoring correlation | Good correlation | Very low |
| Aries, 2012 | 9 | Prosp Obs | AIS | NIRS with SaO2 and blood pressure | Signal response to drops in SaO2 and BP | Good detection of desaturations less good detection of hypotension | Very low |
| Hargroves, 2008 | 7 | Prosp Obs | AIS | NIRS | Signal response to position of head of bed | Good reflection of position-related oxygenation changes | Very low |
| Damian, 2007 | 24 | Retro Obs | AIS | NIRS | Outcome, clinical course, imaging | Bilateral NIRS with interhemispheric difference reflecting clinical course, outcome and effect of decompressive surgery | Low |
| Bonoczk, 2002 | 43 | RCT interv. | AIS | NIRS with TCD | Response of rSO2 and CBFV to vinpocetine | Increase of rSO2 in response to vinpocetine | Moderate |
| Naidech, 2008 | 6 | Prosp CS | SAH | NIRS with TCD, DSA | Change of NIRS signal in vasospasm | No reliable detection of vasospasm by NIRS | Very low |
| Yokose, 2010 | 11 | Prosp CS | SAH | NIRS With TCD, DSA | Change of NIRS signal in vasospasm | Good detection of vasospasm | Very low |
| Mutoh, 2010 | 7 | Prosp CS | SAH | NIRS | Response of NIRS signal to dobutamine | Detection of vasospasm by NIRS, NIRS signal increasing with incremental dobutamine | Low |
| Gopinath, 1993 | 40 | Prosp Obs | TBI | NIRS with CT | Detection of secondary hematoma | Detection secondary hematoma by NIRS | Low |
| Gopinath, 1995 | 167 | Prosp Obs | TBI | NIRS with CT, ICP, clinical | Time to detection of secondary hematomy | Earlier detection of secondary hematoma than by ICP, clinical signs or CT | Moderate |
| Budohoski, 2012 | 121 | Prosp Obs | TBI | NIRS with TCD CBFV, PbtO2, MAP, ICP | Time to cerebral parameter changes MAP and ICP increases | Earlier reflection of MAP and ICP changes by NIRS than by TCD and PbtO2 | Moderate |