Table 2.
Device | Hct level | Coefficient “C2” | Coefficient “C1” | Coefficient “B” | SEE (%) |
---|---|---|---|---|---|
Masimo radical | 40 | − 3.11E−03 | 1.29E+00 | 3.50E+00 | 0.35 |
30 | − 3.26E−03 | 1.31E+00 | 3.07E+00 | 0.36 | |
20 | − 3.31E−03 | 1.34E+00 | 1.03E+00 | 0.66 | |
10 | 1.93E−03 | 3.93E−01 | 4.42E+01 | 0.44 | |
Acare AH-M1 | 40 | − 9.91E−03 | 2.44E+00 | − 4.48E+01 | 0.51 |
30 | − 7.31E−03 | 2.06E+00 | − 3.20E+01 | 0.40 | |
20 | − 8.79E−03 | 2.36E+00 | − 4.60E+01 | 0.40 | |
10 | − 5.42E−03 | 1.95E+00 | − 3.81E+01 | 3.74 | |
CMS 50-DL | 40 | 2.98E−03 | 3.34E−01 | 3.84E+01 | 1.26 |
30 | − 1.50E−03 | 1.04E+00 | 1.16E+01 | 0.94 | |
20 | 2.56E−03 | 3.40E−01 | 4.24E+01 | 1.20 | |
10 | 1.64E−03 | 3.69E−01 | 4.69E+01 | 1.00 |
Higher SEE values indicate greater variability and thus decreased precision