. 2023 Oct 30;22:15330338231210786. doi: 10.1177/15330338231210786

Table 2.

The Anthropomorphic Phantom Study. (Left) Setup Discrepancies Between ExacTrac X-ray and kV-CBCT and (Right) ExacTrac X-ray Accuracy When Known Offsets for the 6 Translational and Rotational Movements Were Applied.

Motions Discrepancies ExacTrac X-ray versus kV-CBCT phantom study ExacTrac X-ray accuracy phantom study
RMS SD Max p-value RMS SD Max
Translations(mm)
Vert. (AP) 0.23 0.14 0.61 .042 0.17 0.10 0.38
Long. (SI) 0.47 0.23 0.84 <.001 0.13 0.10 0.47
Lat. (LR) 0.52 0.19 0.97 <.001 0.11 0.06 0.68
Rotations (°)
Pitch (Rx) 0.20 0.12 0.48 <.001 0.11 0.06 0.25
Roll (Rz) 0.31 0.18 0.68 .005 0.19 0.11 0.37
Yaw (Ry) 0.30 0.14 0.53 <.001 0.12 0.07 0.25

kV-CBCT, kV-cone beam computed tomography; RMS = root-mean-square.