Table 3.

Setup Discrepancies Between ExacTrac X-ray 6D and CBCT for the Patient Study. Left, for the Spinal Patients Group and Right for the Nonspinal Cases.

Motions Discrepancies ExacTrac X-ray versus kV-CBCT spinal Discrepancies ExacTrac X-ray versus kV-CBCT nonspinal
RMS SD Max p-value RMS SD Max p-value
Translations(mm)
Vert. (AP) 0.42 0.42 2.14 .463 0.58 0.55 2.42 <.0001
Long. (SI) 0.51 0.39 1.46 <.0001 0.66 0.61 1.75 <.0001
Lat. (LR) 0.75 0.47 1.58 <.0001 0.73 0.70 2.4 .099
Rotations (°)
Pitch (Rx) 0.52 0.51 2.03 .067 0.66 0.66 2.27 .147
Roll (Rz) 0.44 0.38 1.31 <.0001 0.77 0.67 1.93 <.0001
Yaw (Ry) 0.57 0.42 1.92 <.0001 0.65 0.59 2.23 <.0001

kV-CBCT, kV-cone beam computed tomography.