Published in final edited form as: Ann Surg Oncol. 2023 Apr 11;30(7):4097–4108. doi: 10.1245/s10434-023-13364-z

TABLE 3.

Margin-level and specimen-level assessment results and performance metrics with 95% Clopper–Pearson confidence intervals for all 100 initial resections

Parameter Value
Margin-level results Path+, Image+ (TP) Path−, Image− (TN) Path−, Image+ (FP) Path+, Image− (FN)
SIA 7 555 24 14
Micro-CT Reader 1 6 514 65 15
Micro-CT Reader 2 5 520 59 16
Micro-CT Reader 3 9 508 71 12

Margin-level performance, n (95% CI) Sensitivity (%) Specificity (%) PPV (%) NPV (%)

SIA 33.3 (14.6–57.0) 95.9 (93.9–97.3) 22.6 (12.4–37.5) 97.5 (96.7–98.2)
Micro-CT Reader 1 28.6 (11.3–52.2) 88.8 (85.9–91.2) 8.5 (4.3–15.9) 97.2 (96.3–97.8)
Micro-CT Reader 2 23.8 (8.2–47.2) 89.8 (87.0–92.1) 7.8 (3.7–15.9) 97.0 (96.2–97.6)
Micro-CT Reader 3 42.9 (21.8–66.0) 87.7 (84.8–90.3) 11.3 (6.9–17.9) 97.7 (96.7–98.4)

Specimen-level results Path+, Image+ (TP) Path−, Image− (TN) Path−, Image+ (FP) Path+, Image− (FN)

SIA 6 66 20 8
Micro-CT Reader 1 6 54 32 8
Micro-CT Reader 2 5 59 27 9
Micro-CT Reader 3 7 48 38 7

Specimen-level performance, n (95% CI) Sensitivity (%) Specificity (%) PPV (%) NPV (%)

SIA 42.9 (17.7–71.1) 76.7 (66.4–85.2) 23.1 (12.8–38.1) 89.2 (83.8–93.0)
Micro-CT Reader 1 42.9 (17.7–71.1) 62.8 (51.7–73.0) 15.8 (8.8–26.7) 87.1 (80.7–91.6)
Micro-CT Reader 2 35.7 (12.8–64.9) 68.6 (57.7–78.2) 15.6 (7.9–28.6) 86.8 (81.2–90.9)
Micro-CT Reader 3 50.0 (23.0–77.0) 55.8 (44.7–66.5) 15.6 (9.4–24.7) 87.3 (79.7–92.3)

Path pathology; Image intraoperative margin assessment technique; TP true positive; TN true negative; FP false positive; FN false negative; SIA standard-of-care intraoperative margin assessment; CI confidence interval; PPV positive predictive value; NPV negative predictive value