Published in final edited form as: IEEE Trans Biomed Eng. 2018 Oct 1;65(11):2405–2416. doi: 10.1109/TBME.2018.2873297

Table II -.

Comparison between cuff and ultrasound approach for volunteer 1 minutes-study. Absolute relative error is abbreviated ARE.

Mean (mmHg) Standard Dev. (mmHg) ARE Mean ARE Median Precision (mmHg) Accuracy (mmHg)
Systolic Test 135.6 2.5 2.0% 1.8% 0.2 3.3
Full 135.7 2.1 2.2% 2.1% 0.06 3.7

Diastolic Test 85.8 0.4 2.3% 1.6% −1.3 2.2
Full 85.8 0.8 2.5% 2.3% −0.6 2.6