Skip to main content
. 2023 Oct 24;3(10):e0001681. doi: 10.1371/journal.pgph.0001681

Table 2. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy of the BEMPU TempWatch.

Sensitivity (%) Specificity (%) PPV (%) NPV (%) Accuracy (%)
All neonates 250/369 (67.8%) 1,538/1,604 (95.9%) 250/316 (79.1%) 1,538/1,657 (92.8%) 90.6%
Incubator 87/149 (58.4%) 1,039/1,064 (97.7%) 87/112 (77.7%) 1,039/1,101 (94.4%) 92.8%
Not in incubators 139/168 (82.7%) 433/470 (92.1%) 139/176 (79.0%) 433/462 (93.7%) 89.7%
Low birth weight (1,500–2,500 g) 125/170 (73.5%) 810/848 (95.5%) 125/163 (76.7%) 810/855 (94.7%) 91.8%
Very or extremely low birth weight (<1,500 g) 125/199 (62.8%) 728/756 (96.3%) 125/153 (81.7%) 728/802 (90.8%) 89.3%