Table 2 Referenced optimal electrocautery settings based on location and procedure type for ConMed Beamer and the equivalent nomenclature for ERBE.

CONMED Beamer ESU ERBE ESU equivalent setting
ESD, endoscopic submucosal dissection; POEM, peroral endoscopic myotomy.
POEM
  • Incision and myotomy: G2 35W

Incision and myotomy: Endocut I Mode
  • Submucosal tunneling: 37W spray Coagulation

Submucosal tunneling: 37W spray Coagulation
  • Coagulation grasper (hemostasis): 50W gentle coagulation

Coagulation grasper (hemostasis): 50W soft coagulation
ESD esophageal
  • Incision: G5 35W

Incision: EndoCut I mode
  • Dissection: 35W hot biopsy setting

Dissection: 35W forced coagulation
  • Coagulation grasper (hemostasis): 50W gentle coagulation

Coagulation grasper (Hemostasis): 50W soft coagulation
ESD gastric
  • Incision: G5 35W

Incision: EndoCut I mode
  • Dissection: 35W hot biopsy setting

Dissection: 35W forced coagulation
  • Coagulation grasper (hemostasis): 50W Gentle coagulation

Coagulation grasper (hemostasis): 50W soft coagulation
ESD duodenum
  • Incision: G5 35W

Incision: EndoCut I mode
  • Dissection: 30W hot biopsy setting

Dissection: 30W forced coagulation
  • Coagulation grasper (hemostasis): 50W gentle coagulation

Coagulation grasper (hemostasis): 50W Soft coagulation
ESD colon
  • Incision: G5 35W

Incision: EndoCut I mode
  • Dissection: 30 to 35W hot biopsy setting

Dissection: 30 to 35W forced coagulation
  • Coagulation grasper (hemostasis): 50W Gentle coagulation

Coagulation grasper (Hemostasis): 50W soft coagulation