Skip to main content
. 2023 Apr 19;37(16):3429–3434. doi: 10.1038/s41433-023-02528-6

Table 3.

Repeatability for keratometry and pachymetry measurements in eyes with very severe corneal thinning versus severe corneal thinning.

Parameter Mean ± SD Repeatability limit CV (%) ICC ICC 95% CI
Very severe thinning with TCT ≤ 330 µm (n = 27)
Kmax (D) 81.29 ± 2.28 8.10 2.82 0.94 0.89–0.97
K1A (D) 65.85 ± 1.52 5.43 2.35 0.98 0.95–0.99
K2A (D) 70.17 ± 1.14 4.05 1.60 0.99 0.98–0.99
K1P (D) 10.27 ± 0.33 1.41 3.24 0.96 0.93–0.98
K2P (D) 11.34 ± 0.36 1.28 3.17 0.96 0.92–0.98
ASTIG (D) 5.28 ± 1.83 6.50 39.03 0.82 0.66–0.91
TCT (µm) 292.68 ± 17.01 57.72 5.74 0.86 0.71–0.93
Severe thinning with TCT 331–400 µm (n = 87)
Kmax (D) 70.25 ± 1.18 5.63 1.64 0.99 0.98–0.99
K1A (D) 55.68 ± 0.57 3.11 1.04 0.99 0.99–0.99
K2A (D) 61.32 ± 0.69 2.93 1.15 0.99 0.99–0.99
K1P (D) 8.43 ± 0.18 0.87 2.22 0.98 0.97–0.99
K2P (D) 9.50 ± 0.19 0.87 1.98 0.98 0.97–0.98
ASTIG (D) 5.77 ± 0.72 2.49 18.44 0.97 0.96–0.98
TCT (µm) 374.66 ± 5.66 22.16 1.52 0.95 0.93–0.96

SD standard deviation, CV coefficient of variation, ICC intraclass correlation coefficient, D dioptre, K1A flat keratometry of anterior surface, K2A steep keratometry of anterior surface, K1P flat keratometry of posterior surface, K2P steep keratometry of posterior surface, ASTIG astigmatism, TCT thinnest corneal thickness.