. 2023 Oct 3;24(11):e14166. doi: 10.1002/acm2.14166

TABLE 3.

Mean CT attenuation and noise of the reference, uncorrected, MAC, VMI, and AI‐MAC images with the phantom using the Φ 6.5 × 30‐mm and Φ 7.5 × 40‐mm screws.

Φ 6.5 × 30‐mm screws Φ 7.5 × 40‐mm screws
Reference Uncorrected MAC VMI AI‐MAC Uncorrected MAC VMI AI‐MAC
Bone
CT attenuation (HU) 363.98 ± 33.58 579.90 ± 56.43 440.80 ± 55.10 230.46 ± 32.25 389.79 ± 64.58 690.53 ± 91.72 526.79 ± 114.77 263.97 ± 28.88 404.83 ± 37.40
Noise (HU) 52.89 ± 2.94 74.46 ± 14.86 56.32 ± 7.36 53.56 ± 8.66 55.31 ± 5.60 82.43 ± 10.42 62.60 ± 9.58 56.43 ± 5.92 59.18 ± 4.59
Muscle
CT attenuation (HU) 61.04 ± 4.80 102.91 ± 20.22 86.78 ± 19.84 64.50 ± 6.70 64.69 ± 4.81 128.83 ± 13.92 103.32 ± 13.69 60.28 ± 10.86 65.35 ± 8.46
Noise (HU) 11.00 ± 1.04 28.93 ± 1.51 18.78 ± 2.78 16.73 ± 4.35 12.04 ± 1.49 31.90 ± 9.02 18.77 ± 3.61 25.51 ± 1.79 13.03 ± 1.70

Abbreviations: AI‐MAC, artificial intelligence metal artifacts correction; MAC, metal artifact correction; VMI, virtual monochromatic imaging.