Skip to main content
. Author manuscript; available in PMC: 2023 Oct 30.
Published in final edited form as: IEEE Trans Biomed Eng. 2018 Oct 1;65(11):2405–2416. doi: 10.1109/TBME.2018.2873297

Table VI-.

Results for hypertensive volunteer in longitudinal study over 13 non-consecutive days.

Hypertensive Absolute relative error
Mean Median Precision (mmHg) Accuracy (mmHg)
Systolic Full 8.7% 9.0% −2.0 11.1
Diastolic Full 4.9% 3.8% −2.1 3.7