Table 1.
Statistical data of back-calculated concentrations of PACAP, from 10 independent runs using six calibration points.
Expected Concentration (pmol/L) | Observed Concentration (pmol/L) | Accuracy (Observed/ Expected) × 100 (%) |
||
---|---|---|---|---|
Mean | SD | %CV | ||
400 | 404 | 4.3 | 1.1 | 101 |
200 | 195 | 4.0 | 2.0 | 97.6 |
80 | 85.2 | 4.0 | 4.7 | 107 |
32 | 32.2 | 2.6 | 8.0 | 101 |
12.8 | 13.5 | 1.1 | 8.2 | 105 |
5.12 | 3.84 | 0.6 | 15.7 | 75.1 |
CV: Coefficient of variation; SD: standard deviation, n = 10 plates.