AF signal levels in CH1 and CH2, and relative changes to AF signal baselines. All results are expressed in an arbitrary unit of grayscale values calculated from 16-bit images.

Application protocols Samples (n=10) AF Signal in CH1 AF Signal in CH2
Mean ± SD Relative change Mean ± SD Relative change
Pre-treatment Baseline (n=50) 33189 32,905
0.3% SBB 32,780±3.6 -409 32,765±4.7 -140
0.1% NaBH4 33,275±35 86 32,941±17 36
1% NaBH4 33,347±44 158 32,945±17 40
TrueVIEW® 33,132±34 -57 32,898±12 -7
TrueBlack® 32,830±11 -359 32,778±4.4 -127
Post-treatment Baseline (n=50) 33,069 32,884
0.3% SBB 32,789 ± 4.5 -280 32,786 ± 8.8 -98
0.1% NaBH4 33,222±20 153 32,893±9.5 9
1% NaBH4 33,251 ± 39 182 32,928±15 44
TrueVIEW® 32,943±26 -126 32,820±12 -64
TrueBlack® 32,811±7.8 -258 32,772±5.0 -162