Skip to main content
. 2023 Feb 8;15(10):1355–1364. doi: 10.1039/d2ay02019j

UHPLC-HRMS quality parameters for the analysis of PLTX samples (System I, Method 1, QM-B).

Concentration %
Run-to-run precision (RSD, %) Low level (75 ng mL−1) 1.3
Medium level (800 ng mL−1) 1.0
Day-to-day precision (RSD, %) Low level (75 ng mL−1) 1.6
Medium level (800 ng mL−1) 3.4
Trueness (rel. error, %) Low level (75 ng mL−1) 7.6
Medium level (800 ng mL−1) 9.7
LOQ 6.5 ng mL−1