Table 2.
Mean values and standard deviation (SD), intraclass correlation coefficient (ICC), 95% confidence interval (95% CI), standard error of measurement (SEM), minimal detectable change (MDC), and coefficient of variation (CV) for inter-rater measurements of the cross-sectional area (CSA) of the Patellar Tendon
| Mean ± SD (cm2) | ICC | 95% CI | p value | SEM (cm2) | MDC (cm2) | CV (%) | |
|---|---|---|---|---|---|---|---|
| Inter-rater–without device | |||||||
| CSA25% | 0.793 ± 0.181 | 0.36 | 0.11–0.58 | 0.02 | 0.14 | 0.40 | 22.65 |
| CSA50% | 0.792 ± 0.160 | 0.65 | 0.47–0.78 | < 0.001 | 0.09 | 0.26 | 13.10 |
| CSA75% | 0.780 ± 0.179 | 0.47 | 0.25–0.66 | < 0.001 | 0.13 | 0.36 | 20.96 |
| Inter-rater–with device | |||||||
| CSA25% | 0.817 ± 0.213 | 0.46 | 0.21–0.64 | < 0.001 | 0.15 | 0.43 | 12.64 |
| CSA50% | 0.807 ± 0.180 | 0.65 | 0.47–0.78 | < 0.001 | 0.10 | 0.29 | 14.44 |
| CSA75% | 0.780 ± 0.169 | 0.73 | 0.58–0.83 | < 0.001 | 0.08 | 0.34 | 12.64 |