Skip to main content
. 2023 Oct 24;22:102. doi: 10.1186/s12938-023-01165-0

Table 2.

Identification performance compared to clinical inspection by type of asynchrony and for non-asynchrony breaths

Asynchrony types TP FN TN FP Sensitivity (%) Specificity (%) Accuracy (%)
fa 68 8 4912 12 89.5 99.8 99.6
rt 595 13 4384 8 97.9 99.8 99.6
pc 383 22 4581 14 94.6 99.7 99.3
dt 241 5 4748 6 98.0 99.9 99.8
dc 202 15 4741 42 93.1 99.1 98.9
ie 133 9 4848 10 93.7 99.8 99.6
at 415 34 4543 8 92.4 99.8 99.2
Non-asynchrony breaths 2542 52 2328 78 98.0 96.8 97.4