Skip to main content
. 2023 Oct 12;13(20):3186. doi: 10.3390/diagnostics13203186

Table 1.

Standard MRI protocol for implant-based breast reconstruction.

Sequence T1 TSE T2 TSE Dynamic 3D Fat Suppressed DWI Silicone Suppression (Fat Suppressed) Silicone Excited (Water Suppressed)
Slice thickness (mm) 2 4 2 4 4 4
FOV (mm) 360 360 220 360 360 360
TR 5.9 3500 4.8 3520 6470 4500
TE 2.46 86 1.73 54 88 54
Time (min) 1:19 2:27 6:46 4:00 4:46 3:20

FOV—Field Of View, TSE—turbo spin echo, DWI—diffusion weighted imaging.