. 2023 Nov 4;4(4):102691. doi: 10.1016/j.xpro.2023.102691

Table 2.

Orbitrap Lumos method summary

Global settings
 Method duration (min) 120
 Ion source type NSI
 Positive Ion (V) 2200
 Ion transfer tube temp (oC) 275
 Default charge state 2
Master scan
 Detector type Orbitrap
 Resolution 120,000
 Scan range (m/z) 350-1700
 Max injection time (ms) 50
 AGC target Standard
 Data type Profile
Filters
 Monoisotopic peak detection Peptide
 Intensity threshold 1.0 × 104
 Include charge state 2-6
 Exclude after n times 1
 Exclusion duration (s) 60
 Mass tolerance (ppm) 10
ddMS2
 Data dependent mode Cycle time
 Time between master scans (s) 2
 Isolation mode Quadrupole
 Isolation window (m/z) 1.6
 Activation type CID
 Collision energy (%) 30
 Detector type Ion trap
 Ion trap scan rate Rapid
 Max injection time (ms) Dynamic
 AGC target Standard