Regression and validation data for the determination of harmaline and harmine using the proposed method.

Parameters Harmaline Harmine
Wavelength (nm) 419 456
Linearity range (ng/mL) 10–200 10–200
LOD (ng/mL) 2.51 1.82
LOQ (ng/mL) 7.53 5.46
Slope 0.011 0.024
Intercept 0.762 0.750
Coefficient of determination (r2) 0.9998 0.9997
Accuracy (%R)a 98.20 99.65
Repeatability Precision (RSD)b 0.805 0.878
Intermediate precision (RSD)b 0.780 0.799
Robustness (%R ± RSD)
 Buffer pH (± 0.5) 100.33 ± 0.804 98.55 ± 0.651
 Buffer volume (± 0.25 mL) 98.26 ± 0.708 99.20 ± 0.840
 HP-β-CD volume (± 0.1 mL) 98.54 ± 0.698 100.24 ± 0.925

aAverage of nine determinations (three concentrations repeated three times).

bRSD of nine determinations (three concentrations repeated three times).